  "name": "Maximum permissible error of volumetric and graduated pipettes by class and nominal capacity",
  "description": "Maximum permissible errors in ml from the ISO 648:2008 overview (Class A volumetric examples at 1 and 10 ml) and Table 1 of EN ISO 835:2007 (graduated pipettes, Class A/AS and Class B, 1 to 25 ml), compiled into one long-format table.",
  "page": "https://casrai.org/compare/graduated-pipette-vs-volumetric-pipette",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-18T19:03:05Z",
  "temporalCoverage": "2007/2008",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Pipette standard and class",
      "unit": "",
      "description": ""
    },
    {
      "name": "Nominal capacity",
      "unit": "ml",
      "description": ""
    },
    {
      "name": "Maximum permissible error",
      "unit": "ml",
      "description": ""
    }
  ],
  "rows": [
    {
      "Pipette standard and class": "Volumetric pipette, Class A (ISO 648)",
      "Nominal capacity": 1,
      "Maximum permissible error": 0.008
    },
    {
      "Pipette standard and class": "Volumetric pipette, Class A (ISO 648)",
      "Nominal capacity": 10,
      "Maximum permissible error": 0.02
    },
    {
      "Pipette standard and class": "Graduated pipette, Class A or AS (ISO 835)",
      "Nominal capacity": 1,
      "Maximum permissible error": 0.007
    },
    {
      "Pipette standard and class": "Graduated pipette, Class A or AS (ISO 835)",
      "Nominal capacity": 2,
      "Maximum permissible error": 0.01
    },
    {
      "Pipette standard and class": "Graduated pipette, Class A or AS (ISO 835)",
      "Nominal capacity": 5,
      "Maximum permissible error": 0.03
    },
    {
      "Pipette standard and class": "Graduated pipette, Class A or AS (ISO 835)",
      "Nominal capacity": 10,
      "Maximum permissible error": 0.05
    },
    {
      "Pipette standard and class": "Graduated pipette, Class A or AS (ISO 835)",
      "Nominal capacity": 25,
      "Maximum permissible error": 0.1
    },
    {
      "Pipette standard and class": "Graduated pipette, Class B (ISO 835)",
      "Nominal capacity": 1,
      "Maximum permissible error": 0.01
    },
    {
      "Pipette standard and class": "Graduated pipette, Class B (ISO 835)",
      "Nominal capacity": 2,
      "Maximum permissible error": 0.02
    },
    {
      "Pipette standard and class": "Graduated pipette, Class B (ISO 835)",
      "Nominal capacity": 5,
      "Maximum permissible error": 0.05
    },
    {
      "Pipette standard and class": "Graduated pipette, Class B (ISO 835)",
      "Nominal capacity": 10,
      "Maximum permissible error": 0.1
    },
    {
      "Pipette standard and class": "Graduated pipette, Class B (ISO 835)",
      "Nominal capacity": 25,
      "Maximum permissible error": 0.2
    }
  ],
  "sources": [
    {
      "title": "ISO 648:2008 overview, iTeh Standards",
      "url": "https://standards.iteh.ai/catalog/standards/iso/034d83b6-819a-42cf-9e51-314841ad6998/iso-648-2008",
      "accessed": "2026-11-18"
    },
    {
      "title": "EN ISO 835:2007 overview, iTeh Standards",
      "url": "https://standards.iteh.ai/catalog/standards/cen/bea4bf98-fab1-437f-9df3-7d52c532bafb/en-iso-835-2007",
      "accessed": "2026-11-18"
    }
  ]
}