  "name": "NanoDrop 2000 protein measurement applications: quantitative ranges and intended samples",
  "description": "Compares the protein quantitation applications of the Thermo NanoDrop 2000/2000c software on lower and upper limits, typical reproducibility and intended sample type, compiled from the measurement-ranges table and application chapters of the Thermo user manual V1.0.",
  "page": "https://casrai.org/guides/nanodrop-2000",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-18T20:47:47Z",
  "temporalCoverage": "Manual revision March 2009",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Application",
      "unit": "",
      "description": ""
    },
    {
      "name": "Lower limit",
      "unit": "mg/mL",
      "description": ""
    },
    {
      "name": "Upper limit",
      "unit": "mg/mL",
      "description": ""
    },
    {
      "name": "Typical reproducibility",
      "unit": "",
      "description": "As printed in the manual"
    },
    {
      "name": "Standard curve needed",
      "unit": "",
      "description": ""
    },
    {
      "name": "Intended sample (manual)",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Application": "Protein A280 (pedestal)",
      "Lower limit": 0.1,
      "Upper limit": 400,
      "Typical reproducibility": "0.10-10 mg/mL: ±0.10 mg/mL; above 10 mg/mL: ±2%",
      "Standard curve needed": "No",
      "Intended sample (manual)": "Purified proteins with Trp, Tyr or Cys-Cys; limits stated as purified BSA"
    },
    {
      "Application": "Proteins and Labels (pedestal)",
      "Lower limit": 0.1,
      "Upper limit": 20,
      "Typical reproducibility": "0.10-10 mg/mL: ±0.10 mg/mL",
      "Standard curve needed": "No",
      "Intended sample (manual)": "Protein A280 plus fluorescent dye concentration for protein array conjugates"
    },
    {
      "Application": "BCA (20:1 reagent to sample)",
      "Lower limit": 0.2,
      "Upper limit": 8,
      "Typical reproducibility": "2% over entire range",
      "Standard curve needed": "Yes",
      "Intended sample (manual)": "Colorimetric assay for uncharacterised solutions and cell lysates"
    },
    {
      "Application": "Modified Lowry",
      "Lower limit": 0.2,
      "Upper limit": 4,
      "Typical reproducibility": "2% over entire range",
      "Standard curve needed": "Yes",
      "Intended sample (manual)": "Colorimetric assay for uncharacterised solutions and cell lysates"
    },
    {
      "Application": "Bradford (50:1 reagent to sample)",
      "Lower limit": 0.1,
      "Upper limit": 8,
      "Typical reproducibility": "100-500 µg/mL: ±25 µg/mL; 500-8000 µg/mL: ±5%",
      "Standard curve needed": "Yes",
      "Intended sample (manual)": "Colorimetric assay for uncharacterised solutions and cell lysates"
    }
  ],
  "sources": [
    {
      "title": "Thermo Fisher Scientific: NanoDrop 2000/2000c Spectrophotometer V1.0 User Manual",
      "url": "https://assets.thermofisher.com/TFS-Assets/CAD/manuals/NanoDrop-2000-User-Manual-EN.pdf",
      "accessed": "2026-11-18"
    }
  ]
}