  "name": "Dust monitor measurement principles compared",
  "description": "Compares the three measurement principles used by dust monitors (filter and balance, tapered element oscillating microbalance, optical light scattering) by what each reports, its size selection and when the result is available. Compiled from NIOSH Method 0600, the NIOSH CPDM article and the US EPA Air Sensor Toolbox sensor table.",
  "page": "https://casrai.org/guides/dust-monitor",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-19T15:01:05Z",
  "temporalCoverage": "2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Principle",
      "unit": "",
      "description": ""
    },
    {
      "name": "Quantity reported",
      "unit": "",
      "description": ""
    },
    {
      "name": "Size selection or range",
      "unit": "",
      "description": ""
    },
    {
      "name": "When the result is available",
      "unit": "",
      "description": ""
    },
    {
      "name": "Example named in source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Principle": "Filter and balance (gravimetric)",
      "Quantity reported": "Mass of respirable dust on a filter, divided by air volume",
      "Size selection or range": "Cyclone with 4 µm median cut point; 5 µm pore PVC filter; 1.7, 2.2 or 2.5 L/min by cyclone",
      "When the result is available": "After the filter is equilibrated and weighed (0.001 mg balance)",
      "Example named in source": "NIOSH Method 0600, Issue 3"
    },
    {
      "Principle": "Tapered element oscillating microbalance (TEOM)",
      "Quantity reported": "Respirable coal dust mass exposure",
      "Size selection or range": "Respirable coal dust",
      "When the result is available": "Live display of previous 30 minutes and cumulative shift average",
      "Example named in source": "PDM3700 continuous personal dust monitor"
    },
    {
      "Principle": "Optical particle counter",
      "Quantity reported": "Particle counts in size ranges, in some units converted to PM2.5 and PM10 mass estimates",
      "Size selection or range": "Above 0.5 and above 2.5 µm (one unit); 0.38 to 17 µm (another)",
      "When the result is available": "Data as fast as 1 minute (counter) or 1 minute averages (OPC-N2)",
      "Example named in source": "Dylos DC1100; Alphasense OPC-N2"
    },
    {
      "Principle": "Volume-scattering sensor",
      "Quantity reported": "Estimated mass concentration from total scattered light",
      "Size selection or range": "May include particles larger than 2.5 µm",
      "When the result is available": "Data as fast as 1 second",
      "Example named in source": "Shinyei sensor"
    }
  ],
  "sources": [
    {
      "title": "NIOSH Manual of Analytical Methods, Method 0600: Particulates not otherwise regulated, respirable",
      "url": "https://www.cdc.gov/niosh/docs/2003-154/pdfs/0600.pdf",
      "accessed": "2026-11-19"
    },
    {
      "title": "NIOSH: Continuous personal dust monitor",
      "url": "https://www.cdc.gov/niosh/bulletin/2017/dust-monitor.html",
      "accessed": "2026-11-19"
    },
    {
      "title": "US EPA: Evaluation of Emerging Air Sensor Performance",
      "url": "https://www.epa.gov/air-sensor-toolbox/evaluation-emerging-air-sensor-performance",
      "accessed": "2026-11-19"
    }
  ]
}