  "name": "Ellman's assay reference values: extinction coefficients, buffer, stability and sensitivity",
  "description": "Compiled from the Thermo Fisher Pierce Ellman's Reagent product page, BMG LABTECH Application Note 327 and the Riener et al. 2002 abstract. Each row gives one working parameter of the DTNB thiol assay, the condition it applies to and the source it comes from, so values that differ between sources can be compared in one place.",
  "page": "https://casrai.org/guides/ellmans-reagent",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-18T10:25:00Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Parameter",
      "unit": "",
      "description": ""
    },
    {
      "name": "Value",
      "unit": "",
      "description": "Numeric where a single number is stated; text otherwise"
    },
    {
      "name": "Unit",
      "unit": "",
      "description": "Unit of the value, where applicable"
    },
    {
      "name": "Condition",
      "unit": "",
      "description": "Conditions the source attaches to the value"
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Parameter": "TNB extinction coefficient at 412 nm",
      "Value": 14150,
      "Unit": "M-1 cm-1",
      "Condition": "Value given by the vendor for TNB at 412 nm",
      "Source": "Thermo Fisher"
    },
    {
      "Parameter": "TNB extinction coefficient at 412 nm (Ellman's original)",
      "Value": 13600,
      "Unit": "M-1 cm-1",
      "Condition": "pH 8.0; still often cited",
      "Source": "Thermo Fisher; BMG LABTECH AN327"
    },
    {
      "Parameter": "Detection wavelength",
      "Value": 412,
      "Unit": "nm",
      "Condition": "TNB absorbance maximum used for endpoint reading",
      "Source": "BMG LABTECH AN327; Thermo Fisher"
    },
    {
      "Parameter": "pH range with unaffected TNB absorbance",
      "Value": "7.6 to 8.6",
      "Unit": "pH",
      "Condition": "Standard reaction pH is 8.0",
      "Source": "Thermo Fisher"
    },
    {
      "Parameter": "Reaction buffer",
      "Value": "0.1 M sodium phosphate, 1 mM EDTA",
      "Unit": "pH 8.0",
      "Condition": "Tris pH 8.0 also acceptable per Thermo Fisher",
      "Source": "Thermo Fisher; BMG LABTECH AN327"
    },
    {
      "Parameter": "Working solution stability in phosphate or Tris",
      "Value": "About 1 week",
      "Unit": "",
      "Condition": "pH 8.0",
      "Source": "Thermo Fisher"
    },
    {
      "Parameter": "Working solution stability in PBS",
      "Value": "1 to 2 days",
      "Unit": "",
      "Condition": "pH 7.2",
      "Source": "Thermo Fisher"
    },
    {
      "Parameter": "Cysteine standard range in plate protocol",
      "Value": "1.6 mM stock, two-fold dilutions to 0.1 mM",
      "Unit": "",
      "Condition": "Cysteine hydrochloride monohydrate, MW 175.6",
      "Source": "BMG LABTECH AN327"
    },
    {
      "Parameter": "Detection limit with cystamine mediator",
      "Value": 0.2,
      "Unit": "nmol/mL",
      "Condition": "5-minute assay for protein sulfhydryls",
      "Source": "Riener et al. 2002"
    },
    {
      "Parameter": "Determination limit with cystamine mediator",
      "Value": 0.6,
      "Unit": "nmol/mL",
      "Condition": "5-minute assay for protein sulfhydryls",
      "Source": "Riener et al. 2002"
    }
  ],
  "sources": [
    {
      "title": "Thermo Fisher Scientific, Pierce Ellman's Reagent (22582)",
      "url": "https://www.thermofisher.com/order/catalog/product/22582",
      "accessed": "2026-11-14"
    },
    {
      "title": "BMG LABTECH Application Note 327",
      "url": "https://www.bmglabtech.com/fileadmin/06_Support/Download_Documents/App_Notes/AN327.pdf",
      "accessed": "2026-11-14"
    },
    {
      "title": "Riener et al. 2002, Anal Bioanal Chem 373:266-276 (abstract)",
      "url": "https://pubmed.ncbi.nlm.nih.gov/12110978/",
      "accessed": "2026-11-14"
    }
  ]
}