  "name": "DAPI Staining Parameters Across Biological Specimen Types and Assays",
  "description": "Operational staining parameters for DAPI across six biological specimen preparations and detection modalities, compiled from manufacturer technical datasheets (Thermo Fisher Scientific Molecular Probes, Cell Signaling Technology, Biotium) and peer-reviewed methods literature.",
  "page": "https://casrai.org/guides/dapi-staining-protocol",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-14T18:47:50Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Application or Specimen Type",
      "unit": "",
      "description": "Experimental context and biological sample preparation"
    },
    {
      "name": "Recommended Working Concentration",
      "unit": "µg/mL",
      "description": "Standard working dye concentration for optimal signal-to-noise ratio"
    },
    {
      "name": "Standard Incubation Time",
      "unit": "minutes",
      "description": "Incubation duration at room temperature protected from light"
    },
    {
      "name": "Equilibration and Wash Buffer",
      "unit": "",
      "description": "Buffer composition used for dilution and post-stain rinses"
    },
    {
      "name": "Excitation Wavelength Peak",
      "unit": "nm",
      "description": "Nominal excitation wavelength for instrument filter configuration"
    },
    {
      "name": "Emission Wavelength Peak",
      "unit": "nm",
      "description": "Nominal peak emission wavelength when complexed with nucleic acid target"
    },
    {
      "name": "Primary Source Organization",
      "unit": "",
      "description": "Authoritative manufacturer technical bulletin or protocol source"
    }
  ],
  "rows": [
    {
      "Application or Specimen Type": "Fixed Adherent Mammalian Cells (Immunofluorescence)",
      "Recommended Working Concentration": 1,
      "Standard Incubation Time": 5,
      "Equilibration and Wash Buffer": "1X Phosphate-Buffered Saline (PBS, pH 7.4)",
      "Excitation Wavelength Peak": 358,
      "Emission Wavelength Peak": 461,
      "Primary Source Organization": "Thermo Fisher Scientific MP 01306"
    },
    {
      "Application or Specimen Type": "Suspension Cells for Flow Cytometry (DNA Content)",
      "Recommended Working Concentration": 1.05,
      "Standard Incubation Time": 15,
      "Equilibration and Wash Buffer": "100 mM Tris pH 7.4, 150 mM NaCl, 1 mM CaCl2, 0.5 mM MgCl2, 0.1% NP-40",
      "Excitation Wavelength Peak": 358,
      "Emission Wavelength Peak": 461,
      "Primary Source Organization": "Thermo Fisher Scientific MP 01306"
    },
    {
      "Application or Specimen Type": "Metaphase Chromosome Spread (FISH Counterstain)",
      "Recommended Working Concentration": 0.01,
      "Standard Incubation Time": 30,
      "Equilibration and Wash Buffer": "1X Phosphate-Buffered Saline (PBS) with dH2O pre-rinse",
      "Excitation Wavelength Peak": 358,
      "Emission Wavelength Peak": 461,
      "Primary Source Organization": "Thermo Fisher Scientific MP 01306"
    },
    {
      "Application or Specimen Type": "Fixed Cells (CST Recommended Protocol)",
      "Recommended Working Concentration": 0.5,
      "Standard Incubation Time": 5,
      "Equilibration and Wash Buffer": "1X Phosphate-Buffered Saline (PBS)",
      "Excitation Wavelength Peak": 358,
      "Emission Wavelength Peak": 461,
      "Primary Source Organization": "Cell Signaling Technology #4083"
    },
    {
      "Application or Specimen Type": "Bacterial Suspension (E. coli and Mixed Culture)",
      "Recommended Working Concentration": 10,
      "Standard Incubation Time": 30,
      "Equilibration and Wash Buffer": "1X PBS or 150 mM NaCl solution",
      "Excitation Wavelength Peak": 352,
      "Emission Wavelength Peak": 464,
      "Primary Source Organization": "Biotium PSF006"
    },
    {
      "Application or Specimen Type": "Yeast Culture (Saccharomyces cerevisiae)",
      "Recommended Working Concentration": 10,
      "Standard Incubation Time": 30,
      "Equilibration and Wash Buffer": "1X Phosphate-Buffered Saline (PBS)",
      "Excitation Wavelength Peak": 352,
      "Emission Wavelength Peak": 464,
      "Primary Source Organization": "Biotium PSF006"
    }
  ],
  "sources": [
    {
      "title": "Thermo Fisher Scientific Molecular Probes DAPI Product Manual MP 01306",
      "url": "https://documents.thermofisher.com/TFS-Assets/LSG/manuals/mp01306.pdf",
      "accessed": "2026-11-14"
    },
    {
      "title": "Cell Signaling Technology DAPI Product Datasheet #4083",
      "url": "https://www.cellsignal.com/products/cellular-dyes/dapi/4083",
      "accessed": "2026-11-14"
    },
    {
      "title": "Biotium DAPI Product Information Sheet PSF006",
      "url": "https://biotium.com/wp-content/uploads/2016/12/PI-DAPI.pdf",
      "accessed": "2026-11-14"
    }
  ]
}