  "name": "Cleanroom and Laboratory Biosafety Cabinet HEPA Filter Specifications",
  "description": "Cross-manufacturer technical specifications comparing nominal dimensions, actual physical casing dimensions, media pack depth, rated laminar airflow capacity, face velocity, clean initial resistance, and frame seal configurations for cleanroom ceiling modules and Class II biosafety cabinets. Compiled from IEST-RP-CC001.7, NSF/ANSI 49, Camfil Megalam data sheets, and Labconco engineering manuals.",
  "page": "https://casrai.org/guides/hepa-filter-sizes",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-18T17:00:00Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Equipment and Filter Application",
      "unit": "",
      "description": "Standard cleanroom module, laminar flow bench, or biosafety cabinet model class"
    },
    {
      "name": "Nominal Width",
      "unit": "in",
      "description": "Standard catalog nominal width"
    },
    {
      "name": "Nominal Height",
      "unit": "in",
      "description": "Standard catalog nominal height or length"
    },
    {
      "name": "Nominal Depth",
      "unit": "in",
      "description": "Standard casing depth"
    },
    {
      "name": "Actual Width",
      "unit": "in",
      "description": "Exact measured outer frame width"
    },
    {
      "name": "Actual Height",
      "unit": "in",
      "description": "Exact measured outer frame height"
    },
    {
      "name": "Actual Depth",
      "unit": "in",
      "description": "Exact measured outer frame depth"
    },
    {
      "name": "Rated Airflow",
      "unit": "CFM",
      "description": "Nominal volumetric airflow capacity at design face velocity"
    },
    {
      "name": "Face Velocity",
      "unit": "FPM",
      "description": "Operating airflow velocity across filter face"
    },
    {
      "name": "Initial Resistance",
      "unit": "in. wg",
      "description": "Clean static pressure drop at rated face velocity"
    },
    {
      "name": "Frame Seal Type",
      "unit": "",
      "description": "Sealing mechanism interface against housing"
    }
  ],
  "rows": [
    {
      "Equipment and Filter Application": "Cleanroom Ceiling FFU Module 2x2",
      "Nominal Width": 24,
      "Nominal Height": 24,
      "Nominal Depth": 3,
      "Actual Width": 23.625,
      "Actual Height": 23.625,
      "Actual Depth": 2.75,
      "Rated Airflow": 450,
      "Face Velocity": 90,
      "Initial Resistance": 0.45,
      "Frame Seal Type": "Knife-Edge Gel Seal"
    },
    {
      "Equipment and Filter Application": "Cleanroom Ceiling FFU Module 2x4",
      "Nominal Width": 24,
      "Nominal Height": 48,
      "Nominal Depth": 3,
      "Actual Width": 23.625,
      "Actual Height": 47.625,
      "Actual Depth": 2.75,
      "Rated Airflow": 900,
      "Face Velocity": 90,
      "Initial Resistance": 0.45,
      "Frame Seal Type": "Knife-Edge Gel Seal"
    },
    {
      "Equipment and Filter Application": "Cleanroom Terminal Panel 2x2 (Gasket)",
      "Nominal Width": 24,
      "Nominal Height": 24,
      "Nominal Depth": 5.875,
      "Actual Width": 23.625,
      "Actual Height": 23.625,
      "Actual Depth": 5.875,
      "Rated Airflow": 500,
      "Face Velocity": 100,
      "Initial Resistance": 0.52,
      "Frame Seal Type": "Neoprene Compression Gasket"
    },
    {
      "Equipment and Filter Application": "Cleanroom Terminal Panel 2x4 (Gasket)",
      "Nominal Width": 24,
      "Nominal Height": 48,
      "Nominal Depth": 5.875,
