  "name": "Trypan blue hemocytometer counting parameters stated by two primary protocols (NIH/Current Protocols and Thermo Fisher Gibco)",
  "description": "Side-by-side of the working parameters for the trypan blue exclusion count on a hemocytometer, compiled from the Strober 2015 Current Protocols in Immunology appendix and the Thermo Fisher Gibco trypan blue exclusion protocol, plus Thermo Fisher's published manual-versus-automated counting comparison. Values are copied as stated; 'not stated' marks a parameter the source does not give.",
  "page": "https://casrai.org/guides/hemocytometer-cell-counting-viability-guide",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-18T17:43:58Z",
  "temporalCoverage": "2015/2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Parameter",
      "unit": "",
      "description": ""
    },
    {
      "name": "Unit",
      "unit": "",
      "description": ""
    },
    {
      "name": "Strober 2015 (Curr Protoc Immunol)",
      "unit": "",
      "description": ""
    },
    {
      "name": "Thermo Fisher Gibco protocol",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Parameter": "Trypan blue stock concentration",
      "Unit": "% w/v",
      "Strober 2015 (Curr Protoc Immunol)": 0.4,
      "Thermo Fisher Gibco protocol": 0.4
    },
    {
      "Parameter": "Dye to cell suspension ratio",
      "Unit": "parts dye : parts suspension",
      "Strober 2015 (Curr Protoc Immunol)": "1:1",
      "Thermo Fisher Gibco protocol": "1:1"
    },
    {
      "Parameter": "Volume of each component mixed",
      "Unit": "µL",
      "Strober 2015 (Curr Protoc Immunol)": "10 to 20",
      "Thermo Fisher Gibco protocol": 10
    },
    {
      "Parameter": "Volume loaded into chamber",
      "Unit": "µL",
      "Strober 2015 (Curr Protoc Immunol)": "not stated",
      "Thermo Fisher Gibco protocol": 10
    },
    {
      "Parameter": "Time between mixing and counting",
      "Unit": "minutes",
      "Strober 2015 (Curr Protoc Immunol)": "count within 3 to 5",
      "Thermo Fisher Gibco protocol": "examine immediately"
    },
    {
      "Parameter": "Pre-count cell wash",
      "Unit": "",
      "Strober 2015 (Curr Protoc Immunol)": "centrifuge 100 x g for 5 min, resuspend in PBS or serum-free medium",
      "Thermo Fisher Gibco protocol": "not stated"
    },
    {
      "Parameter": "Serum in counting solution",
      "Unit": "",
      "Strober 2015 (Curr Protoc Immunol)": "serum-free required (serum proteins stain)",
      "Thermo Fisher Gibco protocol": "not stated"
    },
    {
      "Parameter": "Viability benchmark for healthy log-phase culture",
      "Unit": "% viable",
      "Strober 2015 (Curr Protoc Immunol)": "not stated",
      "Thermo Fisher Gibco protocol": 95
    },
    {
      "Parameter": "Concentration formula",
      "Unit": "",
      "Strober 2015 (Curr Protoc Immunol)": "viable count x dilution factor (2 for 1:1)",
      "Thermo Fisher Gibco protocol": "viable count x 10^4 x dilution factor"
    }
  ],
  "sources": [
    {
      "title": "Strober W. Trypan Blue Exclusion Test of Cell Viability. Current Protocols in Immunology 111:A3.B.1-3",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC6716531",
      "accessed": "2026-11-18"
    },
    {
      "title": "Thermo Fisher Gibco: Trypan Blue Exclusion",
      "url": "https://www.thermofisher.com/us/en/home/references/gibco-cell-culture-basics/cell-culture-protocols/trypan-blue-exclusion.html",
      "accessed": "2026-11-18"
    }
  ]
}