  "description": "Comparison matrix detailing the nine subsections of FDA 21 CFR 820.70 production and process controls, required regulatory documentation, critical inspection focus areas, and corresponding clauses under the Quality Management System Regulation (QMSR) incorporating ISO 13485:2016. Compiled from Title 21 of the Code of Federal Regulations, the QMSR Final Rule (89 FR 7496), and FDA software validation guidance.",
  "page": "https://casrai.org/guides/21-cfr-820-70",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-20T10:55:40Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "CFR Subsection",
      "unit": "",
      "description": "Section 820.70 subsection code and title"
    },
    {
      "name": "Core Requirement",
      "unit": "",
      "description": "Primary regulatory obligation mandated by FDA"
    },
    {
      "name": "Mandatory Records",
      "unit": "",
      "description": "Specific documented evidence required during an audit"
    },
    {
      "name": "Inspection Focus Area",
      "unit": "",
      "description": "Key items evaluated by FDA investigators during inspections"
    },
    {
      "name": "QMSR / ISO 13485 Alignment",
      "unit": "",
      "description": "Corresponding requirement in ISO 13485:2016 incorporated by reference under QMSR"
    }
  ],
  "rows": [
    {
      "CFR Subsection": "820.70(a) General Process Controls",
      "Core Requirement": "Develop, conduct, control, and monitor production processes to ensure device conforms to specifications",
      "Mandatory Records": "Documented SOPs, work instructions, routing sheets, approved golden samples, parameter logs",
      "Inspection Focus Area": "Shop floor adherence to procedures, parameter monitoring logs, workmanship defect catalogs",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 7.5.1 (Control of production and service provision)"
    },
    {
      "CFR Subsection": "820.70(b) Production and Process Changes",
      "Core Requirement": "Establish procedures for changes to specifications, methods, processes, or procedures with verification/validation",
      "Mandatory Records": "Engineering change orders (ECOs), change review records, validation reports per § 820.75, sign-offs",
      "Inspection Focus Area": "Unauthorized parameter tweaks, lack of re-validation before commercial release",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 7.3.9 (Control of design and development changes) and Clause 7.5.6"
    },
    {
      "CFR Subsection": "820.70(c) Environmental Control",
      "Core Requirement": "Control and monitor environmental conditions that could adversely affect product quality",
      "Mandatory Records": "Cleanroom monitoring procedures, viable/non-viable particulate logs, differential pressure charts, HVAC service logs",
      "Inspection Focus Area": "Uninvestigated cleanroom excursions, negative pressure issues, overdue HEPA certifications",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 6.4.1 (Work environment) and Clause 7.5.2 (Cleanliness of product)"
    },
    {
      "CFR Subsection": "820.70(d) Personnel Cleanliness & Health",
      "Core Requirement": "Establish requirements for health, hygiene, clothing, and gowning to prevent contamination",
      "Mandatory Records": "Gowning SOPs, hygiene training records, contractor escort logs, medical exclusion records",
      "Inspection Focus Area": "Operator gowning violations, exposed facial hair, untrained temporary technicians in cleanrooms",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 6.4.1 (Work environment - personnel requirements)"
    },
    {
      "CFR Subsection": "820.70(e) Contamination Control",
      "Core Requirement": "Establish procedures to prevent equipment or product contamination by foreign substances",
      "Mandatory Records": "Line clearance logs, cleanroom material transfer SOPs, tool cleaning records, waste handling procedures",
      "Inspection Focus Area": "Cardboard shipping containers in cleanrooms, chemical residue accumulation, mix-up prevention",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 7.5.2 (Cleanliness of product) and Clause 6.4.2 (Contamination control)"
    },
    {
      "CFR Subsection": "820.70(f) Buildings and Facility Design",
      "Core Requirement": "Maintain suitable building design, layout, and space to perform operations and prevent mix-ups",
      "Mandatory Records": "Facility layout diagrams, material flow maps, spatial segregation SOPs, quarantine area controls",
      "Inspection Focus Area": "Congestion in assembly areas, lack of physical segregation between quarantine and released stock",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 6.3 (Infrastructure) and Clause 6.4 (Work environment)"
    },
    {
      "CFR Subsection": "820.70(g) Equipment Maintenance & Adjustment",
      "Core Requirement": "Ensure equipment is clean, maintained, and inspected with posted allowable tolerances",
      "Mandatory Records": "Preventive maintenance schedules, technician work order logs, periodic inspection records, tolerance placards",
      "Inspection Focus Area": "Overdue maintenance tasks, missing tolerance stickers on adjustable fixtures, uncalibrated tools",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 6.3 (Infrastructure maintenance) and Clause 7.5.1"
    },
    {
      "CFR Subsection": "820.70(h) Manufacturing Material Management",
      "Core Requirement": "Control and remove manufacturing materials that could adversely affect device quality",
      "Mandatory Records": "Manufacturing material inventory, removal procedures, chemical residue test data, TOC wash logs",
      "Inspection Focus Area": "Unverified solvent or mold release residues, lack of residue toxicological evaluation",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 7.5.2 (Cleanliness of product - removal of manufacturing materials)"
    },
    {
      "CFR Subsection": "820.70(i) Automated Processes & Software",
      "Core Requirement": "Validate computer software used in production or quality systems for its intended use",
      "Mandatory Records": "Validation master plans, IQ/OQ/PQ protocols, challenge test results, change re-validation reports",
      "Inspection Focus Area": "Unvalidated spreadsheets for device release, untested PLC script updates, lack of protocol sign-off",
      "QMSR / ISO 13485 Alignment": "ISO 13485:2016 Clause 4.1.6 (QMS software validation) and Clause 7.5.6 (Process software validation)"
    }
  ],
  "sources": [
    {
      "title": "Legal Information Institute, Cornell Law School: 21 CFR § 820.70 - Production and process controls",
      "url": "https://www.law.cornell.edu/cfr/text/21/820.70",
      "accessed": "2026-11-20"
    },
    {
      "title": "U.S. Government Publishing Office: 21 CFR 820.70 - Production and process controls (eCFR XML)",
      "url": "https://www.govinfo.gov/content/pkg/CFR-2023-title21-vol8/xml/CFR-2023-title21-vol8-sec820-70.xml",
      "accessed": "2026-11-20"
    },
    {
      "title": "Federal Register: Medical Devices; Quality System Regulation Amendments (QMSR Final Rule, 89 FR 7496)",
      "url": "https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments",
      "accessed": "2026-11-20"
    },
    {
      "title": "U.S. Food and Drug Administration: Quality Management System Regulation (QMSR) Overview",
      "url": "https://www.fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr",
      "accessed": "2026-11-20"
    },
    {
      "title": "U.S. Food and Drug Administration: General Principles of Software Validation",
      "url": "https://www.fda.gov/regulatory-information/search-fda-guidance-documents/general-principles-software-validation",
      "accessed": "2026-11-20"
    }
  ]
}