  "name": "Griess assay kit formats: Promega G2930 vs Cell Signaling #13547",
  "description": "Side-by-side of the published protocol parameters of two commercial Griess nitrite kits, compiled from the Promega TB229 technical bulletin and the Cell Signaling Technology #13547 datasheet, both read on 2026-11-15. Values are as stated by each vendor for its own kit and are not a performance comparison.",
  "page": "https://casrai.org/guides/griess-assay",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-15T11:02:41Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Parameter",
      "unit": "",
      "description": ""
    },
    {
      "name": "Promega Griess Reagent System (G2930)",
      "unit": "",
      "description": ""
    },
    {
      "name": "Cell Signaling Technology #13547",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Parameter": "Reagent delivery",
      "Promega Griess Reagent System (G2930)": "Sequential: sulfanilamide first, then NED",
      "Cell Signaling Technology #13547": "Premixed: reagents A and B mixed in equal volumes as a working reagent"
    },
    {
      "Parameter": "Volume per well",
      "Promega Griess Reagent System (G2930)": "50 µl sample + 50 µl sulfanilamide + 50 µl NED",
      "Cell Signaling Technology #13547": "100 µl sample or standard + 100 µl working reagent"
    },
    {
      "Parameter": "Incubation",
      "Promega Griess Reagent System (G2930)": "5-10 min after each reagent, room temperature, protected from light",
      "Cell Signaling Technology #13547": "Not stated on the datasheet read"
    },
    {
      "Parameter": "Read wavelength",
      "Promega Griess Reagent System (G2930)": "520-550 nm filter",
      "Cell Signaling Technology #13547": "550 nm"
    },
    {
      "Parameter": "Standard curve range",
      "Promega Griess Reagent System (G2930)": "100 to 1.56 µM nitrite plus 0 µM",
      "Cell Signaling Technology #13547": "100 to 1.56 µM nitrite"
    },
    {
      "Parameter": "Reactions per kit",
      "Promega Griess Reagent System (G2930)": "1,000",
      "Cell Signaling Technology #13547": "500"
    },
    {
      "Parameter": "Stated sensitivity",
      "Promega Griess Reagent System (G2930)": "2.5 µM (125 pmol) nitrite in ultrapure water; depends on matrix",
      "Cell Signaling Technology #13547": "Micromolar nitrite; range depends on matrix or buffer"
    },
    {
      "Parameter": "Intended use",
      "Promega Griess Reagent System (G2930)": "Research; matrices include plasma, serum, urine, tissue culture medium",
      "Cell Signaling Technology #13547": "Research use only, not diagnostic; example RAW 264.7 supernatants"
    }
  ],
  "sources": [
    {
      "title": "Promega TB229 Griess Reagent System",
      "url": "https://promega.com/~/media/files/resources/protocols/technical%20bulletins/0/griess%20reagent%20system%20protocol.pdf",
      "accessed": "2026-11-15"
    },
    {
      "title": "Cell Signaling Technology #13547 datasheet",
      "url": "https://media.cellsignal.com/pdf/13547.pdf",
      "accessed": "2026-11-15"
    }
  ]
}