  "name": "Class A Volumetric Pipette Tolerances and Specifications Across ASTM E969 and ISO 648",
  "description": "Original compilation comparing volumetric capacity tolerances, color coding, minimum outflow durations, and waiting times between ASTM E969 and ISO 648 standards across nominal volumes from 0.5 mL to 100 mL, compiled from ASTM E969-02(2017) and ISO 648:2008 specifications.",
  "page": "https://casrai.org/guides/class-a-volumetric-pipette",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-14T11:20:12Z",
  "temporalCoverage": "2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Nominal Capacity",
      "unit": "mL",
      "description": "Nominal liquid volume delivered by the pipette"
    },
    {
      "name": "ASTM E969 Class A Tolerance",
      "unit": "mL",
      "description": "Maximum permissible capacity error under ASTM E969 Class A"
    },
    {
      "name": "ISO 648 Class A/AS Tolerance",
      "unit": "mL",
      "description": "Maximum permissible capacity error under ISO 648 Class A and AS"
    },
    {
      "name": "ASTM E969 Class B Tolerance",
      "unit": "mL",
      "description": "Maximum permissible capacity error under ASTM E969 Class B"
    },
    {
      "name": "ASTM Color Code",
      "unit": "",
      "description": "Identification band color conforming to ASTM E1273"
    },
    {
      "name": "ASTM Min Outflow Time",
      "unit": "s",
      "description": "Minimum outflow time for delivery under ASTM E969"
    },
    {
      "name": "ISO Class AS Waiting Time",
      "unit": "s",
      "description": "Mandatory waiting time after flow cessation under ISO 648 Class AS"
    }
  ],
  "rows": [
    {
      "Nominal Capacity": "0.5",
      "ASTM E969 Class A Tolerance": 0.006,
      "ISO 648 Class A/AS Tolerance": 0.005,
      "ASTM E969 Class B Tolerance": 0.012,
      "ASTM Color Code": "Two Black Bands",
      "ASTM Min Outflow Time": 5,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "1",
      "ASTM E969 Class A Tolerance": 0.006,
      "ISO 648 Class A/AS Tolerance": 0.008,
      "ASTM E969 Class B Tolerance": 0.012,
      "ASTM Color Code": "Blue",
      "ASTM Min Outflow Time": 8,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "2",
      "ASTM E969 Class A Tolerance": 0.006,
      "ISO 648 Class A/AS Tolerance": 0.01,
      "ASTM E969 Class B Tolerance": 0.012,
      "ASTM Color Code": "Orange",
      "ASTM Min Outflow Time": 8,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "3",
      "ASTM E969 Class A Tolerance": 0.01,
      "ISO 648 Class A/AS Tolerance": 0.01,
      "ASTM E969 Class B Tolerance": 0.02,
      "ASTM Color Code": "Black",
      "ASTM Min Outflow Time": 10,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "4",
      "ASTM E969 Class A Tolerance": 0.01,
      "ISO 648 Class A/AS Tolerance": 0.015,
      "ASTM E969 Class B Tolerance": 0.02,
      "ASTM Color Code": "Two Red Bands",
      "ASTM Min Outflow Time": 10,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "5",
      "ASTM E969 Class A Tolerance": 0.01,
      "ISO 648 Class A/AS Tolerance": 0.015,
      "ASTM E969 Class B Tolerance": 0.02,
      "ASTM Color Code": "White",
      "ASTM Min Outflow Time": 10,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "10",
      "ASTM E969 Class A Tolerance": 0.02,
      "ISO 648 Class A/AS Tolerance": 0.02,
      "ASTM E969 Class B Tolerance": 0.04,
      "ASTM Color Code": "Red",
      "ASTM Min Outflow Time": 15,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "15",
      "ASTM E969 Class A Tolerance": 0.03,
      "ISO 648 Class A/AS Tolerance": 0.03,
      "ASTM E969 Class B Tolerance": 0.06,
      "ASTM Color Code": "Green",
      "ASTM Min Outflow Time": 15,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "20",
      "ASTM E969 Class A Tolerance": 0.03,
      "ISO 648 Class A/AS Tolerance": 0.03,
      "ASTM E969 Class B Tolerance": 0.06,
      "ASTM Color Code": "Yellow",
      "ASTM Min Outflow Time": 20,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "25",
      "ASTM E969 Class A Tolerance": 0.03,
      "ISO 648 Class A/AS Tolerance": 0.03,
      "ASTM E969 Class B Tolerance": 0.06,
      "ASTM Color Code": "Blue",
      "ASTM Min Outflow Time": 20,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "50",
      "ASTM E969 Class A Tolerance": 0.05,
      "ISO 648 Class A/AS Tolerance": 0.05,
      "ASTM E969 Class B Tolerance": 0.1,
      "ASTM Color Code": "Red",
      "ASTM Min Outflow Time": 25,
      "ISO Class AS Waiting Time": 5
    },
    {
      "Nominal Capacity": "100",
      "ASTM E969 Class A Tolerance": 0.08,
      "ISO 648 Class A/AS Tolerance": 0.08,
      "ASTM E969 Class B Tolerance": 0.16,
      "ASTM Color Code": "Yellow",
      "ASTM Min Outflow Time": 30,
      "ISO Class AS Waiting Time": 5
    }
  ],
  "sources": [
    {
      "title": "ASTM E969-02(2017) Standard Specification for Glass Volumetric (Transfer) Pipets",
      "url": "https://www.astm.org/e0969-02r17.html",
      "accessed": "2026-11-14"
    },
    {
      "title": "ISO 648:2008 Laboratory glassware — Single-volume pipettes",
      "url": "https://www.iso.org/standard/36424.html",
      "accessed": "2026-11-14"
    }
  ]
}