  "name": "Biophysical Interaction Method Specifications and Operational Parameters",
  "description": "Comparative specifications, sample consumption thresholds, dynamic affinity ranges, and immobilization requirements across solution and surface biophysical binding methods compiled from peer-reviewed literature and instrument technical specifications.",
  "page": "https://casrai.org/guides/microscale-thermophoresis",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-19T13:25:00Z",
  "temporalCoverage": "2010/2026",
  "spatialCoverage": "International",
  "columns": [
    {
      "name": "Biophysical Method",
      "unit": "",
      "description": "Instrumental interaction analysis platform"
    },
    {
      "name": "Primary Readout Mechanism",
      "unit": "",
      "description": "Physical signal tracked during molecular interaction"
    },
    {
      "name": "Sample Environment",
      "unit": "",
      "description": "Solution phase or sensor surface interface"
    },
    {
      "name": "Immobilization Required",
      "unit": "",
      "description": "Whether one binding partner must be covalently or affinity-tethered"
    },
    {
      "name": "Minimum Sample Volume",
      "unit": "microliters",
      "description": "Minimum volume per data point or injection"
    },
    {
      "name": "Typical Dynamic Affinity Range",
      "unit": "molar",
      "description": "Resolvable equilibrium dissociation constant Kd window"
    },
    {
      "name": "Kinetic Rate Resolution",
      "unit": "",
      "description": "Ability to resolve association kon and dissociation koff rate constants"
    },
    {
      "name": "Matrix Tolerance",
      "unit": "",
      "description": "Compatibility with unpurified or crude physiological matrices"
    }
  ],
  "rows": [
    {
      "Biophysical Method": "Microscale Thermophoresis (MST)",
      "Primary Readout Mechanism": "Fluorescence change in microscale thermal gradient",
      "Sample Environment": "Free solution in glass capillary",
      "Immobilization Required": "No",
      "Minimum Sample Volume": 4,
      "Typical Dynamic Affinity Range": "1 pM to 1 mM",
      "Kinetic Rate Resolution": "Limited (primarily equilibrium Kd)",
      "Matrix Tolerance": "High (lysates, serum, plasma)"
    },
    {
      "Biophysical Method": "Surface Plasmon Resonance (SPR)",
      "Primary Readout Mechanism": "Refractive index shift on optical sensor surface",
      "Sample Environment": "Solid surface interface / microfluidic flow",
      "Immobilization Required": "Yes",
      "Minimum Sample Volume": 50,
      "Typical Dynamic Affinity Range": "10 pM to 1 mM",
      "Kinetic Rate Resolution": "High (real-time kon and koff)",
      "Matrix Tolerance": "Low to Moderate (requires clean analyte)"
    },
    {
      "Biophysical Method": "Isothermal Titration Calorimetry (ITC)",
      "Primary Readout Mechanism": "Heat of reaction (enthalpy change) upon titration",
      "Sample Environment": "Free solution in sample cell",
      "Immobilization Required": "No",
      "Minimum Sample Volume": 200,
      "Typical Dynamic Affinity Range": "10 nM to 100 uM",
      "Kinetic Rate Resolution": "No (thermodynamic equilibrium only)",
      "Matrix Tolerance": "Low (susceptible to buffer mismatch heat)"
    },
    {
      "Biophysical Method": "Bio-Layer Interferometry (BLI)",
      "Primary Readout Mechanism": "Optical interference pattern shift at fiber-optic tip",
      "Sample Environment": "Solid surface interface / dip-and-read",
      "Immobilization Required": "Yes",
      "Minimum Sample Volume": 40,
      "Typical Dynamic Affinity Range": "10 pM to 1 mM",
      "Kinetic Rate Resolution": "High (real-time kon and koff)",
      "Matrix Tolerance": "Moderate to High (fluidics-free tip format)"
    },
    {
      "Biophysical Method": "Mass Photometry (MP)",
      "Primary Readout Mechanism": "Interferometric light scattering of single particles",
      "Sample Environment": "Landing on glass coverslip interface",
      "Immobilization Required": "No",
      "Minimum Sample Volume": 10,
      "Typical Dynamic Affinity Range": "10 nM to 1 uM",
      "Kinetic Rate Resolution": "Limited (single-particle equilibrium mass)",
      "Matrix Tolerance": "Low (requires low background scattering)"
    }
  ],
  "sources": [
    {
      "title": "Wienken et al. (2010) Nature Communications - Protein-binding assays in biological liquids using microscale thermophoresis",
      "url": "https://www.nature.com/articles/ncomms1093",
      "accessed": "2026-11-15"
    },
    {
      "title": "Jerabek-Willemsen et al. (2014) Journal of Molecular Structure - MicroScale Thermophoresis: Interaction analysis and beyond",
      "url": "https://doi.org/10.1016/j.molstruc.2014.03.009",
      "accessed": "2026-11-15"
    },
    {
      "title": "Seidel et al. (2013) Methods - Microscale thermophoresis quantifies biomolecular interactions under previously challenging conditions",
      "url": "https://doi.org/10.1016/j.ymeth.2012.12.005",
      "accessed": "2026-11-15"
    },
    {
      "title": "Sparks et al. (2020) JoVE - Use of Microscale Thermophoresis to Measure Protein-Lipid Interactions",
      "url": "https://www.jove.com/t/60607",
      "accessed": "2026-11-15"
    }
  ]
}