  "name": "Reported equilibration and reoxygenation times in hypoxic cell culture",
  "description": "Timing figures for reaching or losing low oxygen in cell culture, compiled from three peer-reviewed reviews (theoretical calculations and reported observations), with the conditions each source states.",
  "page": "https://casrai.org/guides/hypoxia-chamber",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-20T08:40:00Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Scenario",
      "unit": "",
      "description": ""
    },
    {
      "name": "Stated conditions",
      "unit": "",
      "description": ""
    },
    {
      "name": "Reported time",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Scenario": "Theoretical equilibration to 2% O2",
      "Stated conditions": "1.72 mm medium height; within 1.2-fold of input",
      "Reported time": "38 minutes",
      "Source": "Wenger et al. 2015"
    },
    {
      "Scenario": "Theoretical equilibration to 1% O2",
      "Stated conditions": "1.72 mm medium height; within 1.2-fold of input",
      "Reported time": "45 minutes",
      "Source": "Wenger et al. 2015"
    },
    {
      "Scenario": "Theoretical equilibration to 0.2% O2",
      "Stated conditions": "1.72 mm medium height; within 1.2-fold of input",
      "Reported time": "60 minutes",
      "Source": "Wenger et al. 2015"
    },
    {
      "Scenario": "Recovery after brief door opening",
      "Stated conditions": "Hypoxic culture re-established",
      "Reported time": "Up to 1 hour",
      "Source": "Wenger et al. 2015"
    },
    {
      "Scenario": "Thin medium barrier",
      "Stated conditions": "170 µm medium, modular chamber",
      "Reported time": "About 1.5 minutes",
      "Source": "Pavlacky and Polak 2020"
    },
    {
      "Scenario": "Thicker medium overlay",
      "Stated conditions": "1 mm medium overlay",
      "Reported time": "About 10 times longer than 170 µm",
      "Source": "Pavlacky and Polak 2020"
    },
    {
      "Scenario": "Pericellular O2 after medium exchange",
      "Stated conditions": "Equilibration after exchanging medium",
      "Reported time": "About 2 hours",
      "Source": "Pavlacky and Polak 2020"
    },
    {
      "Scenario": "Culture-ware reaching adjusted O2",
      "Stated conditions": "Gas-controlled incubator",
      "Reported time": "Several hours",
      "Source": "Yazdani 2016"
    }
  ],
  "sources": [
    {
      "title": "Frequently asked questions in hypoxia research",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC5045069/",
      "accessed": "2026-11-14"
    },
    {
      "title": "Technical Feasibility and Physiological Relevance of Hypoxic Cell Culture Models",
      "url": "https://www.frontiersin.org/articles/10.3389/fendo.2020.00057/full",
      "accessed": "2026-11-14"
    },
    {
      "title": "Technical aspects of oxygen level regulation in primary cell cultures: a review",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC5464680/",
      "accessed": "2026-11-14"
    }
  ]
}