  "name": "Commercial Allergen-Specific IgE Immunoassay Platforms and Analytical Specifications",
  "description": "Comparative analytical parameters of major in vitro allergen-specific IgE testing platforms compiled from CLSI I/LA20 guidelines, FDA 510(k) classification records under 21 CFR 866.5750, manufacturer assay specifications, and WHO international standardization protocols. Evaluates assay chemistry, solid-phase matrix, specimen volume requirements, analytical limits of quantitation, and calibration traceability.",
  "page": "https://casrai.org/guides/specific-ige-testing",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-26T12:10:00Z",
  "temporalCoverage": "2024/2026",
  "spatialCoverage": "Global / United States",
  "columns": [
    {
      "name": "Assay Platform",
      "unit": "",
      "description": "Analyzer system and commercial manufacturer"
    },
    {
      "name": "Analytical Technology",
      "unit": "",
      "description": "Core immunoassay detection methodology"
    },
    {
      "name": "Solid-Phase Substrate",
      "unit": "",
      "description": "Antigen-coupling solid-phase matrix"
    },
    {
      "name": "Sample Volume",
      "unit": "µL",
      "description": "Patient serum or plasma volume required per test"
    },
    {
      "name": "Lower Measuring Limit",
      "unit": "kUA/L",
      "description": "Analytical lower limit of quantitation (LoQ)"
    },
    {
      "name": "Upper Measuring Limit",
      "unit": "kUA/L",
      "description": "Direct measuring range upper boundary without dilution"
    },
    {
      "name": "Calibration Reference",
      "unit": "",
      "description": "International standard calibration anchor"
    },
    {
      "name": "Regulatory Status",
      "unit": "",
      "description": "FDA classification or international clearance"
    }
  ],
  "rows": [
    {
      "Assay Platform": "Phadia 250 / ImmunoCAP (Thermo Fisher)",
      "Analytical Technology": "Fluorescence Enzyme Immunoassay (FEIA)",
      "Solid-Phase Substrate": "Hydrophilic cellulose polymer solid-phase sponge",
      "Sample Volume": 40,
      "Lower Measuring Limit": 0.1,
      "Upper Measuring Limit": 100,
      "Calibration Reference": "WHO 3rd International Standard 11/234",
      "Regulatory Status": "21 CFR 866.5750 (Class II, Product Code DHB)"
    },
    {
      "Assay Platform": "IMMULITE 2000 3gAllergy (Siemens Healthineers)",
      "Analytical Technology": "Chemiluminescent Enzyme Immunoassay (CLIA)",
      "Solid-Phase Substrate": "Polystyrene bead / Liquid allergen biotin-streptavidin",
      "Sample Volume": 25,
      "Lower Measuring Limit": 0.1,
      "Upper Measuring Limit": 100,
      "Calibration Reference": "WHO 3rd International Standard 11/234",
      "Regulatory Status": "21 CFR 866.5750 (Class II, Product Code DHB)"
    },
    {
      "Assay Platform": "NOVEOS System (HYCOR Biomedical)",
      "Analytical Technology": "Chemiluminescent Microparticle Immunoassay (CLIA)",
      "Solid-Phase Substrate": "Fluorescent paramagnetic microparticles",
      "Sample Volume": 4,
      "Lower Measuring Limit": 0.1,
      "Upper Measuring Limit": 100,
      "Calibration Reference": "WHO 3rd International Standard 11/234",
      "Regulatory Status": "21 CFR 866.5750 (Class II, Product Code DHB)"
    },
    {
      "Assay Platform": "ImmunoCAP ISAC 112 (Thermo Fisher)",
      "Analytical Technology": "Multiplex Fluorometric Microarray Chip",
      "Solid-Phase Substrate": "Chemically activated polymer slide with immobilized native/recombinant proteins",
      "Sample Volume": 30,
      "Lower Measuring Limit": 0.3,
      "Upper Measuring Limit": 100,
      "Calibration Reference": "Internal ISAC Standardized Units (ISU-E)",
      "Regulatory Status": "CE-IVD / 21 CFR 866.5750"
    },
    {
      "Assay Platform": "ALEX2 Allergy Explorer (MacroArray Diagnostics)",
      "Analytical Technology": "Multiplex Colorimetric / Nanoparticle Microarray",
      "Solid-Phase Substrate": "Nitrocellulose membrane array chip with integrated CCD inhibitor",
      "Sample Volume": 100,
      "Lower Measuring Limit": 0.15,
      "Upper Measuring Limit": 50,
      "Calibration Reference": "WHO 3rd International Standard 11/234 (kUA/L)",
      "Regulatory Status": "CE-IVD / FDA 510(k) cleared components"
    }
  ],
  "sources": [
    {
      "title": "CLSI Guideline I/LA20: Analytical Performance Characteristics, Quality Assurance, and Clinical Utility of Immunological Assays for Human Immunoglobulin E Antibodies",
      "url": "https://clsi.org/standards/products/immunology-and-ligand-assay/documents/ila20/",
      "accessed": "2026-11-14"
    },
    {
      "title": "FDA 21 CFR 866.5750 Radioallergosorbent (RAST) Immunological Test System Classification",
      "url": "https://www.law.cornell.edu/cfr/text/21/866.5750",
      "accessed": "2026-11-14"
    },
    {
      "title": "WHO 3rd International Standard for Human Serum Immunoglobulin E (NIBSC Code 11/234)",
      "url": "https://www.nibsc.org/products/brm_product_sections/who_international_standards.aspx",
      "accessed": "2026-11-14"
    },
    {
      "title": "EAACI Molecular Allergology User's Guide (Component-Resolved Diagnostics)",
      "url": "https://www.eaaci.org/resources/educational-materials/molecular-allergology-users-guide.html",
      "accessed": "2026-11-14"
    }
  ]
}