  "name": "SEM sample preparation steps as described by three university core facilities",
  "description": "Side-by-side comparison of the fixation, dehydration, drying, mounting and coating stages for scanning electron microscopy as described in the published protocol or service pages of Utah State University, the University of Gothenburg and the University of Cambridge, compiled by CASRAI from those pages.",
  "page": "https://casrai.org/guides/sem-sample-preparation",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-20T16:11:31Z",
  "temporalCoverage": "2026",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Stage",
      "unit": "",
      "description": ""
    },
    {
      "name": "Utah State University",
      "unit": "",
      "description": ""
    },
    {
      "name": "University of Gothenburg",
      "unit": "",
      "description": ""
    },
    {
      "name": "University of Cambridge",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Stage": "Primary fixation",
      "Utah State University": "2% buffered glutaraldehyde overnight, 10-20x sample volume; rinse 3 x 5 min in 0.1 M HEPES",
      "University of Gothenburg": "Aldehyde fixation crosslinks proteins",
      "University of Cambridge": "Formaldehyde or glutaraldehyde for biological specimens (before critical point drying)"
    },
    {
      "Stage": "Secondary fixation",
      "Utah State University": "1-2% osmium tetroxide for 1 h for tissue; skipped for non-tissue",
      "University of Gothenburg": "Osmium fixes lipids; black precipitate raises conductivity",
      "University of Cambridge": "Osmium tetroxide listed among chemical fixatives; some materials processed unfixed"
    },
    {
      "Stage": "Dehydration",
      "Utah State University": "Ethanol 50, 70, 95% (2 x 10 min each), 100% (3 x 15 min)",
      "University of Gothenburg": "Ethanol or acetone, applied gradually to avoid shrinkage",
      "University of Cambridge": "Commonly ethanol, after washing in deionised water"
    },
    {
      "Stage": "Drying",
      "Utah State University": "Critical point drying, or HMDS series (2:1, 1:1, 1:2 ethanol:HMDS, 15 min each; HMDS 3 x 15 min; last HMDS evaporated overnight in fume hood)",
      "University of Gothenburg": "HMDS (about 3 min incubation then evaporation) or liquid CO2 critical point drying",
      "University of Cambridge": "Critical point drying with liquid CO2; or plunge-freezing in liquid ethane then freeze-drying"
    },
    {
      "Stage": "Mounting",
      "Utah State University": "Specimen stub",
      "University of Gothenburg": "Metal stub with sticky carbon disc; silver glue optional",
      "University of Cambridge": "Mounting after drying (method not specified)"
    },
    {
      "Stage": "Coating",
      "Utah State University": "Sputter coat, 10 nm Au/Pd",
      "University of Gothenburg": "Sputter coat, most commonly gold, about 10 nm",
      "University of Cambridge": "Sputter and carbon coating offered"
    }
  ],
  "sources": [
    {
      "title": "Biological Sample Fixation for SEM (Utah State University Microscopy Core Facility)",
      "url": "https://research.usu.edu/microscopy/files/SEM-sample-prep4.pdf",
      "accessed": "2026-11-20"
    },
    {
      "title": "SEM sample preparation techniques (University of Gothenburg Centre for Cellular Imaging)",
      "url": "https://www.gu.se/en/core-facilities/sem-sample-preparation-techniques",
      "accessed": "2026-11-20"
    },
    {
      "title": "Specimen Preparation (University of Cambridge Department of Biology Electron Microscopy)",
      "url": "https://www.bio.cam.ac.uk/node/2903574",
      "accessed": "2026-11-20"
    }
  ]
}