      "Actual Width": 23.625,
      "Actual Height": 47.625,
      "Actual Depth": 5.875,
      "Rated Airflow": 1000,
      "Face Velocity": 100,
      "Initial Resistance": 0.52,
      "Frame Seal Type": "Neoprene Compression Gasket"
    },
    {
      "Equipment and Filter Application": "Class II BSC 3-Foot Supply Downflow",
      "Nominal Width": 18,
      "Nominal Height": 36,
      "Nominal Depth": 3.06,
      "Actual Width": 18,
      "Actual Height": 36,
      "Actual Depth": 3.06,
      "Rated Airflow": 350,
      "Face Velocity": 55,
      "Initial Resistance": 0.48,
      "Frame Seal Type": "Knife-Edge Gel Seal"
    },
    {
      "Equipment and Filter Application": "Class II BSC 3-Foot Exhaust Filter",
      "Nominal Width": 18,
      "Nominal Height": 18,
      "Nominal Depth": 5.88,
      "Actual Width": 18,
      "Actual Height": 18,
      "Actual Depth": 5.88,
      "Rated Airflow": 180,
      "Face Velocity": 80,
      "Initial Resistance": 0.65,
      "Frame Seal Type": "Neoprene Compression Gasket"
    },
    {
      "Equipment and Filter Application": "Class II BSC 4-Foot Supply Downflow",
      "Nominal Width": 18,
      "Nominal Height": 48,
      "Nominal Depth": 3.06,
      "Actual Width": 18,
      "Actual Height": 48,
      "Actual Depth": 3.06,
      "Rated Airflow": 475,
      "Face Velocity": 55,
      "Initial Resistance": 0.48,
      "Frame Seal Type": "Knife-Edge Gel Seal"
    },
    {
      "Equipment and Filter Application": "Class II BSC 4-Foot Exhaust Filter",
      "Nominal Width": 18,
      "Nominal Height": 24,
      "Nominal Depth": 11.5,
      "Actual Width": 18,
      "Actual Height": 24,
      "Actual Depth": 11.5,
      "Rated Airflow": 290,
      "Face Velocity": 95,
      "Initial Resistance": 0.58,
      "Frame Seal Type": "Neoprene Compression Gasket"
    },
    {
      "Equipment and Filter Application": "Class II BSC 6-Foot Supply Downflow",
      "Nominal Width": 18,
      "Nominal Height": 72,
      "Nominal Depth": 3.06,
      "Actual Width": 18,
      "Actual Height": 72,
      "Actual Depth": 3.06,
      "Rated Airflow": 710,
      "Face Velocity": 55,
      "Initial Resistance": 0.48,
      "Frame Seal Type": "Knife-Edge Gel Seal"
    },
    {
      "Equipment and Filter Application": "Class II BSC 6-Foot Exhaust Filter",
      "Nominal Width": 18,
      "Nominal Height": 30,
      "Nominal Depth": 11.5,
      "Actual Width": 18,
      "Actual Height": 30,
      "Actual Depth": 11.5,
      "Rated Airflow": 435,
      "Face Velocity": 95,
      "Initial Resistance": 0.58,
      "Frame Seal Type": "Neoprene Compression Gasket"
    }
  ],
  "sources": [
    {
      "title": "IEST-RP-CC001.7: HEPA and ULPA Filters",
      "url": "https://www.iest.org/Standards-RPs/Recommended-Practices/IEST-RP-CC001",
      "accessed": "2026-11-14"
    },
    {
      "title": "NSF/ANSI Standard 49: Biosafety Cabinetry Design and Certification",
      "url": "https://www.nsf.org/standards-development/standards-portfolio/biosafety-cabinetry-standards",
      "accessed": "2026-11-14"
    },
    {
      "title": "Camfil Megalam Cleanroom Panel Filter Product Specifications",
      "url": "https://www.camfil.com/en-us/products/high-efficiency-filters",
      "accessed": "2026-11-14"
    },
    {
      "title": "Labconco Purifier Logic+ Biosafety Cabinet Replacement HEPA Specifications",
      "url": "https://www.labconco.com/category/class-ii-biosafety-cabinets",
      "accessed": "2026-11-14"
    }
  ]
}