  "name": "Freeze-pump-thaw parameters as stated in four laboratory procedures",
  "description": "Side-by-side comparison of the freezing bath, pump time per cycle, thaw method, number of cycles and stopping rule given in four openly readable freeze-pump-thaw procedures: an Oregon State SOP, the LibreTexts Schlenk Line Survival Guide, a University of Washington handout and Jeffrey Bode's Not Voodoo guide. Values are copied as stated by each source; no source gives a vacuum pressure.",
  "page": "https://casrai.org/guides/freeze-pump-thaw",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-19T20:54:03Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Source",
      "unit": "",
      "description": ""
    },
    {
      "name": "Freezing bath",
      "unit": "",
      "description": ""
    },
    {
      "name": "Pump time per cycle",
      "unit": "minutes",
      "description": ""
    },
    {
      "name": "Thaw method",
      "unit": "",
      "description": ""
    },
    {
      "name": "Cycles",
      "unit": "",
      "description": ""
    },
    {
      "name": "Stopping rule",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Source": "Oregon State SOP (Beaudry)",
      "Freezing bath": "Liquid nitrogen (preferred)",
      "Pump time per cycle": "10-30, or until headspace evacuated",
      "Thaw method": "Tepid water bath until it just melts",
      "Cycles": "At least 3",
      "Stopping rule": "Minimum of three cycles, then backfill with inert gas"
    },
    {
      "Source": "LibreTexts Schlenk Line Survival Guide",
      "Freezing bath": "Dry ice/acetone at -78 C; liquid nitrogen for valuable liquids",
      "Pump time per cycle": "3-5, flask in cooling bath",
      "Thaw method": "Remove from bath and warm slowly; no warm water bath or heat gun",
      "Cycles": "3",
      "Stopping rule": "Three cycles in total"
    },
    {
      "Source": "University of Washington (Lonergan)",
      "Freezing bath": "Liquid nitrogen (usually best)",
      "Pump time per cycle": "10-30",
      "Thaw method": "Tepid water bath until it just melts",
      "Cycles": "Minimum 3",
      "Stopping rule": "Repeat until no gas evolves on thawing, minimum three cycles"
    },
    {
      "Source": "Not Voodoo (Bode)",
      "Freezing bath": "Liquid nitrogen",
      "Pump time per cycle": "2-3, high vacuum, flask in liquid nitrogen",
      "Thaw method": "Warm until completely melted",
      "Cycles": "Usually 3",
      "Stopping rule": "Three cycles usually; strict procedures may prescribe 10 for sensitive catalysts"
    }
  ],
  "sources": [
    {
      "title": "Oregon State SOP Freeze-Pump-Thaw Degassing of Liquids",
      "url": "https://sites.science.oregonstate.edu/chemistry/blakemore/safety/SOPs/SOP_freezepumpthaw/SOP_freezepumpthaw.html",
      "accessed": "2026-11-19"
    },
    {
      "title": "LibreTexts Schlenk Line Survival Guide 1.08 Freeze-Pump-Thaw",
      "url": "https://chem.libretexts.org/Bookshelves/Inorganic_Chemistry/The_Schlenk_Line_Survival_Guide_(Borys)/01%3A_Guides/1.08%3A_Freeze-Pump-Thaw",
      "accessed": "2026-11-19"
    },
    {
      "title": "University of Washington Freeze-Pump-Thaw Degassing of Liquids",
      "url": "https://depts.washington.edu/eooptic/linkfiles/Freeze_Pump_Thaw.pdf",
      "accessed": "2026-11-19"
    },
    {
      "title": "How to Degas Solvents (Bode, Not Voodoo; EPFL LCSO copy)",
      "url": "https://www.epfl.ch/labs/lcso/wp-content/uploads/2018/06/Solvent-Degassing.pdf",
      "accessed": "2026-11-19"
    }
  ]
}