  "name": "USP Dissolution Apparatus Engineering and Operating Specifications",
  "description": "Comparison of mechanical operating parameters, standard vessel volumes, agitation mechanisms, and compendial tolerances across USP dissolution apparatus types 1 through 7, compiled from USP General Chapter 711, ASTM E2503-20, and FDA mechanical calibration guidance.",
  "page": "https://casrai.org/guides/dissolution-tester",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-15T08:33:24Z",
  "temporalCoverage": "2020/2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Apparatus Designation",
      "unit": "",
      "description": "Compendial apparatus type under USP 711 and 724"
    },
    {
      "name": "Agitation Mechanism",
      "unit": "",
      "description": "Mechanical drive and stirring geometry"
    },
    {
      "name": "Agitation Speed or Rate",
      "unit": "",
      "description": "Standard operating agitation range"
    },
    {
      "name": "Agitation Tolerance",
      "unit": "",
      "description": "Compendial or enhanced mechanical speed tolerance"
    },
    {
      "name": "Standard Vessel Volume",
      "unit": "mL",
      "description": "Standard vessel capacity for routine testing"
    },
    {
      "name": "Primary Dosage Form Application",
      "unit": "",
      "description": "Target pharmaceutical delivery systems"
    }
  ],
  "rows": [
    {
      "Apparatus Designation": "USP Apparatus 1 (Rotating Basket)",
      "Agitation Mechanism": "Rotating 40-mesh stainless steel basket cylinder",
      "Agitation Speed or Rate": "50 to 100 rpm",
      "Agitation Tolerance": "± 4% (USP) or ± 1-2 rpm (ASTM)",
      "Standard Vessel Volume": 1000,
      "Primary Dosage Form Application": "Capsules, floating tablets, beads, delayed-release dosage forms"
    },
    {
      "Apparatus Designation": "USP Apparatus 2 (Paddle)",
      "Agitation Mechanism": "Rotating two-bladed paddle assembly",
      "Agitation Speed or Rate": "50 to 75 rpm",
      "Agitation Tolerance": "± 4% (USP) or ± 1-2 rpm (ASTM)",
      "Standard Vessel Volume": 1000,
      "Primary Dosage Form Application": "Standard immediate-release tablets, disintegrating tablets, suspensions"
    },
    {
      "Apparatus Designation": "USP Apparatus 3 (Reciprocating Cylinder)",
      "Agitation Mechanism": "Vertical reciprocating inner cylinder with mesh screens",
      "Agitation Speed or Rate": "5 to 35 dpm",
      "Agitation Tolerance": "± 5%",
      "Standard Vessel Volume": 250,
      "Primary Dosage Form Application": "Extended-release beads, pellets, chewable tablets, multi-media profiles"
    },
    {
      "Apparatus Designation": "USP Apparatus 4 (Flow-Through Cell)",
      "Agitation Mechanism": "Continuous unidirectional fluid perfusion via piston pump",
      "Agitation Speed or Rate": "4, 8, or 16 mL/min",
      "Agitation Tolerance": "± 5%",
      "Standard Vessel Volume": 1000,
      "Primary Dosage Form Application": "Poorly soluble APIs, microparticles, liposomes, implants, stents"
    },
    {
      "Apparatus Designation": "USP Apparatus 5 (Paddle Over Disk)",
      "Agitation Mechanism": "Paddle rotation above fixed transdermal mesh disk assembly",
      "Agitation Speed or Rate": "25 to 50 rpm",
      "Agitation Tolerance": "± 4%",
      "Standard Vessel Volume": 900,
      "Primary Dosage Form Application": "Transdermal delivery systems and patches"
    },
    {
      "Apparatus Designation": "USP Apparatus 6 (Rotating Cylinder)",
      "Agitation Mechanism": "Rotating stainless steel cylinder with adhered transdermal patch",
      "Agitation Speed or Rate": "25 to 50 rpm",
      "Agitation Tolerance": "± 4%",
      "Standard Vessel Volume": 1000,
      "Primary Dosage Form Application": "Transdermal patches requiring continuous surface flow"
    },
    {
      "Apparatus Designation": "USP Apparatus 7 (Reciprocating Holder)",
      "Agitation Mechanism": "Vertical reciprocating motor shaft holding sample mounts",
      "Agitation Speed or Rate": "20 to 50 dpm",
      "Agitation Tolerance": "± 5%",
      "Standard Vessel Volume": 50,
      "Primary Dosage Form Application": "Low-dose transdermal patches, osmotic oral tablets, stent coatings"
    }
  ],
  "sources": [
    {
      "title": "USP General Chapter 711 Dissolution",
      "url": "https://www.uspnf.com",
      "accessed": "2026-11-15"
    },
    {
      "title": "ASTM E2503-20 Standard Practice for Qualification of Basket and Paddle Dissolution Apparatus",
      "url": "https://www.astm.org/e2503-20.html",
      "accessed": "2026-11-15"
    },
    {
      "title": "FDA Guidance for Industry: The Use of Mechanical Calibration of Dissolution Apparatus 1 and 2",
      "url": "https://www.fda.gov/media/74549/download",
      "accessed": "2026-11-15"
    }
  ]
}