  "name": "Closed-wound surgical drainage evacuators: mechanical and regulatory specifications",
  "description": "Comparative technical specifications of closed-wound surgical drainage evacuators compiled from manufacturer instructions for use, AccessGUDID device records, and US FDA 510(k) premarket notifications. Compares reservoir mechanism, nominal volume capacity, negative pressure profile, FDA product code, and 21 CFR regulatory classification.",
  "page": "https://casrai.org/compare/accordion-drain-vs-hemovac",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-25T04:02:21Z",
  "temporalCoverage": "2026",
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Device model and manufacturer",
      "unit": "",
      "description": "Proprietary brand name, model series, and manufacturer of record"
    },
    {
      "name": "Reservoir mechanism",
      "unit": "",
      "description": "Mechanical component generating negative suction pressure"
    },
    {
      "name": "Nominal capacity",
      "unit": "mL",
      "description": "Maximum calibrated fluid collection volume"
    },
    {
      "name": "Typical negative pressure",
      "unit": "mmHg",
      "description": "Measured initial operational negative suction pressure"
    },
    {
      "name": "FDA regulation number",
      "unit": "",
      "description": "Title 21 Code of Federal Regulations classification section"
    },
    {
      "name": "FDA product code and class",
      "unit": "",
      "description": "Three-letter product code and regulatory device class"
    }
  ],
  "rows": [
    {
      "Device model and manufacturer": "Zimmer Hemovac 400 mL Evacuator (00-2500-700-10)",
      "Reservoir mechanism": "Internal stainless steel coil springs (3-spring assembly)",
      "Nominal capacity": 400,
      "Typical negative pressure": "-70 to -95",
      "FDA regulation number": "21 CFR 878.4680",
      "FDA product code and class": "GCY (Class I)"
    },
    {
      "Device model and manufacturer": "Snyder Mini Hemovac Kit (K800542)",
      "Reservoir mechanism": "Internal stainless steel coil springs (miniature format)",
      "Nominal capacity": 100,
      "Typical negative pressure": "-60 to -85",
      "FDA regulation number": "21 CFR 878.4680",
      "FDA product code and class": "GCY (Class I)"
    },
    {
      "Device model and manufacturer": "C. R. Bard ReliaVac 400 Evacuator (0070510)",
      "Reservoir mechanism": "Pleated polymer bellows with elastomeric memory",
      "Nominal capacity": 400,
      "Typical negative pressure": "-30 to -55",
      "FDA regulation number": "21 CFR 878.4680",
      "FDA product code and class": "GCY (Class I)"
    },
    {
      "Device model and manufacturer": "C. R. Bard ReliaVac 100 Evacuator (0090370)",
      "Reservoir mechanism": "Pleated polymer bellows with elastomeric memory",
      "Nominal capacity": 100,
      "Typical negative pressure": "-25 to -45",
      "FDA regulation number": "21 CFR 878.4680",
      "FDA product code and class": "GCY (Class I)"
    },
    {
      "Device model and manufacturer": "Medline 400 cc Bellows Closed Wound Evacuator (DYND400)",
      "Reservoir mechanism": "Corrugated accordion bellows with anti-reflux valve",
      "Nominal capacity": 400,
      "Typical negative pressure": "-30 to -50",
      "FDA regulation number": "21 CFR 878.4680",
      "FDA product code and class": "GCY (Class I)"
    },
    {
      "Device model and manufacturer": "Zimmer Hemovac Autotransfusion System (H.A.S. K041525)",
      "Reservoir mechanism": "Internal spring canister paired with blood filtration chamber",
      "Nominal capacity": 600,
      "Typical negative pressure": "-70 to -90",
      "FDA regulation number": "21 CFR 868.5830",
      "FDA product code and class": "CAC (Class II)"
    }
  ],
  "sources": [
    {
      "title": "AccessGUDID DI 00889024378834: Zimmer Hemovac Wound Drainage System",
      "url": "https://accessgudid.nlm.nih.gov/devices/00889024378834",
      "accessed": "2026-11-24"
    },
    {
      "title": "AccessGUDID DI 20801741049525: C. R. Bard ReliaVac 400 Closed Wound Suction Kit",
      "url": "https://accessgudid.nlm.nih.gov/devices/20801741049525",
      "accessed": "2026-11-24"
    },
    {
      "title": "openFDA Device Classification: 21 CFR 878.4680 (Product Code GCY)",
      "url": "https://api.fda.gov/device/classification.json?search=product_code:GCY",
      "accessed": "2026-11-24"
    },
    {
      "title": "openFDA 510(k) Premarket Notification K800577 (Snyder Hemovac Vented Wound Kit)",
      "url": "https://api.fda.gov/device/510k.json?search=k_number:K800577",
      "accessed": "2026-11-24"
    }
  ]
}