  "name": "EPA Method 6010D Analyte Performance and Spectral Interferences",
  "description": "Compilation of analytical emission wavelengths, primary interelement spectral interferences, analyte concentration equivalents at 100 mg/L interferent levels, and interlaboratory aqueous precision and accuracy data compiled from EPA SW-846 Method 6010D Tables 1 and 4.",
  "page": "https://casrai.org/guides/epa-6010d",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-17T00:58:22Z",
  "temporalCoverage": "2018/2026",
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Analyte",
      "unit": "",
      "description": "Target elemental analyte"
    },
    {
      "name": "Wavelength",
      "unit": "nm",
      "description": "Primary analytical emission wavelength"
    },
    {
      "name": "Primary Interferent",
      "unit": "",
      "description": "Element causing notable spectral overlap"
    },
    {
      "name": "Interferent Equivalent",
      "unit": "mg/L",
      "description": "Analyte concentration equivalent from 100 mg/L interferent"
    },
    {
      "name": "Aqueous Mean Conc",
      "unit": "mg/L",
      "description": "Mean concentration in interlaboratory aqueous test solution"
    },
    {
      "name": "Relative Standard Deviation",
      "unit": "%",
      "description": "Interlaboratory relative standard deviation (RSD)"
    },
    {
      "name": "Method Accuracy",
      "unit": "%",
      "description": "Mean analytical accuracy percentage of nominal concentration"
    }
  ],
  "rows": [
    {
      "Analyte": "Aluminum",
      "Wavelength": 308.215,
      "Primary Interferent": "Vanadium",
      "Interferent Equivalent": 1.4,
      "Aqueous Mean Conc": 14.8,
      "Relative Standard Deviation": 6.3,
      "Method Accuracy": 100
    },
    {
      "Analyte": "Antimony",
      "Wavelength": 206.833,
      "Primary Interferent": "Chromium",
      "Interferent Equivalent": 2.9,
      "Aqueous Mean Conc": 15.1,
      "Relative Standard Deviation": 7.7,
      "Method Accuracy": 102
    },
    {
      "Analyte": "Arsenic",
      "Wavelength": 193.696,
      "Primary Interferent": "Aluminum",
      "Interferent Equivalent": 1.3,
      "Aqueous Mean Conc": 14.7,
      "Relative Standard Deviation": 6.4,
      "Method Accuracy": 99
    },
    {
      "Analyte": "Barium",
      "Wavelength": 455.403,
      "Primary Interferent": "None",
      "Interferent Equivalent": 0,
      "Aqueous Mean Conc": 3.66,
      "Relative Standard Deviation": 3.1,
      "Method Accuracy": 99
    },
    {
      "Analyte": "Beryllium",
      "Wavelength": 313.042,
      "Primary Interferent": "Vanadium",
      "Interferent Equivalent": 0.05,
      "Aqueous Mean Conc": 3.78,
      "Relative Standard Deviation": 5.8,
      "Method Accuracy": 102
    },
    {
      "Analyte": "Cadmium",
      "Wavelength": 226.502,
      "Primary Interferent": "Iron",
      "Interferent Equivalent": 0.03,
      "Aqueous Mean Conc": 3.61,
      "Relative Standard Deviation": 7,
      "Method Accuracy": 97
    },
    {
      "Analyte": "Chromium",
      "Wavelength": 267.716,
      "Primary Interferent": "Manganese",
      "Interferent Equivalent": 0.04,
      "Aqueous Mean Conc": 3.75,
      "Relative Standard Deviation": 8.2,
      "Method Accuracy": 101
    },
    {
      "Analyte": "Copper",
      "Wavelength": 324.754,
      "Primary Interferent": "Titanium",
      "Interferent Equivalent": 0.05,
      "Aqueous Mean Conc": 3.58,
      "Relative Standard Deviation": 5.6,
      "Method Accuracy": 97
    },
    {
      "Analyte": "Iron",
      "Wavelength": 259.94,
      "Primary Interferent": "Manganese",
      "Interferent Equivalent": 0.12,
      "Aqueous Mean Conc": 14.8,
      "Relative Standard Deviation": 5.9,
      "Method Accuracy": 100
    },
    {
      "Analyte": "Lead",
      "Wavelength": 220.353,
      "Primary Interferent": "Aluminum",
      "Interferent Equivalent": 0.17,
      "Aqueous Mean Conc": 14.4,
      "Relative Standard Deviation": 5.9,
      "Method Accuracy": 97
    },
    {
      "Analyte": "Manganese",
      "Wavelength": 257.61,
      "Primary Interferent": "Chromium",
      "Interferent Equivalent": 0.01,
      "Aqueous Mean Conc": 3.7,
      "Relative Standard Deviation": 4.3,
      "Method Accuracy": 100
    },
    {
      "Analyte": "Nickel",
      "Wavelength": 231.604,
      "Primary Interferent": "None",
      "Interferent Equivalent": 0,
      "Aqueous Mean Conc": 3.7,
      "Relative Standard Deviation": 5.7,
      "Method Accuracy": 100
    },
    {
      "Analyte": "Selenium",
      "Wavelength": 196.026,
      "Primary Interferent": "Aluminum",
      "Interferent Equivalent": 0.23,
      "Aqueous Mean Conc": 15.3,
      "Relative Standard Deviation": 7.5,
      "Method Accuracy": 104
    },
    {
      "Analyte": "Vanadium",
      "Wavelength": 292.402,
      "Primary Interferent": "Chromium",
      "Interferent Equivalent": 0.05,
      "Aqueous Mean Conc": 3.51,
      "Relative Standard Deviation": 6.6,
      "Method Accuracy": 95
    },
    {
      "Analyte": "Zinc",
      "Wavelength": 213.856,
      "Primary Interferent": "Nickel",
      "Interferent Equivalent": 0.29,
      "Aqueous Mean Conc": 3.57,
      "Relative Standard Deviation": 8.3,
      "Method Accuracy": 96
    }
  ],
  "sources": [
    {
      "title": "SW-846 Test Method 6010D: Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES)",
      "url": "https://www.epa.gov/hw-sw846/sw-846-test-method-6010d-inductively-coupled-plasma-optical-emission-spectrometry-icp-oes",
      "accessed": "2026-11-16"
    }
  ]
}