  "name": "Benchtop Nanoparticle Characterization Platform Specifications and Operating Limits",
  "description": "An original cross-platform benchmark comparing measurement principles, size detection limits, particle concentration ranges, sample volume requirements, and optical configurations across commercial nanoparticle tracking analysis instruments and orthogonal characterization platforms. Compiled from manufacturer technical specifications, ISO 19430:2024 standards documentation, and peer-reviewed characterization protocols.",
  "page": "https://casrai.org/guides/nanoparticle-tracking-analysis",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-14T05:52:03Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "International",
  "columns": [
    {
      "name": "Instrument Platform",
      "unit": "",
      "description": "Instrument model and primary manufacturer"
    },
    {
      "name": "Measurement Principle",
      "unit": "",
      "description": "Core physical detection and sizing modality"
    },
    {
      "name": "Lower Size Limit",
      "unit": "nm",
      "description": "Lower hydrodynamic or physical diameter detection limit"
    },
    {
      "name": "Upper Size Limit",
      "unit": "nm",
      "description": "Upper hydrodynamic or physical diameter detection limit"
    },
    {
      "name": "Min Concentration",
      "unit": "particles/mL",
      "description": "Minimum detectable particle concentration in liquid suspension"
    },
    {
      "name": "Max Concentration",
      "unit": "particles/mL",
      "description": "Maximum linear particle concentration before coincidence loss"
    },
    {
      "name": "Min Sample Volume",
      "unit": "µL",
      "description": "Minimum liquid sample volume required for measurement cell"
    },
    {
      "name": "Fluorescence Capability",
      "unit": "",
      "description": "Availability of fluorescent biomarker detection channels"
    }
  ],
  "rows": [
    {
      "Instrument Platform": "Malvern Panalytical NanoSight NS300",
      "Measurement Principle": "Nanoparticle Tracking Analysis (Light Scatter / NTA)",
      "Lower Size Limit": 10,
      "Upper Size Limit": 1000,
      "Min Concentration": 1000000,
      "Max Concentration": 1000000000,
      "Min Sample Volume": 300,
      "Fluorescence Capability": "Optional 6-position motorized filter wheel"
    },
    {
      "Instrument Platform": "Malvern Panalytical NanoSight Pro",
      "Measurement Principle": "Nanoparticle Tracking Analysis (Automated Flow NTA)",
      "Lower Size Limit": 10,
      "Upper Size Limit": 1000,
      "Min Concentration": 1000000,
      "Max Concentration": 1000000000,
      "Min Sample Volume": 250,
      "Fluorescence Capability": "Integrated dual-channel fluorescence"
    },
    {
      "Instrument Platform": "Particle Metrix ZetaView PMX-120",
      "Measurement Principle": "Multi-Parameter NTA & Micro-Electrophoresis",
      "Lower Size Limit": 10,
      "Upper Size Limit": 2000,
      "Min Concentration": 100000,
      "Max Concentration": 10000000000,
      "Min Sample Volume": 500,
      "Fluorescence Capability": "Multi-laser fluorescence and colocalization"
    },
    {
      "Instrument Platform": "Malvern Panalytical Zetasizer Ultra",
      "Measurement Principle": "Dynamic Light Scattering (Multi-Angle DLS / MADLS)",
      "Lower Size Limit": 0.3,
      "Upper Size Limit": 10000,
      "Min Concentration": 100000,
      "Max Concentration": 1000000000000,
      "Min Sample Volume": 20,
      "Fluorescence Capability": "Not applicable (ensemble scatter)"
    },
    {
      "Instrument Platform": "Izon Science Exoid",
      "Measurement Principle": "Tunable Resistive Pulse Sensing (TRPS)",
      "Lower Size Limit": 40,
      "Upper Size Limit": 2000,
      "Min Concentration": 10000000,
      "Max Concentration": 100000000000,
      "Min Sample Volume": 35,
      "Fluorescence Capability": "Not applicable (electrical pore transit)"
    }
  ],
  "sources": [
    {
      "title": "Malvern Panalytical NanoSight Range Technical Specifications",
      "url": "https://www.malvernpanalytical.com/en/products/product-range/nanosight-range",
      "accessed": "2026-11-14"
    },
    {
      "title": "Particle Metrix ZetaView Technical Documentation",
      "url": "https://www.particle-metrix.com/products/zetaview/",
      "accessed": "2026-11-14"
    },
    {
      "title": "ISO 19430:2024 Particle tracking analysis (PTA) method",
      "url": "https://www.iso.org/standard/83431.html",
      "accessed": "2026-11-14"
    }
  ]
}