  "name": "Aerogen Solo (vibrating mesh) versus PARI Vios with LC Sprint (compressor and jet nebulizer): attributes stated in FDA and manufacturer documents",
  "description": "Side-by-side of attributes that the FDA 510(k) summary K232507, the Aerogen Solo instruction manual Rev L and PARI's Vios with LC Sprint product page each state. Cells marked Not stated in source were not published in the document read; no values were estimated.",
  "page": "https://casrai.org/guides/aerogen-solo",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-13T01:22:28Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Attribute",
      "unit": "",
      "description": ""
    },
    {
      "name": "Aerogen Solo",
      "unit": "",
      "description": ""
    },
    {
      "name": "PARI Vios with LC Sprint",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Attribute": "Technology",
      "Aerogen Solo": "Vibrating mesh, approx. 128 kHz, electrically driven",
      "PARI Vios with LC Sprint": "Compressor-driven reusable nebulizer (LC Sprint)"
    },
    {
      "Attribute": "Compressed gas source needed",
      "Aerogen Solo": "No",
      "PARI Vios with LC Sprint": "Compressor supplied (Vios)"
    },
    {
      "Attribute": "Device role named by source",
      "Aerogen Solo": "Nebulizer (FDA product code CAF, Class 2, 21 CFR 868.5630)",
      "PARI Vios with LC Sprint": "Vios is the compressor; LC Sprint is the nebulizer"
    },
    {
      "Attribute": "Reuse",
      "Aerogen Solo": "Nebulizer single-patient use, not to be cleaned; controller reusable",
      "PARI Vios with LC Sprint": "Reusable nebulizer plus backup nebulizer in bundle"
    },
    {
      "Attribute": "Maximum reservoir capacity",
      "Aerogen Solo": "6 mL",
      "PARI Vios with LC Sprint": "Not stated in source"
    },
    {
      "Attribute": "Maximum pressure / flow stated",
      "Aerogen Solo": "Operates up to 90 cm H2O circuit pressure",
      "PARI Vios with LC Sprint": "Compressor max 34 PSI (2.3 bar), 8.5 L/min; nominal 18 PSI (1.2 bar), 4.5 L/min with PARI LC reusable nebulizer"
    },
    {
      "Attribute": "Weight",
      "Aerogen Solo": "Nebulizer 13.5 g",
      "PARI Vios with LC Sprint": "Compressor 3 lbs"
    },
    {
      "Attribute": "Age groups stated",
      "Aerogen Solo": "Pediatric 29 days or older, and adult",
      "PARI Vios with LC Sprint": "Adult and pediatric"
    },
    {
      "Attribute": "Setting stated",
      "Aerogen Solo": "Hospital; homecare on ventilation only for USB Controller System",
      "PARI Vios with LC Sprint": "Not stated in source"
    },
    {
      "Attribute": "HCPCS codes listed",
      "Aerogen Solo": "Not stated in source",
      "PARI Vios with LC Sprint": "E0570 (compressor), A7005 (reusable nebulizer)"
    },
    {
      "Attribute": "Warranty stated",
      "Aerogen Solo": "Not stated in source",
      "PARI Vios with LC Sprint": "5 years (compressor)"
    }
  ],
  "sources": [
    {
      "title": "FDA 510(k) K232507 decision letter and summary",
      "url": "https://www.accessdata.fda.gov/cdrh_docs/pdf23/K232507.pdf",
      "accessed": "2026-11-12"
    },
    {
      "title": "Aerogen Solo System Instruction Manual 30-674 Rev L (US)",
      "url": "https://www.hamilton-medical.com/dam/jcr:765bce7c-3777-4582-8c11-49b485e62e56/Aerogen-Solo-30-674-Rev-L-Aerogen-Solo-System-Instruction-Manual-US_WEB.pdf",
      "accessed": "2026-11-12"
    },
    {
      "title": "PARI Vios with LC Sprint product page",
      "url": "https://www.pari.com/us/products/nebulizer-systems-for-the-lungs-us/vios-with-lc-sprint/",
      "accessed": "2026-11-12"
    }
  ]
}