  "name": "Fiber photometry: artifacts, controls and inference limits",
  "description": "Original qualitative cross-source compilation linking distinct photometry measurement problems to the evidence used to interpret them and the limit of that evidence. Compiled from pyPhotometry, TDT documentation and the Keevers and Legaria studies; source names are given in each row. This is a reporting reference, not measured experimental data or a universal correction protocol.",
  "page": "https://casrai.org/guides/fiber-photometry",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-14T10:20:19Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Measurement problem",
      "unit": "",
      "description": ""
    },
    {
      "name": "Relevant evidence or control",
      "unit": "",
      "description": ""
    },
    {
      "name": "Limit of the evidence",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Measurement problem": "Background illumination",
      "Relevant evidence or control": "LED-off baseline subtraction in pulsed acquisition",
      "Limit of the evidence": "Does not remove patch-cord autofluorescence or restore a saturated raw signal",
      "Source": "pyPhotometry official documentation"
    },
    {
      "Measurement problem": "Autofluorescence and photobleaching",
      "Relevant evidence or control": "Optical-component and tissue contributions; inspect background and fluorescence decay",
      "Limit of the evidence": "Recorded light includes sources other than target-dependent sensor fluorescence",
      "Source": "TDT Overview"
    },
    {
      "Measurement problem": "Sensor-dependent reference contamination",
      "Relevant evidence or control": "Confirm that reference excitation matches the sensor isosbestic behavior",
      "Limit of the evidence": "405 nm can produce negative GCaMP6/newer reference deflections that inflate the corrected response on subtraction",
      "Source": "TDT Run-time"
    },
    {
      "Measurement problem": "Fitted-reference bias",
      "Relevant evidence or control": "Compare the fitted reference and regression method; IRLS outperformed OLS in the study tests",
      "Limit of the evidence": "A successful regression does not establish biological independence of the reference channel",
      "Source": "Keevers and Jean-Richard-dit-Bressel (2025)"
    },
    {
      "Measurement problem": "Equating regional calcium fluorescence with firing",
      "Relevant evidence or control": "Compare to independent measurements; striatal study included electrophysiology and imaging",
      "Limit of the evidence": "Mouse striatal findings support nonsomatic contributions; they are not a universal spike-rate conversion",
      "Source": "Legaria et al. (2022)"
    }
  ],
  "sources": [
    {
      "title": "pyPhotometry official documentation",
      "url": "https://pyphotometry.readthedocs.io/en/latest/",
      "accessed": "2026-11-14"
    },
    {
      "title": "TDT Fiber Photometry User Guide: Overview",
      "url": "https://www.tdt.com/docs/fiber-photometry-user-guide/overview/",
      "accessed": "2026-11-14"
    },
    {
      "title": "TDT Fiber Photometry User Guide: Run-time",
      "url": "https://www.tdt.com/docs/fiber-photometry-user-guide/run-time/",
      "accessed": "2026-11-14"
    },
    {
      "title": "Keevers and Jean-Richard-dit-Bressel (2025): Obtaining artifact-corrected signals",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC11957252/",
      "accessed": "2026-11-14"
    },
    {
      "title": "Legaria et al. (2022): Striatal photometry and nonsomatic calcium",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC10152879/",
      "accessed": "2026-11-14"
    }
  ]
}