  "name": "Factors that change a microscope's point spread function, with the source for each",
  "description": "Compiled from Nikon MicroscopyU, the Evident (Olympus) microscopy resource, and two peer-reviewed methods papers; each row names a factor, its effect on the PSF and the source that states it.",
  "page": "https://casrai.org/guides/point-spread-function",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T14:45:00Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Factor",
      "unit": "",
      "description": ""
    },
    {
      "name": "Effect on the PSF or resolution",
      "unit": "",
      "description": ""
    },
    {
      "name": "Source",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Factor": "Numerical aperture",
      "Effect on the PSF or resolution": "Higher NA improves resolution; magnification is not a term in the resolution formulas",
      "Source": "Nikon MicroscopyU, Resolution"
    },
    {
      "Factor": "Wavelength",
      "Effect on the PSF or resolution": "Shorter wavelength improves resolution",
      "Source": "Nikon MicroscopyU, Resolution"
    },
    {
      "Factor": "Refractive-index mismatch between immersion and mounting medium",
      "Effect on the PSF or resolution": "Causes spherical aberration, which stretches the PSF axially and reduces resolution and signal",
      "Source": "Evident (Olympus) Microscopy Resource, Introduction to Deconvolution"
    },
    {
      "Factor": "Objective, confocal scanning components or relay optics faults",
      "Effect on the PSF or resolution": "PSF geometry measured with fluorescent microspheres indicates problems with these components",
      "Source": "Cole, Jinadasa and Brown, Nature Protocols 2011"
    },
    {
      "Factor": "Depth into the sample",
      "Effect on the PSF or resolution": "Refractive-index mismatch can be quantified as a function of depth by measuring the 3D PSF at multiple positions and depths",
      "Source": "van der Wee et al., Communications Biology 2021"
    }
  ],
  "sources": [
    {
      "title": "Nikon MicroscopyU: Resolution",
      "url": "https://www.microscopyu.com/microscopy-basics/resolution",
      "accessed": "2026-11-14"
    },
    {
      "title": "Evident (Olympus) Microscopy Resource: Introduction to Deconvolution",
      "url": "https://evidentscientific.com/en/microscope-resource/knowledge-hub/digital-imaging/deconvolution/deconintro",
      "accessed": "2026-11-14"
    },
    {
      "title": "Cole RW, Jinadasa T, Brown CM. Nature Protocols 6, 1929-1941 (2011)",
      "url": "https://doi.org/10.1038/nprot.2011.407",
      "accessed": "2026-11-14"
    },
    {
      "title": "van der Wee et al. Communications Biology (2021)",
      "url": "https://doi.org/10.1038/s42003-021-02432-3",
      "accessed": "2026-11-14"
    }
  ]
}