  "name": "Shear stress devices compared: ranges and equations for parallel plate, cone-and-plate, microfluidic and radial chambers",
  "description": "Compares in vitro fluid shear devices by the wall shear stress range and governing equation each source states, compiled from a peer-reviewed design review and the standard description of the parallel-plate flow chamber. Ranges are those reported for each device type, not guarantees for any particular product.",
  "page": "https://casrai.org/guides/parallel-plate-flow-chamber",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T03:39:30Z",
  "temporalCoverage": null,
  "spatialCoverage": "Laboratory devices, not location specific",
  "columns": [
    {
      "name": "Device",
      "unit": "",
      "description": ""
    },
    {
      "name": "Minimum shear",
      "unit": "dyn/cm2",
      "description": ""
    },
    {
      "name": "Maximum shear",
      "unit": "dyn/cm2",
      "description": ""
    },
    {
      "name": "Shear equation",
      "unit": "",
      "description": ""
    },
    {
      "name": "Limit stated by source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Device": "Parallel plate flow chamber (original design)",
      "Minimum shear": 0.01,
      "Maximum shear": 30,
      "Shear equation": "Wall shear = 6 x viscosity x flow rate / (width x height^2)",
      "Limit stated by source": "Assumes Newtonian, incompressible, laminar flow; cell shear approximates wall shear"
    },
    {
      "Device": "Radial impinging flow chamber",
      "Minimum shear": 10,
      "Maximum shear": 200,
      "Shear equation": "Shear roughly 3 x viscosity x flow rate / (pi x radius x height^2)",
      "Limit stated by source": "Inertial effects kept below 5% of viscous forces; laminar for Re below 2000 at radius above 0.05 mm"
    },
    {
      "Device": "Cone-and-plate device (laminar)",
      "Minimum shear": 0.01,
      "Maximum shear": 100,
      "Shear equation": "Shear = viscosity x angular velocity / cone angle (radians)",
      "Limit stated by source": "Evaporation in long studies; no live imaging unless transparent"
    },
    {
      "Device": "Microfluidic channel",
      "Minimum shear": 0.7,
      "Maximum shear": 130,
      "Shear equation": "Shear = 12 x viscosity x flow rate / (width x height^2)",
      "Limit stated by source": "PDMS adsorbs hydrophobic molecules; bubbles disrupt flow"
    },
    {
      "Device": "Physiological reference: straight arteries (pulsatile)",
      "Minimum shear": 10,
      "Maximum shear": 60,
      "Shear equation": "Not applicable",
      "Limit stated by source": "Reference range for unidirectional pulsatile shear, not a device"
    }
  ],
  "sources": [
    {
      "title": "Wikipedia: Parallel-plate flow chamber",
      "url": "https://en.wikipedia.org/wiki/Parallel-plate_flow_chamber",
      "accessed": "2026-11-20"
    },
    {
      "title": "Fallon, Mathews, Hinds: In Vitro Flow Chamber Design for the Study of Endothelial Cell (Patho)Physiology",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC8628846",
      "accessed": "2026-11-20"
    }
  ]
}