  "name": "CO2 incubator maintenance intervals compared across a manufacturer manual, a manufacturer guide and a biosafety office",
  "description": "Compares how three sources schedule routine CO2 incubator maintenance tasks: the Thermo Scientific 3010/3020 operating manual, Esco Lifesciences' maintenance guide and the University of Utah Institutional Biosafety Committee. Values are transcribed as each source states them; where a source gives no value the cell says so.",
  "page": "https://casrai.org/guides/incubator-repair",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T15:01:17Z",
  "temporalCoverage": "2021/2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Task",
      "unit": "",
      "description": ""
    },
    {
      "name": "Thermo Scientific manual 7013010 Rev. 6",
      "unit": "",
      "description": ""
    },
    {
      "name": "Esco Lifesciences guide V.B Mar2021",
      "unit": "",
      "description": ""
    },
    {
      "name": "University of Utah IBC",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Task": "Humidity pan water",
      "Thermo Scientific manual 7013010 Rev. 6": "Check level; sterile, distilled or demineralised water, fill to within 1/2 inch of top",
      "Esco Lifesciences guide V.B Mar2021": "Refill with sterile distilled water once a week",
      "University of Utah IBC": "Sterile distilled water, pH 7-9; change every two weeks; keep pan full"
    },
    {
      "Task": "Interior cleaning",
      "Thermo Scientific manual 7013010 Rev. 6": "Complete decontamination between experiments; more often depending on use and conditions",
      "Esco Lifesciences guide V.B Mar2021": "Weekly interior and exterior cleaning; inner chamber decontamination as needed",
      "University of Utah IBC": "Clean 1-2 times per month depending on number of users; top every 2 weeks"
    },
    {
      "Task": "Air, HEPA or ULPA filters",
      "Thermo Scientific manual 7013010 Rev. 6": "Air sample and access port filters every 6 months recommended; HEPA and gas filters or as needed",
      "Esco Lifesciences guide V.B Mar2021": "ULPA filter (170 L/240 L) yearly or when visibly dirty",
      "University of Utah IBC": "HEPA filters every 6 months to a year depending on use"
    },
    {
      "Task": "Gas inlet filters",
      "Thermo Scientific manual 7013010 Rev. 6": "Change as needed (preventive maintenance table)",
      "Esco Lifesciences guide V.B Mar2021": "At least yearly or when dirty",
      "University of Utah IBC": "6 months to 1 year"
    },
    {
      "Task": "Calibration",
      "Thermo Scientific manual 7013010 Rev. 6": "Verify and document CO2 and temperature calibration; interval depends on use, environment and accuracy",
      "Esco Lifesciences guide V.B Mar2021": "Yearly for temperature, CO2 and humidity sensors; oxygen sensor yearly and as needed",
      "University of Utah IBC": "CO2 calibration monthly to quarterly depending on door traffic"
    },
    {
      "Task": "Door gasket",
      "Thermo Scientific manual 7013010 Rev. 6": "Inspect door gasket seal",
      "Esco Lifesciences guide V.B Mar2021": "Replace outer magnetic gasket as needed (brittleness, corrosion, wear, damage)",
      "University of Utah IBC": "Not covered"
    },
    {
      "Task": "Heat sterilisation cycle",
      "Thermo Scientific manual 7013010 Rev. 6": "Not covered",
      "Esco Lifesciences guide V.B Mar2021": "Run decontamination or sterilisation cycle if fitted",
      "University of Utah IBC": "Once a month to once every six months"
    },
    {
      "Task": "Gas tank",
      "Thermo Scientific manual 7013010 Rev. 6": "Check CO2 tank levels",
      "Esco Lifesciences guide V.B Mar2021": "Daily; regulator gauge not below 15 psig",
      "University of Utah IBC": "Not covered"
    }
  ],
  "sources": [
    {
      "title": "Thermo Scientific Operating and Maintenance Manual 7013010 Rev. 6",
      "url": "https://documents.thermofisher.com/TFS-Assets/LED/manuals/7013010r6-EN-3010-Series-Water-Jacket.pdf",
      "accessed": "2026-11-21"
    },
    {
      "title": "Esco Lifesciences CO2 Incubator Maintenance Tips and Procedures V.B Mar2021",
      "url": "https://www.escolifesciences.com/pdf/CO2%20Incubator%20Maintenance%20Tips%20and%20Procedures-V.B-Mar2021.pdf",
      "accessed": "2026-11-21"
    },
    {
      "title": "University of Utah IBC Best Practices for CO2 Incubator Maintenance",
      "url": "https://ibc.utah.edu/library/best-practices-for-co2-incubator--maintenance.php",
      "accessed": "2026-11-21"
    }
  ]
}