  "name": "Method 21 instrument performance tests: requirement, criterion and timing",
  "description": "Compares the three instrument performance evaluations in section 8.1 of EPA Method 21 (response factor, calibration precision, response time) with the daily calibration drift check in 40 CFR 60.485a, giving the pass criterion, number of measurements and when each is required. Compiled from the EPA method text (3 August 2017) and the Cornell LII copy of 40 CFR 60.485a.",
  "page": "https://casrai.org/guides/epa-method-21",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-20T15:55:04Z",
  "temporalCoverage": null,
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Check",
      "unit": "",
      "description": ""
    },
    {
      "name": "Pass criterion",
      "unit": "",
      "description": ""
    },
    {
      "name": "Measurements",
      "unit": "",
      "description": "Number of repeated measurements or trials"
    },
    {
      "name": "When required",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Check": "Response factor (8.1.1)",
      "Pass criterion": "Below 10 for each compound unless the regulation says otherwise",
      "Measurements": "3 alternating calibration gas and zero gas measurements",
      "When required": "Before placing the analyzer in service; not repeated at later intervals; published factors may be referenced",
      "Source": "EPA Method 21 section 8.1.1"
    },
    {
      "Check": "Calibration precision (8.1.2)",
      "Pass criterion": "At or below 10 percent of the calibration gas value",
      "Measurements": "3 alternating zero gas and calibration gas measurements",
      "When required": "Before service, then at 3-month intervals or at next use, whichever is later",
      "Source": "EPA Method 21 section 8.1.2"
    },
    {
      "Check": "Response time (8.1.3)",
      "Pass criterion": "At or below 30 seconds to reach 90 percent of final reading",
      "Measurements": "3 trials, with pump, probe and filter in place",
      "When required": "Before service; again after a pump or flow change that alters response time",
      "Source": "EPA Method 21 section 8.1.3"
    },
    {
      "Check": "Calibration drift assessment (60.485a(b)(2))",
      "Pass criterion": "Negative drift over 10 percent requires re-monitoring equipment checked since last calibration",
      "Measurements": "Check with the same calibration gases used before the survey",
      "When required": "At least at the end of each monitoring day",
      "Source": "40 CFR 60.485a"
    }
  ],
  "sources": [
    {
      "title": "EPA Method 21 Determination of Volatile Organic Compound Leaks (PDF, 8/3/2017)",
      "url": "https://www.epa.gov/sites/default/files/2017-08/documents/method_21.pdf",
      "accessed": "2026-11-20"
    },
    {
      "title": "40 CFR 60.485a Test methods and procedures (Cornell LII)",
      "url": "https://www.law.cornell.edu/cfr/text/40/60.485a",
      "accessed": "2026-11-20"
    }
  ]
}