  "name": "EPA Method 1633A holding times by sample type and storage condition",
  "description": "Holding times in days from Section 8.5 and the 8.5.6 summary of Method 1633A (December 2024), compiled by sample type and storage temperature, with the shorter limit that applies when specific analytes are targets.",
  "page": "https://casrai.org/guides/epa-method-1633a",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-29T12:50:00Z",
  "temporalCoverage": "2024",
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Sample type",
      "unit": "",
      "description": ""
    },
    {
      "name": "Storage condition",
      "unit": "",
      "description": ""
    },
    {
      "name": "Standard holding time",
      "unit": "days",
      "description": ""
    },
    {
      "name": "Shorter holding time",
      "unit": "days",
      "description": ""
    },
    {
      "name": "Condition for the shorter limit",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Sample type": "Aqueous, including leachates",
      "Storage condition": "0 to 6 °C, dark",
      "Standard holding time": 28,
      "Shorter holding time": 7,
      "Condition for the shorter limit": "NMeFOSE, NEtFOSE, NMeFOSAA or NEtFOSAA is a target analyte (or store at or below -20 °C)"
    },
    {
      "Sample type": "Aqueous, including leachates",
      "Storage condition": "At or below -20 °C",
      "Standard holding time": 90,
      "Shorter holding time": "none",
      "Condition for the shorter limit": "Not applicable"
    },
    {
      "Sample type": "Soil and sediment",
      "Storage condition": "Either storage temperature",
      "Standard holding time": 90,
      "Shorter holding time": 3,
      "Condition for the shorter limit": "NFDHA is an analyte of interest"
    },
    {
      "Sample type": "Biosolids",
      "Storage condition": "Frozen preferred; EPA recommends freezing if stored more than a few days",
      "Standard holding time": 90,
      "Shorter holding time": "none",
      "Condition for the shorter limit": "Not applicable"
    },
    {
      "Sample type": "Tissue",
      "Storage condition": "0 to 6 °C",
      "Standard holding time": 90,
      "Shorter holding time": 3,
      "Condition for the shorter limit": "NFDHA is an analyte of interest"
    },
    {
      "Sample type": "Extracts",
      "Storage condition": "0 to 6 °C or at or below -20 °C, dark",
      "Standard holding time": 90,
      "Shorter holding time": "none",
      "Condition for the shorter limit": "Issue noted for 11Cl-PF3OUdS and 9Cl-PF3ONS after 28 days"
    }
  ],
  "sources": [
    {
      "title": "EPA Method 1633A, EPA 820-R-24-007",
      "url": "https://www.epa.gov/system/files/documents/2024-12/method-1633a-december-5-2024-508-compliant.pdf",
      "accessed": "2026-11-21"
    }
  ]
}