  "name": "Laboratory Membrane Filtration Water Testing Specifications and Regulatory Standards",
  "description": "Comparative laboratory specifications, hardware parameters, operating thresholds, and compliance benchmarks for membrane filtration water testing systems across EPA, Standard Methods, ISO, and USP standards.",
  "page": "https://casrai.org/guides/membrane-filtration-water",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-15T09:44:22Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "Global / United States and International",
  "columns": [
    {
      "name": "Standard Specification Parameter",
      "unit": "",
      "description": "Key hardware, consumable, or operational parameter"
    },
    {
      "name": "Standard Methods 9222 (Coliform Group)",
      "unit": "",
      "description": "APHA/AWWA/WEF standard for water and wastewater"
    },
    {
      "name": "EPA Method 1604 (Coliforms and E. coli)",
      "unit": "",
      "description": "US EPA simultaneous detection using MI medium"
    },
    {
      "name": "ISO 7704 / ISO 9308-1 (Water Quality)",
      "unit": "",
      "description": "International standard for membrane evaluation & testing"
    },
    {
      "name": "USP Chapter 1231 / 61 (Pharmaceutical Water)",
      "unit": "",
      "description": "United States Pharmacopeia bulk water monitoring"
    }
  ],
  "rows": [
    {
      "Standard Specification Parameter": "Nominal Membrane Pore Size",
      "Standard Methods 9222 (Coliform Group)": "0.45 µm",
      "EPA Method 1604 (Coliforms and E. coli)": "0.45 µm",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "0.45 µm (or 0.22 µm)",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "0.45 µm preferred"
    },
    {
      "Standard Specification Parameter": "Standard Sample Volume",
      "Standard Methods 9222 (Coliform Group)": "100 mL",
      "EPA Method 1604 (Coliforms and E. coli)": "100 mL (pre-fill 10 mL if under 20 mL)",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "100 mL",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "100 mL to 200 mL"
    },
    {
      "Standard Specification Parameter": "Standard Membrane Diameter",
      "Standard Methods 9222 (Coliform Group)": "47 mm",
      "EPA Method 1604 (Coliforms and E. coli)": "47 mm",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "47 mm or 50 mm",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "47 mm"
    },
    {
      "Standard Specification Parameter": "Primary Membrane Polymer",
      "Standard Methods 9222 (Coliform Group)": "Mixed Cellulose Esters (MCE)",
      "EPA Method 1604 (Coliforms and E. coli)": "Mixed Cellulose Esters (MCE) or Nitrocellulose",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "Cellulose Esters or Polyethersulfone",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "Mixed Cellulose Esters, Nylon, or PVDF"
    },
    {
      "Standard Specification Parameter": "Operating Vacuum Differential",
      "Standard Methods 9222 (Coliform Group)": "34 to 51 kPa (10 to 15 in. Hg)",
      "EPA Method 1604 (Coliforms and E. coli)": "34 to 51 kPa (max 68 kPa)",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "20 to 50 kPa (",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "30 to 50 kPa ("
    },
    {
      "Standard Specification Parameter": "Approved Funnel Materials",
      "Standard Methods 9222 (Coliform Group)": "Borosilicate glass, Polysulfone, Stainless steel",
      "EPA Method 1604 (Coliforms and E. coli)": "Glass, Polypropylene, Stainless steel",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "Glass, Polysulfone, Polypropylene",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "316L Stainless steel, Polypropylene, Glass"
    },
    {
      "Standard Specification Parameter": "Funnel Wall Rinse Protocol",
      "Standard Methods 9222 (Coliform Group)": "2 to 3 rinses of 20 to 30 mL sterile buffer",
      "EPA Method 1604 (Coliforms and E. coli)": "2 to 3 rinses of 20 to 30 mL sterile buffer",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "2 to 3 rinses of 20 to 30 mL sterile buffer",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "3 rinses of 20 to 30 mL sterile buffer"
    },
    {
      "Standard Specification Parameter": "Microbial Recovery Benchmark",
      "Standard Methods 9222 (Coliform Group)": "Quantitative coliform retention",
      "EPA Method 1604 (Coliforms and E. coli)": "Simultaneous MI chromogenic/fluorogenic recovery",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "Recovery ratio >= 80% (0.80) vs nonselective",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "TAMC enumeration / non-inhibitory recovery"
    },
    {
      "Standard Specification Parameter": "Standard Incubation Parameter",
      "Standard Methods 9222 (Coliform Group)": "35.0 °C ± 0.5 °C for 24 h (m-Endo)",
      "EPA Method 1604 (Coliforms and E. coli)": "35.0 °C ± 0.5 °C for 24 h (MI agar)",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "36.0 °C ± 2.0 °C for 21 to 24 h (CCA)",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "30.0 °C to 35.0 °C for 48 to 72 h (R2A/TSA)"
    },
    {
      "Standard Specification Parameter": "Primary Regulatory Application",
      "Standard Methods 9222 (Coliform Group)": "Clean Water Act & Safe Drinking Water Act compliance",
      "EPA Method 1604 (Coliforms and E. coli)": "Clean Water Act biological monitoring",
      "ISO 7704 / ISO 9308-1 (Water Quality)": "EU Drinking Water Directive & international testing",
      "USP Chapter 1231 / 61 (Pharmaceutical Water)": "cGMP bulk water quality release (WFI / PW)"
    }
  ],
  "sources": [
    {
      "title": "EPA Method 1604 (EPA-821-R-02-024)",
      "url": "https://www.epa.gov/cwa-methods",
      "accessed": "2026-11-15"
    },
    {
      "title": "Standard Methods for the Examination of Water and Wastewater, Section 9222",
      "url": "https://www.standardmethods.org",
      "accessed": "2026-11-15"
    },
    {
      "title": "40 CFR § 141.21 Revised Total Coliform Rule",
      "url": "https://www.ecfr.gov/current/title-40/chapter-I/subchapter-D/part-141",
      "accessed": "2026-11-15"
    },
    {
      "title": "ISO 7704:2023 Water Quality — Performance testing of membrane filters",
      "url": "https://www.iso.org/standard/79848.html",
      "accessed": "2026-11-15"
    },
    {
      "title": "USP General Chapter 1231 Water for Pharmaceutical Purposes",
      "url": "https://www.usp.org",
      "accessed": "2026-11-15"
    }
  ]
}