  "name": "Research Vibrating Microtomes for Acute Brain Slicing: Technical Specifications and Calibration Standards",
  "description": "Comparative technical specifications, blade oscillation parameters, vertical Z-axis deflection limits, and tissue stabilization mechanisms across laboratory vibrating microtomes utilized for acute brain slice preparation in electrophysiology.",
  "page": "https://casrai.org/guides/brain-slicing",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-14T21:27:43Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Microtome Model",
      "unit": "",
      "description": "Manufacturer and model designation of the vibrating blade microtome"
    },
    {
      "name": "Z-Axis Deflection Limit",
      "unit": "µm",
      "description": "Maximum calibrated vertical blade deflection during oscillation"
    },
    {
      "name": "Oscillation Frequency Min",
      "unit": "Hz",
      "description": "Minimum horizontal blade oscillation frequency"
    },
    {
      "name": "Oscillation Frequency Max",
      "unit": "Hz",
      "description": "Maximum horizontal blade oscillation frequency"
    },
    {
      "name": "Advance Speed Min",
      "unit": "mm/s",
      "description": "Minimum motorized forward blade travel speed"
    },
    {
      "name": "Advance Speed Max",
      "unit": "mm/s",
      "description": "Maximum motorized forward blade travel speed"
    },
    {
      "name": "Tissue Stabilization Method",
      "unit": "",
      "description": "Mechanism used to secure and stabilize soft neural tissue during sectioning"
    },
    {
      "name": "Deflection Calibration Device",
      "unit": "",
      "description": "Integrated or optional measurement tool for verifying vertical blade wobble"
    }
  ],
  "rows": [
    {
      "Microtome Model": "Leica VT1200 S",
      "Z-Axis Deflection Limit": 1,
      "Oscillation Frequency Min": 85,
      "Oscillation Frequency Max": 85,
      "Advance Speed Min": 0.01,
      "Advance Speed Max": 1.5,
      "Tissue Stabilization Method": "Cyanoacrylate glue on magnetic specimen plate with optional agar block backing",
      "Deflection Calibration Device": "Optical VibroCheck measurement sensor with micrometer tuning screw"
    },
    {
      "Microtome Model": "Precisionary Compresstome VF-310-0Z",
      "Z-Axis Deflection Limit": 1,
      "Oscillation Frequency Min": 50,
      "Oscillation Frequency Max": 65,
      "Advance Speed Min": 0.01,
      "Advance Speed Max": 5,
      "Tissue Stabilization Method": "Cylindrical specimen tube with 2.0-2.5% low-melting-point agarose embedding and radial compression",
      "Deflection Calibration Device": "Patented Auto Zero-Z factory-aligned mechanical suspension"
    },
    {
      "Microtome Model": "Campden Instruments 7000smz-2",
      "Z-Axis Deflection Limit": 1,
      "Oscillation Frequency Min": 50,
      "Oscillation Frequency Max": 120,
      "Advance Speed Min": 0.01,
      "Advance Speed Max": 2,
      "Tissue Stabilization Method": "Cyanoacrylate glue on specimen dish with integrated Peltier cold-bath cooling",
      "Deflection Calibration Device": "Optoelectronic vertical deflection verifier with calibrated handheld meter"
    },
    {
      "Microtome Model": "Thomas Scientific Vibratome 1000 Classic",
      "Z-Axis Deflection Limit": 4,
      "Oscillation Frequency Min": 0,
      "Oscillation Frequency Max": 60,
      "Advance Speed Min": 0.05,
      "Advance Speed Max": 2.5,
      "Tissue Stabilization Method": "Specimen vice clamp or cyanoacrylate mounting on Delrin stage",
      "Deflection Calibration Device": "Manual mechanical micrometer alignment"
    },
    {
      "Microtome Model": "Harvard Apparatus Oscillating Tissue Slicer",
      "Z-Axis Deflection Limit": 3.5,
      "Oscillation Frequency Min": 10,
      "Oscillation Frequency Max": 70,
      "Advance Speed Min": 0.02,
      "Advance Speed Max": 1.8,
      "Tissue Stabilization Method": "Removable magnetic specimen chuck with surrounding fluid bath",
      "Deflection Calibration Device": "Factory preset mechanical alignment without active optical sensor"
    }
  ],
  "sources": [
    {
      "title": "Leica VT1200 S Technical Specifications",
      "url": "https://www.leicabiosystems.com",
      "accessed": "2026-11-14"
    },
    {
      "title": "Precisionary Compresstome VF-310-0Z User Guide",
      "url": "https://precisionary.com",
      "accessed": "2026-11-14"
    },
    {
      "title": "Campden Instruments 7000smz-2 Technical Data",
      "url": "https://campdeninstruments.com",
      "accessed": "2026-11-14"
    },
    {
      "title": "Preparation of Acute Brain Slices, CSH Protocols",
      "url": "http://cshprotocols.cshlp.org",
      "accessed": "2026-11-14"
    }
  ]
}