  "name": "ASTM standards named with nitinol F2063 and the role each plays",
  "description": "Original compilation of ASTM designations that appear on the ASTM F2063-26 store page or in FDA's nitinol non-clinical guidance, with the property each addresses and the role the source gives it. Compiled 2026-11-24 from those two sources only; standard text and limits are not reproduced.",
  "page": "https://casrai.org/guides/astm-f2063",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-25T00:04:56Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "ASTM designation",
      "unit": "",
      "description": ""
    },
    {
      "name": "Subject",
      "unit": "",
      "description": ""
    },
    {
      "name": "Role given by the source",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "ASTM designation": "F2063",
      "Subject": "Wrought nickel-titanium shape memory alloys for medical devices and surgical implants",
      "Role given by the source": "Material specification for mill product (bar, plate, coil, tubing); named by FDA as an example consensus standard for composition",
      "Source": "ASTM store page; FDA guidance"
    },
    {
      "ASTM designation": "F2005",
      "Subject": "Terminology for nickel-titanium shape memory alloys",
      "Role given by the source": "Defines nitinol terminology; cited by FDA for its definitions",
      "Source": "FDA guidance"
    },
    {
      "ASTM designation": "F2633",
      "Subject": "Finished nickel-titanium tube",
      "Role given by the source": "ASTM states finished NiTi tube is covered here rather than by F2063",
      "Source": "ASTM store page"
    },
    {
      "ASTM designation": "F2004",
      "Subject": "Transformation temperature of NiTi alloys by thermal analysis",
      "Role given by the source": "Recommended method for transformation temperatures; may not give the same Af as F2082",
      "Source": "FDA guidance"
    },
    {
      "ASTM designation": "F2082",
      "Subject": "Transformation temperature by bend and free recovery",
      "Role given by the source": "Recommended for load-bearing devices when modified to the device's expected strains",
      "Source": "FDA guidance"
    },
    {
      "ASTM designation": "F2516",
      "Subject": "Tension testing of nickel-titanium superelastic materials",
      "Role given by the source": "Source of material-model parameters for computational analyses",
      "Source": "FDA guidance"
    },
    {
      "ASTM designation": "F2477",
      "Subject": "In vitro pulsatile durability testing of vascular stents",
      "Role given by the source": "Cited as an example of pressure control with displacement monitoring in radial pulsatile fatigue testing",
      "Source": "FDA guidance"
    },
    {
      "ASTM designation": "F2129",
      "Subject": "Cyclic potentiodynamic polarization for corrosion susceptibility of small implant devices",
      "Role given by the source": "Recommended method for pitting corrosion testing after simulated assembly and implantation; has no acceptance criterion of its own",
      "Source": "FDA guidance"
    },
    {
      "ASTM designation": "F3306",
      "Subject": "Ion release evaluation of medical implants",
      "Role given by the source": "Recommended method for nickel ion release testing when corrosion criteria are not met",
      "Source": "FDA guidance"
    }
  ],
  "sources": [
    {
      "title": "ASTM International: F2063-26 store page",
      "url": "https://store.astm.org/f2063-26.html",
      "accessed": "2026-11-24"
    },
    {
      "title": "FDA: Technical Considerations for Non-Clinical Assessment of Medical Devices Containing Nitinol",
      "url": "https://www.fda.gov/media/123272/download",
      "accessed": "2026-11-24"
    }
  ]
}