  "name": "NVLAP evidence documents and their limits",
  "description": "Original crosswalk compiled from NIST Handbook 150:2020 update 1, the NVLAP application instructions and Calibration LAP guidance. It distinguishes what each document establishes from claims it cannot establish by itself; source clauses identify the basis for each row.",
  "page": "https://casrai.org/guides/nvlap",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-26T15:25:00Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Evidence object",
      "unit": "",
      "description": ""
    },
    {
      "name": "What it establishes",
      "unit": "",
      "description": ""
    },
    {
      "name": "Limit of the evidence",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source and clause",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Evidence object": "Application-stage Lab Code",
      "What it establishes": "NVLAP has assigned an identifier to the applicant laboratory.",
      "Limit of the evidence": "A code alone does not establish granted accreditation; applicants cannot use the accreditation symbol.",
      "Source and clause": "NIST Handbook 150, 3.1.5 and Annex A.2(b); Apply for NVLAP Accreditation"
    },
    {
      "Evidence object": "Certificate of Accreditation",
      "What it establishes": "Identifies the accredited laboratory, Authorized Representative, effective and expiration dates, and Lab Code.",
      "Limit of the evidence": "Technical coverage must be read with the accompanying Scope of Accreditation.",
      "Source and clause": "NIST Handbook 150, 1.3.5 and 3.6.2-3.6.4"
    },
    {
      "Evidence object": "Testing Scope of Accreditation",
      "What it establishes": "Identifies materials or products tested, methods used and tests performed.",
      "Limit of the evidence": "Does not establish accreditation for tests outside the identified scope.",
      "Source and clause": "NIST Handbook 150, 3.6.4(c) and Annex A.5"
    },
    {
      "Evidence object": "Calibration Scope of Accreditation",
      "What it establishes": "Identifies quantity or reference material, procedure and instrument/material type, range, additional parameters and uncertainty.",
      "Limit of the evidence": "A common NVLAP name does not imply one prescribed procedure or minimum uncertainty for every laboratory.",
      "Source and clause": "NIST Handbook 150, 3.6.4(b); NIST Calibration LAP"
    },
    {
      "Evidence object": "Accreditation reference on a laboratory report",
      "What it establishes": "Identifies use of NVLAP accreditation under reporting conditions; out-of-scope and subcontracted data must be distinguished.",
      "Limit of the evidence": "Does not establish government product certification, approval or endorsement, or accredited status for every reported result.",
      "Source and clause": "NIST Handbook 150, Annex A.2(f), A.5 and A.7; Referencing NVLAP Accreditation"
    }
  ],
  "sources": [
    {
      "title": "NIST Handbook 150:2020 update 1",
      "url": "https://nvlpubs.nist.gov/nistpubs/hb/2020/NIST.HB.150-2020-upd1.pdf",
      "accessed": "2026-11-15"
    },
    {
      "title": "NIST: Apply for NVLAP Accreditation",
      "url": "https://www.nist.gov/nvlap/apply-nvlap-accreditation",
      "accessed": "2026-11-15"
    },
    {
      "title": "NIST: Calibration LAP",
      "url": "https://www.nist.gov/nvlap/calibration-lap",
      "accessed": "2026-11-15"
    },
    {
      "title": "NIST: Referencing NVLAP Accreditation",
      "url": "https://www.nist.gov/nvlap/referencing-nvlap-accreditation-and-use-ilac-combined-mark",
      "accessed": "2026-11-15"
    }
  ]
}