  "name": "Requirements and guidance for new reagent lot checks across CLIA, CAP and CLSI EP26",
  "description": "Compares what each US source says about checking a new reagent lot: scope, trigger, check material, acceptance criterion and documentation. Compiled from the CLIA regulation text, two CAP TODAY Q&As by CAP staff, and the CLSI EP26 abstract and scope.",
  "page": "https://casrai.org/guides/reagent-lot-to-lot-verification",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-14T11:52:55Z",
  "temporalCoverage": null,
  "spatialCoverage": "United States",
  "columns": [
    {
      "name": "Source",
      "unit": "",
      "description": ""
    },
    {
      "name": "Type",
      "unit": "",
      "description": ""
    },
    {
      "name": "Trigger covered",
      "unit": "",
      "description": ""
    },
    {
      "name": "Check material named",
      "unit": "",
      "description": ""
    },
    {
      "name": "Who sets acceptance criterion",
      "unit": "",
      "description": ""
    },
    {
      "name": "Documentation stated",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Source": "42 CFR 493.1256(e)(1)",
      "Type": "Federal regulation (nonwaived testing)",
      "Trigger covered": "Each batch, lot and shipment of reagents, disks, stains, antisera, when prepared or opened",
      "Check material named": "Positive and negative reactivity checks",
      "Who sets acceptance criterion": "Not specified in the paragraph; (f) requires control results to meet laboratory and manufacturer criteria",
      "Documentation stated": "Yes, (g) requires documentation of all control procedures"
    },
    {
      "Source": "42 CFR 493.1256(d)(6)",
      "Type": "Federal regulation (nonwaived testing)",
      "Trigger covered": "Complete change of reagents",
      "Check material named": "Control testing before patient testing resumes",
      "Who sets acceptance criterion": "Laboratory and manufacturer criteria under (f)",
      "Documentation stated": "Yes, (g)"
    },
    {
      "Source": "CAP COM.30450 (CAP TODAY Q&A, qa-column-0116)",
      "Type": "Accreditation checklist requirement",
      "Trigger covered": "New reagent lots and shipments",
      "Check material named": "Old lot or suitable reference materials; patient specimens recommended when possible",
      "Who sets acceptance criterion": "Each laboratory defines it (e.g. tolerance limits)",
      "Documentation stated": "Criteria documented in laboratory policy"
    },
    {
      "Source": "CAP COM.30450 (CAP TODAY Q&A, qa-column-0322, March 2022)",
      "Type": "Accreditation checklist requirement",
      "Trigger covered": "Reagent lot change",
      "Check material named": "Patient samples (best practice, no matrix effect); manufacturer QC with lot-specific targets; PT materials with peer-group means",
      "Who sets acceptance criterion": "Laboratory; recalibration or calibration verification required unless lots shown not to affect accuracy",
      "Documentation stated": "Not stated in the Q&A"
    },
    {
      "Source": "CLSI EP26, 2nd ed. (Apr 2022)",
      "Type": "Consensus guideline",
      "Trigger covered": "New reagent lot replacing current lot, quantitative or continuous-scale results",
      "Check material named": "Patient samples",
      "Who sets acceptance criterion": "Statistical protocol set up once per analyte, then applied to each lot",
      "Documentation stated": "Not stated in the abstract"
    }
  ],
  "sources": [
    {
      "title": "42 CFR 493.1256 Standard: Control procedures",
      "url": "https://www.law.cornell.edu/cfr/text/42/493.1256",
      "accessed": "2026-11-14"
    },
    {
      "title": "CAP TODAY Q&A column (qa-column-0116)",
      "url": "https://www.captodayonline.com/qa-column-0116/2/",
      "accessed": "2026-11-14"
    },
    {
      "title": "CAP TODAY Q&A column (qa-column-0322), March 2022",
      "url": "https://www.captodayonline.com/qa-column-0322/2/",
      "accessed": "2026-11-14"
    },
    {
      "title": "CLSI EP26: User Evaluation of Acceptability of a Reagent Lot Change",
      "url": "https://clsi.org/standards/products/method-evaluation/documents/ep26",
      "accessed": "2026-11-14"
    }
  ]
}