  "name": "HPLC Instrument Module Calibration Parameters and Acceptance Criteria",
  "description": "Original compilation of operational calibration parameters, test standards, and regulatory acceptance thresholds for HPLC modules, synthesized from USP Chapter 1058, WHO Good Chromatography Practices, and NIST SRM 2034.",
  "page": "https://casrai.org/guides/hplc-calibration",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-22T18:40:08Z",
  "temporalCoverage": "2020-2026",
  "spatialCoverage": "International",
  "columns": [
    {
      "name": "Module",
      "unit": "",
      "description": "HPLC functional hardware module"
    },
    {
      "name": "Parameter",
      "unit": "",
      "description": "Specific metrological parameter evaluated"
    },
    {
      "name": "Test Method",
      "unit": "",
      "description": "Operational verification procedure"
    },
    {
      "name": "Traceable Standard or Tool",
      "unit": "",
      "description": "External reference standard or calibrated instrument"
    },
    {
      "name": "Acceptance Limit",
      "unit": "",
      "description": "Numerical pass/fail tolerance"
    }
  ],
  "rows": [
    {
      "Module": "Solvent Delivery System (Pump)",
      "Parameter": "Flow Rate Accuracy",
      "Test Method": "Timed gravimetric water collection or digital flow meter",
      "Traceable Standard or Tool": "Calibrated analytical balance and NIST-calibrated timer",
      "Acceptance Limit": "Within ±1.0% to ±2.0% of setpoint"
    },
    {
      "Module": "Solvent Delivery System (Pump)",
      "Parameter": "Gradient Proportioning Accuracy",
      "Test Method": "Step-gradient tracer absorbance test (0.1% acetone)",
      "Traceable Standard or Tool": "UV detector monitoring at 265 nm",
      "Acceptance Limit": "Within ±1.0% of nominal programmed concentration"
    },
    {
      "Module": "Autosampler / Injector",
      "Parameter": "Injection Precision (Repeatability)",
      "Test Method": "Replicate injections (n=6) of caffeine or uracil test mix",
      "Traceable Standard or Tool": "USP reference standard solution",
      "Acceptance Limit": "Peak area RSD ≤ 1.0% (RSD ≤ 0.5% for high-precision assays)"
    },
    {
      "Module": "Spectrophotometric Detector (UV/Vis)",
      "Parameter": "Wavelength Accuracy",
      "Test Method": "Spectral scan across certified intrinsic absorption bands",
      "Traceable Standard or Tool": "NIST SRM 2034 Holmium Oxide Solution (or D-alpha line)",
      "Acceptance Limit": "Within ±1.0 nm to ±2.0 nm of certified reference values"
    },
    {
      "Module": "Spectrophotometric Detector (UV/Vis)",
      "Parameter": "Response Linearity",
      "Test Method": "Multi-point serial dilution series (caffeine or dichromate)",
      "Traceable Standard or Tool": "Class A certified chemical reference standards",
      "Acceptance Limit": "Linear correlation coefficient r² ≥ 0.999"
    },
    {
      "Module": "Column Compartment",
      "Parameter": "Temperature Accuracy",
      "Test Method": "Contact thermocouple probe measurement at 30, 40, and 60 °C",
      "Traceable Standard or Tool": "NIST-calibrated digital thermometer",
      "Acceptance Limit": "Within ±1.0 °C of nominal setpoint"
    }
  ],
  "sources": [
    {
      "title": "WHO Technical Report Series No. 1025, Annex 4 (Good chromatography practices)",
      "url": "https://cdn.who.int/media/docs/default-source/medicines/norms-and-standards/guidelines/trs1025/trs1025-annex4.pdf",
      "accessed": "2026-11-22"
    },
    {
      "title": "NIST SRM 2034 Holmium Oxide Solution Wavelength Standard",
      "url": "https://tsapps.nist.gov/srmext/certificates/2034.pdf",
      "accessed": "2026-11-22"
    },
    {
      "title": "FDA 21 CFR § 211.160 Laboratory Controls",
      "url": "https://www.law.cornell.edu/cfr/text/21/211.160",
      "accessed": "2026-11-22"
    }
  ]
}