  "name": "Fitted FCS quantities and the errors that corrupt them",
  "description": "Compiled from four peer-reviewed reviews of fluorescence correlation spectroscopy: for each fitted quantity (amplitude, decay time, brightness) what it reports, and for each recognised error source what it does to the result. Each row is stated in at least one of the listed reviews; the table pairs the statements so a reader can see what to check before trusting a number.",
  "page": "https://casrai.org/guides/fluorescence-correlation-spectroscopy",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-19T05:17:51Z",
  "temporalCoverage": "2011/2024",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Item",
      "unit": "",
      "description": ""
    },
    {
      "name": "Kind",
      "unit": "",
      "description": ""
    },
    {
      "name": "What it is or what it does",
      "unit": "",
      "description": ""
    },
    {
      "name": "Effect on the result",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Item": "Amplitude G(0)",
      "Kind": "Fitted parameter",
      "What it is or what it does": "Inversely proportional to the mean number N of molecules in the observation volume, so it reports local concentration",
      "Effect on the result": "Falls toward noise as concentration rises above about 100 nM",
      "Source": "Yu et al. 2021; Tian et al. 2011; Jovicic et al. 2024"
    },
    {
      "Item": "Decay time τD",
      "Kind": "Fitted parameter",
      "What it is or what it does": "Mean time a molecule stays in the observation volume by diffusion; gives the diffusion coefficient D and, via Stokes-Einstein, the hydrodynamic radius",
      "Effect on the result": "Depends on the calibrated volume size, refractive index and cover-slip thickness",
      "Source": "Jovicic et al. 2024; Yu et al. 2021"
    },
    {
      "Item": "Molecular brightness",
      "Kind": "Fitted parameter",
      "What it is or what it does": "Count rate per molecule, returned by the fit alongside N, D and tau_D",
      "Effect on the result": "Reduced by photobleaching and by optical saturation",
      "Source": "Tian et al. 2011"
    },
    {
      "Item": "Photobleaching",
      "Kind": "Error source",
      "What it is or what it does": "Dye destroyed while crossing the volume",
      "Effect on the result": "Lower apparent N and shorter apparent diffusion times",
      "Source": "Yu et al. 2021"
    },
    {
      "Item": "Background and dark counts",
      "Kind": "Error source",
      "What it is or what it does": "Uncorrelated signal from autofluorescence and detector counts",
      "Effect on the result": "Lower curve amplitude, distorting N",
      "Source": "Yu et al. 2021"
    },
    {
      "Item": "Optical saturation",
      "Kind": "Error source",
      "What it is or what it does": "Excitation power high enough that emission no longer scales with intensity",
      "Effect on the result": "Named the most notable cause of inaccuracy and irreproducibility in FCS results",
      "Source": "Jovicic et al. 2024"
    },
    {
      "Item": "Aggregates",
      "Kind": "Error source",
      "What it is or what it does": "A few bright, slow particles among many small ones",
      "Effect on the result": "Distorted curves; FCS is unreliable for aggregating samples",
      "Source": "Yu et al. 2021"
    },
    {
      "Item": "Spectral crosstalk (FCCS)",
      "Kind": "Error source",
      "What it is or what it does": "Dye emission leaking into the other channel",
      "Effect on the result": "False-positive cross-correlation",
      "Source": "Yu et al. 2021"
    },
    {
      "Item": "Imperfect volume overlap (FCCS)",
      "Kind": "Error source",
      "What it is or what it does": "Two foci not coincident",
      "Effect on the result": "False-negative cross-correlation",
      "Source": "Yu et al. 2021; Tian et al. 2011"
    }
  ],
  "sources": [
    {
      "title": "Yu L et al., A Comprehensive Review of Fluorescence Correlation Spectroscopy, Frontiers in Physics (2021)",
      "url": "https://www.frontiersin.org/articles/10.3389/fphy.2021.644450/full",
      "accessed": "2026-11-19"
    },
    {
      "title": "Tian Y, Martinez MM, Pappas D, Fluorescence correlation spectroscopy: a review of biochemical and microfluidic applications, Applied Spectroscopy (2011)",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC3071976",
      "accessed": "2026-11-19"
    },
    {
      "title": "Review of contemporary fluorescence correlation spectroscopy method in diverse solution studies, Quarterly Reviews of Biophysics (2024)",
      "url": "https://www.cambridge.org/core/journals/quarterly-reviews-of-biophysics/article/review-of-contemporary-fluorescence-correlation-spectroscopy-method-in-diverse-solution-studies/46F1C911D3C332CB93DAEC14AE322180",
      "accessed": "2026-11-19"
    }
  ]
}