  "name": "Anion Exchange Chromatography Column Technical Specifications and Operating Limits",
  "description": "Comparative specifications and operating parameters for leading analytical and preparative anion exchange chromatography columns compiled from manufacturer instructions for use and technical manuals.",
  "page": "https://casrai.org/guides/anion-exchange-column",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-16T04:25:40Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Column Model",
      "unit": "",
      "description": "Manufacturer column product name and format"
    },
    {
      "name": "Manufacturer",
      "unit": "",
      "description": "Primary manufacturer or vendor"
    },
    {
      "name": "Exchanger Type",
      "unit": "",
      "description": "Functional classification (Strong Anion Exchanger or Weak Anion Exchanger)"
    },
    {
      "name": "Functional Ligand",
      "unit": "",
      "description": "Surface active chemical functional group"
    },
    {
      "name": "Base Matrix",
      "unit": "",
      "description": "Support bead material and structure"
    },
    {
      "name": "Particle Size",
      "unit": "µm",
      "description": "Mean bead diameter"
    },
    {
      "name": "Operating pH Range",
      "unit": "",
      "description": "Working chemical stability pH interval"
    },
    {
      "name": "Max Pressure",
      "unit": "bar",
      "description": "Maximum rated column or bed backpressure"
    }
  ],
  "rows": [
    {
      "Column Model": "Cytiva Mono Q 10/100 GL",
      "Manufacturer": "Cytiva",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Quaternary ammonium (–N+(CH3)3)",
      "Base Matrix": "Polystyrene/divinylbenzene (MonoBeads)",
      "Particle Size": 10,
      "Operating pH Range": "2–12",
      "Max Pressure": 40
    },
    {
      "Column Model": "Cytiva HiTrap Q HP (1 mL)",
      "Manufacturer": "Cytiva",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Quaternary ammonium (–N+(CH3)3)",
      "Base Matrix": "Spherical agarose (High Performance)",
      "Particle Size": 34,
      "Operating pH Range": "2–12",
      "Max Pressure": 3
    },
    {
      "Column Model": "Cytiva HiTrap Q FF (1 mL)",
      "Manufacturer": "Cytiva",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Quaternary ammonium (–N+(CH3)3)",
      "Base Matrix": "Highly cross-linked agarose (Fast Flow)",
      "Particle Size": 90,
      "Operating pH Range": "2–12",
      "Max Pressure": 3
    },
    {
      "Column Model": "Cytiva Capto Q",
      "Manufacturer": "Cytiva",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Quaternary ammonium (–N+(CH3)3)",
      "Base Matrix": "High-flow cross-linked agarose with dextran",
      "Particle Size": 90,
      "Operating pH Range": "2–12",
      "Max Pressure": 3
    },
    {
      "Column Model": "Thermo Dionex IonPac AS11-HC",
      "Manufacturer": "Thermo Scientific",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Alkanol quaternary ammonium latex",
      "Base Matrix": "Ethylvinylbenzene/55% divinylbenzene",
      "Particle Size": 9,
      "Operating pH Range": "0–14",
      "Max Pressure": 345
    },
    {
      "Column Model": "Tosoh TSKgel SuperQ-5PW",
      "Manufacturer": "Tosoh Bioscience",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Quaternary ammonium / polyamine",
      "Base Matrix": "Hydroxylated polymethacrylate",
      "Particle Size": 10,
      "Operating pH Range": "2–12",
      "Max Pressure": 20
    },
    {
      "Column Model": "Tosoh TSKgel DEAE-5PW",
      "Manufacturer": "Tosoh Bioscience",
      "Exchanger Type": "Weak (WAX)",
      "Functional Ligand": "Diethylaminoethyl (–CH2CH2N+(C2H5)2H)",
      "Base Matrix": "Hydroxylated polymethacrylate",
      "Particle Size": 10,
      "Operating pH Range": "2–12",
      "Max Pressure": 20
    },
    {
      "Column Model": "Bio-Rad ENrich Q (5 x 50 mm)",
      "Manufacturer": "Bio-Rad",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Quaternary ammonium (–N+(CH3)3)",
      "Base Matrix": "Porous polymethacrylate",
      "Particle Size": 10,
      "Operating pH Range": "2–12",
      "Max Pressure": 34
    },
    {
      "Column Model": "Waters Protein-Pak Hi Res Q",
      "Manufacturer": "Waters Corporation",
      "Exchanger Type": "Strong (SAX)",
      "Functional Ligand": "Quaternary ammonium (Q)",
      "Base Matrix": "Non-porous polymethacrylate",
      "Particle Size": 5,
      "Operating pH Range": "3–10",
      "Max Pressure": 150
    }
  ],
  "sources": [
    {
      "title": "Cytiva Mono Q Instructions 71-7053-00",
      "url": "https://www.cytivalifesciences.com",
      "accessed": "2026-11-16"
    },
    {
      "title": "Cytiva HiTrap Q HP Instructions 71-7006-00",
      "url": "https://www.cytivalifesciences.com",
      "accessed": "2026-11-16"
    },
    {
      "title": "Cytiva Capto Q Data File 28-9037-77",
      "url": "https://www.cytivalifesciences.com",
      "accessed": "2026-11-16"
    },
    {
      "title": "Thermo Scientific Dionex IonPac AS11-HC Product Manual 031333",
      "url": "https://www.thermofisher.com",
      "accessed": "2026-11-16"
    },
    {
      "title": "Tosoh Bioscience TSKgel 5PW Column Instruction Manual",
      "url": "https://www.tosohbioscience.com",
      "accessed": "2026-11-16"
    },
    {
      "title": "Bio-Rad ENrich High-Resolution Ion Exchange Columns Instruction Manual 10027798",
      "url": "https://www.bio-rad.com",
      "accessed": "2026-11-16"
    },
    {
      "title": "Waters Protein-Pak Hi Res Columns Care and Use Manual 720002937EN",
      "url": "https://www.waters.com",
      "accessed": "2026-11-16"
    }
  ]
}