  "name": "Enterolert product pages and the NEMI method record compared: reference method, incubation, counting range and storage",
  "description": "Cross-walk of IDEXX's Enterolert and Enterolert-E product pages with the two NEMI (EPA/USGS) summaries of the Enterolert Test Kit Procedure, compiled on 2026-11-21 from the four pages listed as sources. Cells not stated by a source are marked as such.",
  "page": "https://casrai.org/guides/enterolert",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T14:23:58Z",
  "temporalCoverage": "2026-11-21",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Record",
      "unit": "",
      "description": ""
    },
    {
      "name": "Reference method named by the source",
      "unit": "",
      "description": ""
    },
    {
      "name": "Incubation temperature",
      "unit": "°C",
      "description": ""
    },
    {
      "name": "Incubation time",
      "unit": "hours",
      "description": ""
    },
    {
      "name": "Counting range, 51-well or standard Quanti-Tray",
      "unit": "MPN/100 mL",
      "description": ""
    },
    {
      "name": "Counting range, 97-well Quanti-Tray/2000",
      "unit": "MPN/100 mL",
      "description": ""
    },
    {
      "name": "Storage",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Record": "IDEXX Enterolert page",
      "Reference method named by the source": "ASTM D6503-99; US EPA approved; Standard Methods",
      "Incubation temperature": 41,
      "Incubation time": 24,
      "Counting range, 51-well or standard Quanti-Tray": "1-200",
      "Counting range, 97-well Quanti-Tray/2000": "1-2,419",
      "Storage": "2-30 °C, away from light; shelf life up to 12 months"
    },
    {
      "Record": "IDEXX Enterolert-E page",
      "Reference method named by the source": "Correlates with ISO 7899-1",
      "Incubation temperature": 41,
      "Incubation time": 24,
      "Counting range, 51-well or standard Quanti-Tray": "1-200",
      "Counting range, 97-well Quanti-Tray/2000": "1-2,419",
      "Storage": "2-30 °C; shelf life up to 12 months"
    },
    {
      "Record": "NEMI method summary 5441 (ASTM D6503, approved Dec. 10, 1999)",
      "Reference method named by the source": "ASTM D6503; presence/absence or 5-, 10-, 15-tube MPN or Quanti-Tray",
      "Incubation temperature": 41,
      "Incubation time": 24,
      "Counting range, 51-well or standard Quanti-Tray": "Not stated in this record",
      "Counting range, 97-well Quanti-Tray/2000": "Not stated in this record",
      "Storage": "Not stated in this record"
    },
    {
      "Record": "NEMI method summary 5590 (Enterolert Test Kit Procedure, rev. 1999)",
      "Reference method named by the source": "MPN method; Standard Methods 20th ed.; 40 CFR 136; Fed. Reg. 66(169) 2001",
      "Incubation temperature": 41,
      "Incubation time": 24,
      "Counting range, 51-well or standard Quanti-Tray": "1-200",
      "Counting range, 97-well Quanti-Tray/2000": "1-2,419",
      "Storage": "Not stated in this record"
    }
  ],
  "sources": [
    {
      "title": "IDEXX Enterolert",
      "url": "https://www.idexx.com/en/water/water-products-services/enterolert/",
      "accessed": "2026-11-21"
    },
    {
      "title": "IDEXX Enterolert-E",
      "url": "https://www.idexx.com/en/water/water-products-services/enterolert-e/",
      "accessed": "2026-11-21"
    },
    {
      "title": "NEMI: ASTM D6503 Enterococci in Water",
      "url": "https://www.nemi.gov/methods/method_summary/5441/",
      "accessed": "2026-11-21"
    },
    {
      "title": "NEMI: IDEXX Enterolert Test Kit Procedure",
      "url": "https://www.nemi.gov/methods/method_summary/5590/",
      "accessed": "2026-11-21"
    }
  ]
}