  "name": "Mass Analyzer Performance Specifications and Operational Benchmarks",
  "description": "Original benchmark comparison of the six primary mass spectrometry analyzer types, compiled from IUPAC Recommendations (2013), Makarov et al. (2015), Hendrickson et al. (2015), Douglas et al. (2005), Chernushevich et al. (2001), and Koppenaal et al. (2005). Compares separation mechanisms, resolving power, mass accuracy, upper mass limits, scan rates, and vacuum pressure regimes.",
  "page": "https://casrai.org/guides/mass-spectrometry-types",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-18T15:00:00Z",
  "temporalCoverage": "2001/2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Analyzer Type",
      "unit": "",
      "description": "Common name and architectural classification of the mass analyzer"
    },
    {
      "name": "Separation Mechanism",
      "unit": "",
      "description": "Physical governing principle used to sort ions by m/z"
    },
    {
      "name": "Resolving Power (FWHM)",
      "unit": "m/Δm",
      "description": "Typical resolving power at full width at half maximum"
    },
    {
      "name": "Mass Accuracy Limit",
      "unit": "ppm",
      "description": "Standard upper bound of mass measurement accuracy under routine calibration"
    },
    {
      "name": "Upper Mass Limit",
      "unit": "m/z",
      "description": "Standard operational mass transmission or detection ceiling"
    },
    {
      "name": "Maximum Scan Rate",
      "unit": "Hz",
      "description": "Nominal spectral acquisition frequency"
    },
    {
      "name": "Operating Vacuum Pressure",
      "unit": "mbar",
      "description": "Required vacuum regime in analyzer cell"
    }
  ],
  "rows": [
    {
      "Analyzer Type": "Quadrupole (Q / QqQ)",
      "Separation Mechanism": "Dynamic RF/DC electric field stability filter",
      "Resolving Power (FWHM)": 2000,
      "Mass Accuracy Limit": 10,
      "Upper Mass Limit": 4000,
      "Maximum Scan Rate": 40,
      "Operating Vacuum Pressure": "1.0e-5"
    },
    {
      "Analyzer Type": "Time-of-Flight (oa-TOF)",
      "Separation Mechanism": "Flight velocity drift in field-free flight tube",
      "Resolving Power (FWHM)": 60000,
      "Mass Accuracy Limit": 2,
      "Upper Mass Limit": 20000,
      "Maximum Scan Rate": 100,
      "Operating Vacuum Pressure": "1.0e-7"
    },
    {
      "Analyzer Type": "Linear Ion Trap (LIT)",
      "Separation Mechanism": "Resonant dynamic RF ejection from 2D quadrupole",
      "Resolving Power (FWHM)": 4000,
      "Mass Accuracy Limit": 20,
      "Upper Mass Limit": 4000,
      "Maximum Scan Rate": 30,
      "Operating Vacuum Pressure": "1.0e-3"
    },
    {
      "Analyzer Type": "Orbitrap",
      "Separation Mechanism": "Harmonic axial oscillation around central spindle",
      "Resolving Power (FWHM)": 500000,
      "Mass Accuracy Limit": 1,
      "Upper Mass Limit": 6000,
      "Maximum Scan Rate": 40,
      "Operating Vacuum Pressure": "1.0e-10"
    },
    {
      "Analyzer Type": "Fourier Transform ICR",
      "Separation Mechanism": "Cyclotron frequency oscillation in Penning trap",
      "Resolving Power (FWHM)": 2000000,
      "Mass Accuracy Limit": 0.2,
      "Upper Mass Limit": 10000,
      "Maximum Scan Rate": 2,
      "Operating Vacuum Pressure": "1.0e-10"
    },
    {
      "Analyzer Type": "Magnetic Sector",
      "Separation Mechanism": "Momentum and kinetic energy dispersion in B/E fields",
      "Resolving Power (FWHM)": 100000,
      "Mass Accuracy Limit": 1,
      "Upper Mass Limit": 10000,
      "Maximum Scan Rate": 5,
      "Operating Vacuum Pressure": "1.0e-8"
    }
  ],
  "sources": [
    {
      "title": "Definitions of terms relating to mass spectrometry (IUPAC Recommendations 2013)",
      "url": "https://doi.org/10.1351/PAC-REC-06-04-06",
      "accessed": "2026-11-14"
    },
    {
      "title": "Evolution of Orbitrap Mass Spectrometry Instrumentation (Eliuk & Makarov 2015)",
      "url": "https://doi.org/10.1146/annurev-anchem-071114-040325",
      "accessed": "2026-11-14"
    },
    {
      "title": "21 Tesla FT-ICR Mass Spectrometer (Hendrickson et al. 2015)",
      "url": "https://doi.org/10.1007/s13361-015-1182-2",
      "accessed": "2026-11-14"
    },
    {
      "title": "Linear ion traps in mass spectrometry (Douglas et al. 2005)",
      "url": "https://doi.org/10.1002/mas.20004",
      "accessed": "2026-11-14"
    },
    {
      "title": "An introduction to quadrupole–time-of-flight mass spectrometry (Chernushevich et al. 2001)",
      "url": "https://doi.org/10.1002/jms.207",
      "accessed": "2026-11-14"
    },
    {
      "title": "MS Detectors (Koppenaal et al. 2005)",
      "url": "https://doi.org/10.1021/ac053494z",
      "accessed": "2026-11-14"
    }
  ]
}