  "name": "Analytical Specifications and Performance Benchmarks of Ultra-Sensitive Immunoassay Platforms",
  "description": "Comparison of analytical limits of detection, linear dynamic range, sample volume requirements, multiplexing capacity, and assay run times across digital ELISA (Simoa), conventional sandwich ELISA, electrochemiluminescence (MSD), single molecule counting (SMC), and suspension bead arrays (Luminex). Compiled from peer-reviewed analytical validation studies and manufacturer technical specifications.",
  "page": "https://casrai.org/guides/simoa-assay",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-22T10:55:00Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Technology",
      "unit": "",
      "description": "Immunoassay technology or analytical platform architecture"
    },
    {
      "name": "Primary Readout Mode",
      "unit": "",
      "description": "Physical signal transduction and quantification mechanism"
    },
    {
      "name": "Typical LOD Lower Bound",
      "unit": "pg/mL",
      "description": "Representative lower limit of detection for high-affinity assays"
    },
    {
      "name": "Dynamic Range",
      "unit": "orders of magnitude",
      "description": "Linear quantitative measurement range across sample concentrations"
    },
    {
      "name": "Required Sample Volume",
      "unit": "µL",
      "description": "Minimum biological sample volume required per test replicate"
    },
    {
      "name": "Multiplexing Capacity",
      "unit": "targets per well",
      "description": "Maximum number of simultaneous analyte targets quantified per well"
    },
    {
      "name": "Assay Incubation Time",
      "unit": "hours",
      "description": "Typical total hands-on and automated incubation time per run"
    }
  ],
  "rows": [
    {
      "Technology": "Simoa Digital ELISA",
      "Primary Readout Mode": "Digital bead counting and analog fluorescence",
      "Typical LOD Lower Bound": 0.005,
      "Dynamic Range": 4.5,
      "Required Sample Volume": 15,
      "Multiplexing Capacity": 6,
      "Assay Incubation Time": 2.5
    },
    {
      "Technology": "Conventional Sandwich ELISA",
      "Primary Readout Mode": "Bulk analog spectrophotometric absorbance",
      "Typical LOD Lower Bound": 5,
      "Dynamic Range": 2.5,
      "Required Sample Volume": 100,
      "Multiplexing Capacity": 1,
      "Assay Incubation Time": 4
    },
    {
      "Technology": "Meso Scale Discovery ECL",
      "Primary Readout Mode": "Electrochemiluminescence on carbon electrode spots",
      "Typical LOD Lower Bound": 0.5,
      "Dynamic Range": 4,
      "Required Sample Volume": 25,
      "Multiplexing Capacity": 10,
      "Assay Incubation Time": 3
    },
    {
      "Technology": "Single Molecule Counting",
      "Primary Readout Mode": "Microfluidic capillary single-photon counting",
      "Typical LOD Lower Bound": 0.1,
      "Dynamic Range": 4,
      "Required Sample Volume": 50,
      "Multiplexing Capacity": 1,
      "Assay Incubation Time": 4
    },
    {
      "Technology": "Luminex xMAP Suspension Array",
      "Primary Readout Mode": "Fluorescent bead flow cytometry or CCD imaging",
      "Typical LOD Lower Bound": 2,
      "Dynamic Range": 3.5,
      "Required Sample Volume": 25,
      "Multiplexing Capacity": 50,
      "Assay Incubation Time": 3.5
    }
  ],
  "sources": [
    {
      "title": "The Simoa HD-1 Analyzer: A Novel Fully Automated Digital Immunoassay Analyzer",
      "url": "https://doi.org/10.1177/2211068215589580",
      "accessed": "2026-11-14"
    },
    {
      "title": "Single-molecule enzyme-linked immunosorbent assay detects serum proteins at subfemtomolar concentrations",
      "url": "https://doi.org/10.1038/nbt.1641",
      "accessed": "2026-11-14"
    },
    {
      "title": "Comparison of digital ELISA and automated microfluidic immunoassay platforms for plasma neurofilament light chain",
      "url": "https://doi.org/10.1515/cclm-2019-1032",
      "accessed": "2026-11-14"
    },
    {
      "title": "Single-molecule arrays for ultrasensitive detection in clinical diagnostics",
      "url": "https://doi.org/10.1146/annurev-anchem-061318-115342",
      "accessed": "2026-11-14"
    }
  ]
}