  "name": "Fluorescent Calcium Indicator Specifications and Photochemical Properties",
  "description": "Comparative photochemical properties, calcium dissociation constants, spectral parameters, and instrumentation requirements for synthetic small-molecule fluorescent calcium indicators compiled from manufacturer technical data and peer-reviewed literature.",
  "page": "https://casrai.org/guides/fluo-4",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-19T21:15:53Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Indicator",
      "unit": "",
      "description": "Chemical name and derivative form of the fluorescent calcium indicator dye"
    },
    {
      "name": "Excitation Peak",
      "unit": "nm",
      "description": "Wavelength of maximum light absorption in physiological buffer"
    },
    {
      "name": "Emission Peak",
      "unit": "nm",
      "description": "Wavelength of maximum fluorescence emission"
    },
    {
      "name": "Dissociation Constant (Kd)",
      "unit": "nM",
      "description": "Equilibrium dissociation constant for calcium at 22 deg C, pH 7.2"
    },
    {
      "name": "Readout Modality",
      "unit": "",
      "description": "Single-wavelength intensity vs dual-wavelength ratiometric measurement"
    },
    {
      "name": "Fluorescence Increase Upon Ca2+ Saturation",
      "unit": "fold",
      "description": "Dynamic range ratio of Ca2+-saturated to Ca2+-free fluorescence intensity"
    },
    {
      "name": "Primary Excitation Light Source",
      "unit": "",
      "description": "Standard laboratory laser or light source matching indicator excitation"
    }
  ],
  "rows": [
    {
      "Indicator": "Fluo-4",
      "Excitation Peak": 494,
      "Emission Peak": 516,
      "Dissociation Constant (Kd)": 345,
      "Readout Modality": "Single-wavelength intensity",
      "Fluorescence Increase Upon Ca2+ Saturation": 100,
      "Primary Excitation Light Source": "488 nm Argon-ion or blue diode laser"
    },
    {
      "Indicator": "Fura-2",
      "Excitation Peak": 340,
      "Emission Peak": 510,
      "Dissociation Constant (Kd)": 145,
      "Readout Modality": "Dual-excitation ratiometric (340/380 nm)",
      "Fluorescence Increase Upon Ca2+ Saturation": 30,
      "Primary Excitation Light Source": "340 nm and 380 nm UV Xenon arc lamp or dual LEDs"
    },
    {
      "Indicator": "Fluo-3",
      "Excitation Peak": 506,
      "Emission Peak": 526,
      "Dissociation Constant (Kd)": 390,
      "Readout Modality": "Single-wavelength intensity",
      "Fluorescence Increase Upon Ca2+ Saturation": 60,
      "Primary Excitation Light Source": "488 nm or 514 nm Argon-ion laser"
    },
    {
      "Indicator": "Indo-1",
      "Excitation Peak": 338,
      "Emission Peak": 405,
      "Dissociation Constant (Kd)": 230,
      "Readout Modality": "Dual-emission ratiometric (405/485 nm)",
      "Fluorescence Increase Upon Ca2+ Saturation": 12,
      "Primary Excitation Light Source": "350 nm UV laser or mercury-arc lamp"
    },
    {
      "Indicator": "Cal-520",
      "Excitation Peak": 492,
      "Emission Peak": 514,
      "Dissociation Constant (Kd)": 320,
      "Readout Modality": "Single-wavelength intensity",
      "Fluorescence Increase Upon Ca2+ Saturation": 100,
      "Primary Excitation Light Source": "488 nm blue diode or Argon-ion laser"
    }
  ],
  "sources": [
    {
      "title": "Fluo-4 AM Product Specifications and Technical Handbook",
      "url": "https://www.thermofisher.com/order/catalog/product/F14201",
      "accessed": "2026-11-19"
    },
    {
      "title": "Fluo-4 AM Technical Data",
      "url": "https://www.aatbio.com/products/fluo-4-am",
      "accessed": "2026-11-19"
    },
    {
      "title": "Calcium imaging: a technique to monitor calcium dynamics in biological systems",
      "url": "https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10784449/",
      "accessed": "2026-11-19"
    }
  ]
}