  "name": "Dot blot symptoms, probable causes and corrective actions in HRP chemiluminescent blots",
  "description": "A symptom-to-cause-to-fix table compiled by CASRAI from two manufacturer technical tips (Thermo Fisher) and two western blot troubleshooting guides (Proteintech, UC Davis Gomes Lab). Only causes tied to blocking, antibody incubation, washing and detection are included, since those steps are shared by a dot blot; gel and transfer causes are excluded. Each row names its source.",
  "page": "https://casrai.org/guides/dot-blot-troubleshooting",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T04:23:10Z",
  "temporalCoverage": "2009/2024",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Symptom",
      "unit": "",
      "description": ""
    },
    {
      "name": "Probable cause",
      "unit": "",
      "description": ""
    },
    {
      "name": "Corrective action",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Symptom": "No signal or very weak signal",
      "Probable cause": "Antibody may have lost activity, be expired, or be stored incorrectly; secondary antibody of the wrong host species; too little primary incubation time",
      "Corrective action": "Test activity with a dot blot dilution series; confirm host species and Ig type of the primary; use fresh antibody; follow storage advice; extend primary incubation (Gomes: overnight at 4 °C)",
      "Source": "Proteintech; UC Davis Gomes Lab"
    },
    {
      "Symptom": "No signal with HRP detection",
      "Probable cause": "Sodium azide in buffers inhibits HRP activity",
      "Corrective action": "Use azide-free buffers",
      "Source": "Proteintech; UC Davis Gomes Lab"
    },
    {
      "Symptom": "Faint signal",
      "Probable cause": "Low-affinity antibody; weak or old ECL reagent; non-fat dry milk masking antigen",
      "Corrective action": "Increase antibody concentration 2 to 4 fold over the recommended start; use fresh ECL reagents; reduce milk or substitute 3% BSA",
      "Source": "UC Davis Gomes Lab"
    },
    {
      "Symptom": "High background across the membrane",
      "Probable cause": "Insufficient blocking or washing; secondary antibody binding nonspecifically; primary nonspecific binding",
      "Corrective action": "Adjust blocking (Gomes starts with 5% dry milk with 0.1 to 0.5% Tween 20, or 3% BSA); increase number of washes and Tween 20 (0.1 to 0.5%); run a secondary-only control; raise NaCl to 0.15 to 0.5 M in antibody and wash buffers",
      "Source": "UC Davis Gomes Lab"
    },
    {
      "Symptom": "Dark spots or dots",
      "Probable cause": "Antibodies binding to the blocking reagent; HRP aggregates",
      "Corrective action": "Change the blocking reagent; filter the blocking buffer; wash excess detection reagent off before exposure; filter the HRP conjugate",
      "Source": "Proteintech; UC Davis Gomes Lab"
    },
    {
      "Symptom": "Patchy, uneven spots across the blot",
      "Probable cause": "Contaminated reagents, too little solution during incubation or washing, uneven agitation, contaminated equipment, aggregated HRP, long exposure",
      "Corrective action": "Make fresh reagents; fully immerse the membrane; agitate uniformly on a rocker or shaker; wash equipment; filter the conjugate; shorten exposure",
      "Source": "UC Davis Gomes Lab"
    },
    {
      "Symptom": "White spots or reverse image",
      "Probable cause": "Excess antibody or protein consumes substrate locally, leaving no light after the reaction ends",
      "Corrective action": "Reduce antibody or protein concentration; test a dilution series",
      "Source": "UC Davis Gomes Lab; Thermo Fisher Tech Tip 22"
    },
    {
      "Symptom": "Signal with a biotin-avidin or streptavidin system",
      "Probable cause": "Non-fat dry milk contains endogenous biotin",
      "Corrective action": "Substitute 3% BSA for milk",
      "Source": "UC Davis Gomes Lab"
    },
    {
      "Symptom": "Inconsistent results from one run to the next",
      "Probable cause": "Antibody dilution outside the optimum, since less antibody can sometimes give more signal",
      "Corrective action": "Optimize primary and secondary dilutions on a dot blot within the supplier's recommended range",
      "Source": "Thermo Fisher TR0024; Thermo Fisher Tech Tip 22"
    }
  ],
  "sources": [
    {
      "title": "Thermo Fisher Tech Tip 22: Determine source of nonspecific background signal",
      "url": "https://assets.thermofisher.com/TFS-Assets/LSG/Application-Notes/TR0022-Determine-background-cause.pdf",
      "accessed": "2026-11-21"
    },
    {
      "title": "Thermo Fisher TR0024: Dot blotting, the key to optimizing your western blot",
      "url": "https://documents.thermofisher.com/TFS-Assets/LSG/Application-Notes/TR0024-Optimize-Ab-dilutions.pdf",
      "accessed": "2026-11-21"
    },
    {
      "title": "Proteintech: Western Blot Troubleshooting, Weak/No Signal and Other",
      "url": "https://www.ptglab.com/support/western-blot-protocol/western-blot-troubleshooting-weak-no-signal-other/",
      "accessed": "2026-11-21"
    },
    {
      "title": "Gomes Lab, UC Davis: Western Blotting Tips and Troubleshooting Guide",
      "url": "https://lab.vanderbilt.edu/wp-content/uploads/sites/21/public_files/083013_WESTERN%20BLOTTING%20TIPS%20AND%20TROUBLESHOOTING%20GUIDE.pdf",
      "accessed": "2026-11-21"
    }
  ]
}