  "name": "Hemocytometer Manual Counting Protocols and Calculations Across Cell Populations",
  "description": "Comparative technical parameters for manual cell enumeration on an Improved Neubauer counting chamber across erythrocytes, leukocytes, thrombocytes, and cultured cells, compiled from 21 CFR 864.6160 and laboratory protocol standards.",
  "page": "https://casrai.org/guides/rbc-count-hemocytometer",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-22T19:50:03Z",
  "temporalCoverage": "2026",
  "spatialCoverage": "Global",
  "columns": [
    {
      "name": "Cell Population",
      "unit": "",
      "description": "Target cellular element enumerated in biological fluid"
    },
    {
      "name": "Standard Diluent",
      "unit": "",
      "description": "Chemical composition of diluting fluid"
    },
    {
      "name": "Dilution Ratio",
      "unit": "",
      "description": "Standard blood or cell suspension dilution factor"
    },
    {
      "name": "Grid Squares Counted",
      "unit": "",
      "description": "Designated squares evaluated on Improved Neubauer grid"
    },
    {
      "name": "Counting Volume",
      "unit": "μL",
      "description": "Total chamber volume represented by counted squares"
    },
    {
      "name": "Calculation Multiplier",
      "unit": "factor",
      "description": "Volume and dilution multiplier to yield cells per μL"
    },
    {
      "name": "Objective Lens",
      "unit": "magnification",
      "description": "Standard microscope objective used for enumeration"
    }
  ],
  "rows": [
    {
      "Cell Population": "Erythrocytes (RBC)",
      "Standard Diluent": "Hayem's solution or Dacie's formol-citrate",
      "Dilution Ratio": "1:200",
      "Grid Squares Counted": "5 medium central squares (80 micro-squares)",
      "Counting Volume": 0.02,
      "Calculation Multiplier": 10000,
      "Objective Lens": "40x"
    },
    {
      "Cell Population": "Leukocytes (WBC)",
      "Standard Diluent": "Turk's solution (acetic acid with gentian violet)",
      "Dilution Ratio": "1:20",
      "Grid Squares Counted": "4 large corner primary squares",
      "Counting Volume": 0.4,
      "Calculation Multiplier": 50,
      "Objective Lens": "10x"
    },
    {
      "Cell Population": "Thrombocytes (Platelets)",
      "Standard Diluent": "1% Ammonium oxalate solution",
      "Dilution Ratio": "1:100",
      "Grid Squares Counted": "5 medium central squares (80 micro-squares)",
      "Counting Volume": 0.02,
      "Calculation Multiplier": 5000,
      "Objective Lens": "40x phase contrast"
    },
    {
      "Cell Population": "Cultured Mammalian Cells",
      "Standard Diluent": "0.4% Trypan blue vital exclusion dye in PBS",
      "Dilution Ratio": "1:2",
      "Grid Squares Counted": "4 large corner primary squares",
      "Counting Volume": 0.4,
      "Calculation Multiplier": 5000,
      "Objective Lens": "10x or 20x"
    }
  ],
  "sources": [
    {
      "title": "21 CFR § 864.6160 - Manual blood cell counting device",
      "url": "https://www.law.cornell.edu/cfr/text/21/864.6160",
      "accessed": "2026-11-22"
    },
    {
      "title": "Thermo Fisher Scientific: Manually Counting Cells in a Hemocytomer",
      "url": "https://www.thermofisher.com/us/en/home/references/gibco-cell-culture-basics/cell-culture-protocols/counting-cells-in-a-hemacytometer.html",
      "accessed": "2026-11-22"
    },
    {
      "title": "PubMed Central / Biochem Mol Biol Educ: Hemocytometer Proficiency and Calculations (PMC11792770)",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC11792770/",
      "accessed": "2026-11-22"
    }
  ]
}