  "name": "BRET system variants compared: donor, substrate, acceptor, assay duration and stated limits",
  "description": "Compares six BRET systems (BRET1, BRET2, eBRET, BRET3, NanoBRET, extended NanoBRET) on luciferase donor, emission peak, substrate, common acceptors, approximate assay duration and the limitation each review states. Compiled by hand from Table 1 of two open-access review articles (Molecules 2019 and Frontiers in Bioengineering and Biotechnology 2019); where the reviews differ slightly, both values are shown.",
  "page": "https://casrai.org/guides/bret-assay",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-26T11:30:00Z",
  "temporalCoverage": "2019",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "BRET system",
      "unit": "",
      "description": ""
    },
    {
      "name": "Luciferase donor",
      "unit": "",
      "description": ""
    },
    {
      "name": "Donor emission peak",
      "unit": "nm",
      "description": ""
    },
    {
      "name": "Substrate",
      "unit": "",
      "description": ""
    },
    {
      "name": "Common acceptors (emission peak, nm)",
      "unit": "",
      "description": ""
    },
    {
      "name": "Approximate assay duration",
      "unit": "",
      "description": ""
    },
    {
      "name": "Stated limitation",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "BRET system": "BRET1",
      "Luciferase donor": "Renilla luciferase (Rluc/Rluc8)",
      "Donor emission peak": "475-480",
      "Substrate": "Coelenterazine h",
      "Common acceptors (emission peak, nm)": "YFP or Venus (527-530); eYFP (530)",
      "Approximate assay duration": "About 1 h",
      "Stated limitation": "Sensitive to solvent polarity, serum and pH; not suited to binding studies needing extracellular luciferase tagging"
    },
    {
      "BRET system": "BRET2",
      "Luciferase donor": "Renilla luciferase (Rluc/Rluc8)",
      "Donor emission peak": "395-400",
      "Substrate": "Coelenterazine 400a (DeepBlueC)",
      "Common acceptors (emission peak, nm)": "GFP10 or GFP2 (about 510)",
      "Approximate assay duration": "Seconds",
      "Stated limitation": "Weak, short-lived light output; needs high expression of partners"
    },
    {
      "BRET system": "eBRET",
      "Luciferase donor": "Renilla luciferase (Rluc/Rluc8)",
      "Donor emission peak": "475-480",
      "Substrate": "EnduRen",
      "Common acceptors (emission peak, nm)": "YFP or Venus (527-530)",
      "Approximate assay duration": "Over 6 h",
      "Stated limitation": "Same as BRET1 plus substrate pre-incubation; EnduRen is expensive per the Molecules review"
    },
    {
      "BRET system": "BRET3",
      "Luciferase donor": "Firefly luciferase (FLuc, 62 kDa)",
      "Donor emission peak": "565",
      "Substrate": "D-luciferin",
      "Common acceptors (emission peak, nm)": "DsRed (583); Cy3/Cy3.5 (570/596)",
      "Approximate assay duration": "Not stated",
      "Stated limitation": "Overlap of donor and acceptor emission; low signal-to-noise; FLuc sensitive to ATP, temperature and ionic strength"
    },
    {
      "BRET system": "NanoBRET",
      "Luciferase donor": "Nanoluciferase (NanoLuc/Nluc, 19 kDa)",
      "Donor emission peak": "About 460 (462 in Molecules)",
      "Substrate": "Furimazine",
      "Common acceptors (emission peak, nm)": "HaloTag ligand (618 in Molecules; 635 for HT-NCT in Frontiers); Venus (527); BODIPY; TAMRA (about 579)",
      "Approximate assay duration": "About 2 h",
      "Stated limitation": "Bright signal can saturate the detector; substrate pre-incubation; furimazine described as expensive in 2019"
    },
    {
      "BRET system": "Extended NanoBRET (eNanoBRET)",
      "Luciferase donor": "Nanoluciferase (NanoLuc/Nluc, 19 kDa)",
      "Donor emission peak": "About 460",
      "Substrate": "Endurazine (Vivazine)",
      "Common acceptors (emission peak, nm)": "HT-NCT (635); Venus (527); BODIPY; TAMRA (about 579)",
      "Approximate assay duration": "Over 6 h",
      "Stated limitation": "Detector saturation and substrate pre-incubation requirement"
    }
  ],
  "sources": [
    {
      "title": "El Khamlichi et al., Molecules 2019: Bioluminescence Resonance Energy Transfer as a Method to Study Protein-Protein Interactions",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC6384791/",
      "accessed": "2026-11-20"
    },
    {
      "title": "Frontiers in Bioengineering and Biotechnology 2019: NanoBRET: The Bright Future of Proximity-Based Assays",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC6443706/",
      "accessed": "2026-11-20"
    }
  ]
}