  "name": "Test-weight tolerances for analytical balance calibration: ASTM E617 Class 0 and 1 vs OIML R 111 E1, E2 and F1",
  "description": "Compares the tolerance (ASTM E617) or maximum permissible error (OIML R 111) in milligrams for eight nominal masses, side by side across the two weight-class systems. ASTM values are from a manufacturer's ASTM E617 tolerance chart; OIML values were cross-checked between two published R 111 (2004) tables. Rows where the two OIML tables disagree (20 g) were left out.",
  "page": "https://casrai.org/guides/analytical-balance-calibration-weighing-technique",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-17T18:45:33Z",
  "temporalCoverage": "R 111 edition 2004",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Nominal mass",
      "unit": "",
      "description": ""
    },
    {
      "name": "ASTM E617 Class 0 tolerance",
      "unit": "mg",
      "description": ""
    },
    {
      "name": "ASTM E617 Class 1 tolerance",
      "unit": "mg",
      "description": ""
    },
    {
      "name": "OIML R 111 E1 maximum permissible error",
      "unit": "mg",
      "description": ""
    },
    {
      "name": "OIML R 111 E2 maximum permissible error",
      "unit": "mg",
      "description": ""
    },
    {
      "name": "OIML R 111 F1 maximum permissible error",
      "unit": "mg",
      "description": ""
    }
  ],
  "rows": [
    {
      "Nominal mass": "1 mg",
      "ASTM E617 Class 0 tolerance": 0.005,
      "ASTM E617 Class 1 tolerance": 0.01,
      "OIML R 111 E1 maximum permissible error": 0.003,
      "OIML R 111 E2 maximum permissible error": 0.006,
      "OIML R 111 F1 maximum permissible error": 0.02
    },
    {
      "Nominal mass": "10 mg",
      "ASTM E617 Class 0 tolerance": 0.005,
      "ASTM E617 Class 1 tolerance": 0.01,
      "OIML R 111 E1 maximum permissible error": 0.003,
      "OIML R 111 E2 maximum permissible error": 0.008,
      "OIML R 111 F1 maximum permissible error": 0.025
    },
    {
      "Nominal mass": "100 mg",
      "ASTM E617 Class 0 tolerance": 0.005,
      "ASTM E617 Class 1 tolerance": 0.01,
      "OIML R 111 E1 maximum permissible error": 0.005,
      "OIML R 111 E2 maximum permissible error": 0.016,
      "OIML R 111 F1 maximum permissible error": 0.05
    },
    {
      "Nominal mass": "1 g",
      "ASTM E617 Class 0 tolerance": 0.017,
      "ASTM E617 Class 1 tolerance": 0.034,
      "OIML R 111 E1 maximum permissible error": 0.01,
      "OIML R 111 E2 maximum permissible error": 0.03,
      "OIML R 111 F1 maximum permissible error": 0.1
    },
    {
      "Nominal mass": "10 g",
      "ASTM E617 Class 0 tolerance": 0.025,
      "ASTM E617 Class 1 tolerance": 0.05,
      "OIML R 111 E1 maximum permissible error": 0.02,
      "OIML R 111 E2 maximum permissible error": 0.06,
      "OIML R 111 F1 maximum permissible error": 0.2
    },
    {
      "Nominal mass": "100 g",
      "ASTM E617 Class 0 tolerance": 0.13,
      "ASTM E617 Class 1 tolerance": 0.25,
      "OIML R 111 E1 maximum permissible error": 0.05,
      "OIML R 111 E2 maximum permissible error": 0.16,
      "OIML R 111 F1 maximum permissible error": 0.5
    },
    {
      "Nominal mass": "200 g",
      "ASTM E617 Class 0 tolerance": 0.25,
      "ASTM E617 Class 1 tolerance": 0.5,
      "OIML R 111 E1 maximum permissible error": 0.1,
      "OIML R 111 E2 maximum permissible error": 0.3,
      "OIML R 111 F1 maximum permissible error": 1
    },
    {
      "Nominal mass": "1 kg",
      "ASTM E617 Class 0 tolerance": 1.3,
      "ASTM E617 Class 1 tolerance": 2.5,
      "OIML R 111 E1 maximum permissible error": 0.5,
      "OIML R 111 E2 maximum permissible error": 1.6,
      "OIML R 111 F1 maximum permissible error": 5
    }
  ],
  "sources": [
    {
      "title": "Troemner: Calibration Weight Tolerance (ASTM E617 and OIML R-111 tables)",
      "url": "https://www.troemner.com/reference-center/weights-reference-center/calibration-weight-tolerance",
      "accessed": "2026-11-17"
    },
    {
      "title": "Gram Group: OIML guidelines (R 111-2004 maximum permissible errors)",
      "url": "https://gram-group.com/oiml-guidelines/",
      "accessed": "2026-11-17"
    }
  ]
}