  "name": "Zwitterionic laboratory buffers: pKa, temperature coefficient, useful pH range and autoclave guidance",
  "description": "Compiled from a university buffer reference table (pKa at 20 C and dpKa/dT per C, useful pH range) and a manufacturer buffer guide (autoclave statements). The pKa shift column is computed as dpKa/dT multiplied by +21 C (4 C to 25 C); it is an approximation that assumes a linear coefficient.",
  "page": "https://casrai.org/guides/buffer-and-solution-preparation-lab-guide",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T02:26:34Z",
  "temporalCoverage": "2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Buffer",
      "unit": "",
      "description": ""
    },
    {
      "name": "pKa at 20 C",
      "unit": "",
      "description": ""
    },
    {
      "name": "dpKa/dT",
      "unit": "per degC",
      "description": ""
    },
    {
      "name": "Useful pH range",
      "unit": "pH",
      "description": ""
    },
    {
      "name": "Computed pKa shift for a 21 C rise (4 to 25 C)",
      "unit": "pH units",
      "description": ""
    },
    {
      "name": "Autoclave statement in manufacturer guide",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Buffer": "Tris",
      "pKa at 20 C": 8.08,
      "dpKa/dT": -0.028,
      "Useful pH range": "7.0-9.0",
      "Computed pKa shift for a 21 C rise (4 to 25 C)": -0.59,
      "Autoclave statement in manufacturer guide": "Not stated in sources"
    },
    {
      "Buffer": "HEPES",
      "pKa at 20 C": 7.55,
      "dpKa/dT": -0.014,
      "Useful pH range": "7.0-8.0",
      "Computed pKa shift for a 21 C rise (4 to 25 C)": -0.29,
      "Autoclave statement in manufacturer guide": "Cannot be autoclaved (degrades on heating)"
    },
    {
      "Buffer": "MOPS",
      "pKa at 20 C": 7.2,
      "dpKa/dT": -0.011,
      "Useful pH range": "6.5-7.9",
      "Computed pKa shift for a 21 C rise (4 to 25 C)": -0.23,
      "Autoclave statement in manufacturer guide": "Cannot be autoclaved (degrades on heating)"
    },
    {
      "Buffer": "MES",
      "pKa at 20 C": 6.15,
      "dpKa/dT": -0.011,
      "Useful pH range": "5.8-6.5",
      "Computed pKa shift for a 21 C rise (4 to 25 C)": -0.23,
      "Autoclave statement in manufacturer guide": "Not stated in sources"
    },
    {
      "Buffer": "PIPES",
      "pKa at 20 C": 6.82,
      "dpKa/dT": -0.009,
      "Useful pH range": "6.4-7.2",
      "Computed pKa shift for a 21 C rise (4 to 25 C)": -0.19,
      "Autoclave statement in manufacturer guide": "Not stated in sources"
    },
    {
      "Buffer": "TES",
      "pKa at 20 C": 7.5,
      "dpKa/dT": -0.02,
      "Useful pH range": "7.0-8.0",
      "Computed pKa shift for a 21 C rise (4 to 25 C)": -0.42,
      "Autoclave statement in manufacturer guide": "Not stated in sources"
    },
    {
      "Buffer": "Bicine",
      "pKa at 20 C": 8.35,
      "dpKa/dT": -0.018,
      "Useful pH range": "7.8-8.8",
      "Computed pKa shift for a 21 C rise (4 to 25 C)": -0.38,
      "Autoclave statement in manufacturer guide": "Not stated in sources"
    }
  ],
  "sources": [
    {
      "title": "Zwitterionic Buffer table, Vanderbilt University Chemistry (Rizzo group)",
      "url": "https://www.vanderbilt.edu/AnS/Chemistry/Rizzo/stuff/Buffers/buffers.html",
      "accessed": "2026-11-20"
    },
    {
      "title": "Buffers for Biochemical Reactions, Promega",
      "url": "https://www.promega.com/Resources/Guides/lab-equipment-and-supplies/buffers-for-biochemical-reactions",
      "accessed": "2026-11-20"
    }
  ]
}