  "name": "Current-clamp amplifier circuit elements and their functions",
  "description": "Compiled from Molecular Devices documentation (Axon Guide chapter 3 and the Axopatch 200B datasheet): each circuit element or setting of a current-clamp amplifier with its stated function.",
  "page": "https://casrai.org/guides/current-clamp",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-13T08:00:42Z",
  "temporalCoverage": "2012/2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Circuit element",
      "unit": "",
      "description": ""
    },
    {
      "name": "Stated function",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source section",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Circuit element": "Unity-gain buffer amplifier",
      "Stated function": "Reads pipette potential with input resistance far above pipette and cell, so pipette current is clamped near zero",
      "Source section": "Axon Guide ch. 3"
    },
    {
      "Circuit element": "Current source (A2 with output resistor Ro)",
      "Stated function": "Injects current I = Vcmd/Ro into the cell through the pipette; pulse, DC or any waveform",
      "Source section": "Axon Guide ch. 3"
    },
    {
      "Circuit element": "Bridge balance",
      "Stated function": "Subtracts a signal proportional to pipette current x pipette resistance from the buffer output",
      "Source section": "Axon Guide ch. 3"
    },
    {
      "Circuit element": "Track circuit",
      "Stated function": "Adjusts pipette offset dynamically so current is zero on average; must be off when recording ionic currents",
      "Source section": "Axon Guide ch. 3"
    },
    {
      "Circuit element": "Current monitor via series resistor",
      "Stated function": "Measures the voltage drop across a series resistor so saturation with a very high-resistance pipette is visible",
      "Source section": "Axon Guide ch. 3"
    },
    {
      "Circuit element": "Discontinuous current clamp (DCC)",
      "Stated function": "Same circuitry as dSEVC with experimenter-set pulse amplitude; more noise and lower bandwidth than bridge mode",
      "Source section": "Axon Guide ch. 3"
    },
    {
      "Circuit element": "Dual-speed current clamp",
      "Stated function": "Allows faster current clamping in small cells on the Axopatch 200B",
      "Source section": "Axopatch 200B datasheet"
    }
  ],
  "sources": [
    {
      "title": "The Axon Guide to Electrophysiology and Biophysics Laboratory Techniques (1-2500-0102 D)",
      "url": "https://www.moleculardevices.com/axon-guide",
      "accessed": "2026-11-13"
    },
    {
      "title": "Axopatch 200B datasheet",
      "url": "https://www.moleculardevices.com/sites/default/files/en/asset/dd/data-sheets/axon-axopatch-200b-datasheet-rev-b.pdf",
      "accessed": "2026-11-13"
    }
  ]
}