  "name": "LOD and LOQ estimation approaches compared: factors, replicates and confirmation rules by guideline",
  "description": "Compares how ICH Q2(R1), ICH Q2(R2), US EPA 40 CFR 136 Appendix B (MDL), CLSI-style LoB/LoD/LoQ protocols and the statistical derivation in LCGC International set the detection and quantitation limit factors, the minimum number of replicates and the confirmation expectations. Compiled by hand from the guideline texts and one peer-reviewed or trade-journal explanation of each statistical approach; factors are numeric multipliers or ratios as stated in the sources.",
  "page": "https://casrai.org/guides/limit-of-detection-vs-limit-of-quantitation-calculation",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-13T18:48:12Z",
  "temporalCoverage": "1994/2023",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Approach (source)",
      "unit": "",
      "description": ""
    },
    {
      "name": "Detection factor",
      "unit": "multiplier or S/N ratio",
      "description": "Factor applied to sigma (or S/N ratio, or t-value) to reach the detection limit"
    },
    {
      "name": "Quantitation factor",
      "unit": "multiplier or S/N ratio",
      "description": "Factor for the quantitation limit where the source defines one"
    },
    {
      "name": "Minimum replicates stated",
      "unit": "replicates",
      "description": "Smallest number of replicates the source states; blank where none is stated"
    },
    {
      "name": "How the value is confirmed or reported",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Approach (source)": "ICH Q2(R1) signal-to-noise",
      "Detection factor": 3,
      "Quantitation factor": 10,
      "Minimum replicates stated": "not stated",
      "How the value is confirmed or reported": "Q2(R1) says an S/N between 3 or 2:1 is generally acceptable for DL and a typical ratio is 10:1 for QL; the estimate is later validated by analysing samples near the limit"
    },
    {
      "Approach (source)": "ICH Q2(R2) signal-to-noise, section 3.2.3.2",
      "Detection factor": 3,
      "Quantitation factor": 10,
      "Minimum replicates stated": "not stated",
      "How the value is confirmed or reported": "3:1 generally acceptable for DL; at least 10:1 acceptable for QL; S/N region should be defined and, if possible, centred on the peak position"
    },
    {
      "Approach (source)": "ICH Q2(R2) standard deviation and slope, section 3.2.3.3",
      "Detection factor": 3.3,
      "Quantitation factor": 10,
      "Minimum replicates stated": "not stated",
      "How the value is confirmed or reported": "DL = 3.3 sigma/S and QL = 10 sigma/S; sigma from blank SD, residual SD of regression, or SD of y-intercepts; estimates validated on samples at or near the limit; QL confirmation may be omitted with justification if QL is about 10 times below the reporting limit"
    },
    {
      "Approach (source)": "US EPA 40 CFR 136 Appendix B (MDL), Revision 2",
      "Detection factor": 3.143,
      "Quantitation factor": "not defined",
      "Minimum replicates stated": 7,
      "How the value is confirmed or reported": "MDL = t(n-1, 0.99) x SD of spiked replicates (3.143 for 7 replicates, 6 degrees of freedom); at least 7 spiked and 7 method blanks in at least 3 batches on 3 dates; the greater of MDLs and MDLb is the initial MDL"
    },
    {
      "Approach (source)": "CLSI-style LoB/LoD/LoQ (Armbruster and Pry 2008, following CLSI EP17)",
      "Detection factor": 1.645,
      "Quantitation factor": "not defined",
      "Minimum replicates stated": 20,
      "How the value is confirmed or reported": "LoB = mean blank + 1.645 SD blank; LoD = LoB + 1.645 SD of low-level sample; 60 replicates to establish (manufacturers), 20 to verify (laboratories); LoQ is the lowest level meeting predefined bias and imprecision goals"
    },
    {
      "Approach (source)": "Statistical derivation with alpha = beta = 0.05 (Boque and Vander Heyden, LCGC 2009)",
      "Detection factor": 3.3,
      "Quantitation factor": "not defined",
      "Minimum replicates stated": 10,
      "How the value is confirmed or reported": "Critical level 1.64 sigma0, detection limit about 3.3 sigma0; at least 10 portions of a low-level test sample through the full procedure; state the approach used when reporting an LOD"
    }
  ],
  "sources": [
    {
      "title": "ICH Q2(R2) Validation of Analytical Procedures (Step 4, adopted 1 November 2023)",
      "url": "https://database.ich.org/sites/default/files/ICH_Q2%28R2%29_Guideline_2023_1130.pdf",
      "accessed": "2026-11-13"
    },
    {
      "title": "ICH Q2(R1) Validation of Analytical Procedures: Text and Methodology",
      "url": "https://database.ich.org/sites/default/files/Q2%28R1%29%20Guideline.pdf",
      "accessed": "2026-11-13"
    },
    {
      "title": "40 CFR Part 136 Appendix B, Definition and Procedure for the Determination of the Method Detection Limit, Revision 2",
      "url": "https://www.ecfr.gov/current/title-40/chapter-I/subchapter-D/part-136/appendix-Appendix%20B%20to%20Part%20136",
      "accessed": "2026-11-13"
    },
    {
      "title": "Armbruster DA, Pry T. Limit of blank, limit of detection and limit of quantitation. Clin Biochem Rev 2008",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC2556583/",
      "accessed": "2026-11-13"
    },
    {
      "title": "Boque R, Vander Heyden Y. The Limit of Detection. LCGC International",
      "url": "https://chromatographyonline.com/view/limit-detection",
      "accessed": "2026-11-13"
    }
  ]
}