  "name": "IVD and Medical Device Clinical Study Regulatory Frameworks",
  "description": "Comparative analysis of clinical study requirements, ethical review rules, informed consent thresholds, and specimen tracking across international standards and regulatory frameworks including ISO 20916, ISO 14155, EU IVDR 2017/746, and US FDA 21 CFR 812.",
  "page": "https://casrai.org/guides/iso-20916",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-24T11:15:00Z",
  "temporalCoverage": "2019/2026",
  "spatialCoverage": "Global, European Union, United States",
  "columns": [
    {
      "name": "Standard or Regulation",
      "unit": "",
      "description": "Official designation and title of the standard or regulation"
    },
    {
      "name": "Target Scope",
      "unit": "",
      "description": "Medical technology or device category governed"
    },
    {
      "name": "Ethics Oversight Requirement",
      "unit": "",
      "description": "Mandatory Institutional Review Board or Ethics Committee approval"
    },
    {
      "name": "Leftover Specimen Consent Rule",
      "unit": "",
      "description": "Informed consent requirements for archived or surplus specimens"
    },
    {
      "name": "Core Study Documentation",
      "unit": "",
      "description": "Primary protocol and clinical performance reporting instruments"
    },
    {
      "name": "European Union Regulatory Status",
      "unit": "",
      "description": "Harmonisation and legal standing under EU MDR and IVDR"
    },
    {
      "name": "US FDA Regulatory Status",
      "unit": "",
      "description": "Recognition in FDA consensus standards database and acceptance"
    }
  ],
  "rows": [
    {
      "Standard or Regulation": "ISO 20916:2019",
      "Target Scope": "In vitro diagnostic medical devices using human specimens",
      "Ethics Oversight Requirement": "Mandatory for interventional studies; tiered for non-interventional",
      "Leftover Specimen Consent Rule": "Exemptions permitted for de-identified leftover specimens under specific conditions",
      "Core Study Documentation": "Clinical Performance Study Protocol (CPSP) and Report (CPSR)",
      "European Union Regulatory Status": "Harmonised under IVDR via Decision (EU) 2024/2625 (presumption of conformity)",
      "US FDA Regulatory Status": "Non-Recognized due to human subject protection differences with 21 CFR 50/56"
    },
    {
      "Standard or Regulation": "ISO 14155:2020",
      "Target Scope": "Therapeutic, invasive, and general medical devices",
      "Ethics Oversight Requirement": "Mandatory approval required for all human clinical investigations",
      "Leftover Specimen Consent Rule": "Not applicable (evaluates physical devices directly in or on human subjects)",
      "Core Study Documentation": "Clinical Investigation Plan (CIP) and Report (CIR)",
      "European Union Regulatory Status": "Harmonised under EU MDR 2017/745 (presumption of conformity)",
      "US FDA Regulatory Status": "Complete Recognition (Recognition #2-282 under 21 CFR 812.28 foreign GCP)"
    },
    {
      "Standard or Regulation": "EU IVDR 2017/746",
      "Target Scope": "All in vitro diagnostic devices placed on the European market",
      "Ethics Oversight Requirement": "Mandatory favorable opinion from competent ethics committee prior to study start",
      "Leftover Specimen Consent Rule": "National member state provisions apply; strict controls for residual biosamples",
      "Core Study Documentation": "Clinical Performance Study Plan and Clinical Performance Study Report",
      "European Union Regulatory Status": "Binding primary European Union regulation for CE marking",
      "US FDA Regulatory Status": "Foreign legal framework; not directly applicable in US jurisdiction"
    },
    {
      "Standard or Regulation": "US FDA 21 CFR Part 812",
      "Target Scope": "Investigational medical devices and IVDs in the United States",
      "Ethics Oversight Requirement": "Mandatory IRB review unless expressly exempt under 21 CFR 56.104",
      "Leftover Specimen Consent Rule": "Enforcement discretion for unlinked leftover specimens under 2006 guidance",
      "Core Study Documentation": "IDE application (significant risk) or exempt study protocol with IRB records",
      "European Union Regulatory Status": "Foreign regulatory framework; not directly applicable in EU jurisdiction",
      "US FDA Regulatory Status": "Binding federal regulation under Federal Food, Drug, and Cosmetic Act"
    }
  ],
  "sources": [
    {
      "title": "Standard Norge: ISO 20916:2019",
      "url": "https://online.standard.no/en/iso-20916-2019",
      "accessed": "2026-11-20"
    },
    {
      "title": "European Commission: MDCG 2022-2 Guidance",
      "url": "https://health.ec.europa.eu/system/files/2022-01/mdcg_2022-2_en.pdf",
      "accessed": "2026-11-20"
    },
    {
      "title": "FDA Recognized Consensus Standards Database",
      "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/search.cfm",
      "accessed": "2026-11-20"
    },
    {
      "title": "FDA Guidance on Leftover Human Specimens",
      "url": "https://www.fda.gov/regulatory-information/search-fda-guidance-documents/guidance-informed-consent-in-vitro-diagnostic-device-studies-using-leftover-human-specimens-are-not",
      "accessed": "2026-11-20"
    },
    {
      "title": "Legal Information Institute: 21 CFR 812.2",
      "url": "https://www.law.cornell.edu/cfr/text/21/812.2",
      "accessed": "2026-11-20"
    }
  ]
}