  "name": "What written SOP rules and guidance require: FDA GLP, EPA GLP, FDA drug cGMP and EPA QA/G-6",
  "description": "Side-by-side comparison of what four U.S. sources say about written standard operating procedures: whether SOPs are required, who authorizes changes, how deviations are handled, availability at the point of use, and revision records. Cells say 'Not addressed' where the cited section does not cover the point.",
  "page": "https://casrai.org/guides/how-to-write-a-lab-sop-template-guide",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-15T16:36:27Z",
  "temporalCoverage": "2007/2026",
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Source",
      "unit": "",
      "description": ""
    },
    {
      "name": "Status",
      "unit": "",
      "description": "Regulation or guidance"
    },
    {
      "name": "Who approves SOPs or significant changes",
      "unit": "",
      "description": ""
    },
    {
      "name": "How deviations are handled",
      "unit": "",
      "description": ""
    },
    {
      "name": "Availability at point of use",
      "unit": "",
      "description": ""
    },
    {
      "name": "Revision records and review interval",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Source": "21 CFR 58.81 (FDA GLP, nonclinical laboratory studies)",
      "Status": "Regulation",
      "Who approves SOPs or significant changes": "Management must be satisfied SOPs are adequate; significant changes authorized in writing by management",
      "How deviations are handled": "Authorized by the study director and documented in the raw data",
      "Availability at point of use": "Manuals and SOPs immediately available in each laboratory area; published literature may supplement SOPs",
      "Revision records and review interval": "Historical file of SOPs and all revisions with revision dates; no review interval stated"
    },
    {
      "Source": "40 CFR 160.81 (EPA FIFRA GLP)",
      "Status": "Regulation",
      "Who approves SOPs or significant changes": "Management must be satisfied SOPs are adequate; significant changes authorized in writing by management",
      "How deviations are handled": "Authorized by the study director and documented in the raw data",
      "Availability at point of use": "Manuals and SOPs immediately available in each laboratory or other study area; published literature may supplement SOPs",
      "Revision records and review interval": "Historical file of SOPs and all revisions with revision dates; no review interval stated"
    },
    {
      "Source": "21 CFR 211.100 (FDA drug cGMP, production and process control)",
      "Status": "Regulation",
      "Who approves SOPs or significant changes": "Procedures and any changes drafted, reviewed and approved by the appropriate organizational units and reviewed and approved by the quality control unit",
      "How deviations are handled": "Any departure from written procedures recorded and justified; work documented when performed",
      "Availability at point of use": "Not addressed in 211.100(a)-(b)",
      "Revision records and review interval": "Not addressed in 211.100(a)-(b)"
    },
    {
      "Source": "EPA QA/G-6 (April 2007)",
      "Status": "Guidance",
      "Who approves SOPs or significant changes": "Generally the immediate supervisor and the quality assurance officer review and approve each SOP",
      "How deviations are handled": "Not addressed",
      "Availability at point of use": "Not addressed as a rule; electronic access can be read-only",
      "Revision records and review interval": "Master list of SOPs with version and date; outdated versions archived; periodic review, e.g. every 1-2 years, with review date added"
    }
  ],
  "sources": [
    {
      "title": "21 CFR 58.81 Standard operating procedures",
      "url": "https://www.law.cornell.edu/cfr/text/21/58.81",
      "accessed": "2026-11-15"
    },
    {
      "title": "40 CFR 160.81 Standard operating procedures",
      "url": "https://www.law.cornell.edu/cfr/text/40/160.81",
      "accessed": "2026-11-15"
    },
    {
      "title": "21 CFR 211.100 Written procedures; deviations",
      "url": "https://www.law.cornell.edu/cfr/text/21/211.100",
      "accessed": "2026-11-15"
    },
    {
      "title": "Guidance for Preparing Standard Operating Procedures (EPA QA/G-6), EPA/600/B-07/001",
      "url": "https://www.epa.gov/sites/default/files/2015-06/documents/g6-final.pdf",
      "accessed": "2026-11-15"
    }
  ]
}