  "name": "Holmium oxide solution band positions: NIST SRM 2034 certified wavelengths at three bandwidths versus the USP-listed 1 nm reference peaks",
  "description": "Certified wavelengths of minimum transmittance for the 14 holmium oxide (4% in 10% perchloric acid) bands, taken from the NIST SRM 2034 certificate at 0.1, 1 and 3 nm spectral bandwidth, set beside the rounded reference peaks listed for USP  wavelength control, to show how far a band moves with instrument bandwidth.",
  "page": "https://casrai.org/guides/spectrophotometer-calibration-wavelength-photometric-accuracy",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-22T14:15:25Z",
  "temporalCoverage": "2016/2025",
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Holmium oxide band",
      "unit": "",
      "description": "NIST numbering of the 14 transmittance minima"
    },
    {
      "name": "NIST SRM 2034 certified minimum, 0.1 nm bandwidth",
      "unit": "nm",
      "description": ""
    },
    {
      "name": "NIST SRM 2034 certified minimum, 1 nm bandwidth",
      "unit": "nm",
      "description": ""
    },
    {
      "name": "NIST SRM 2034 certified minimum, 3 nm bandwidth",
      "unit": "nm",
      "description": ""
    },
    {
      "name": "NIST expanded uncertainty U95 at 1 nm bandwidth",
      "unit": "nm",
      "description": ""
    },
    {
      "name": "USP reference peak listed for 1.0 nm bandwidth",
      "unit": "nm",
      "description": "as reproduced in a USP 48 text mirror"
    }
  ],
  "rows": [
    {
      "Holmium oxide band": "Band 1",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 240.97,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 241.12,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 241.04,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 241.1
    },
    {
      "Holmium oxide band": "Band 2",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 249.78,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 249.89,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 250.07,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 249.9
    },
    {
      "Holmium oxide band": "Band 3",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 278.15,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 278.13,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 278.05,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 278.1
    },
    {
      "Holmium oxide band": "Band 4",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 287.03,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 287.22,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 287.57,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 287.2
    },
    {
      "Holmium oxide band": "Band 5",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 333.48,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 333.48,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 333.47,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 333.5
    },
    {
      "Holmium oxide band": "Band 6",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 345.46,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 345.38,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 345.53,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 345.4
    },
    {
      "Holmium oxide band": "Band 7",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 361.27,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 361.25,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 361.11,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 361.3
    },
    {
      "Holmium oxide band": "Band 8",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 385.36,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 385.61,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 386,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.04,
      "USP reference peak listed for 1.0 nm bandwidth": 385.6
    },
    {
      "Holmium oxide band": "Band 9",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 416.02,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 416.25,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 416.89,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 416.3
    },
    {
      "Holmium oxide band": "Band 10",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": "not certified (band splits)",
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 451.45,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 451.36,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.05,
      "USP reference peak listed for 1.0 nm bandwidth": 451.4
    },
    {
      "Holmium oxide band": "Band 11",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 467.78,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 467.82,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 468.11,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.04,
      "USP reference peak listed for 1.0 nm bandwidth": 467.8
    },
    {
      "Holmium oxide band": "Band 12",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 485.2,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 485.23,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 485.21,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.04,
      "USP reference peak listed for 1.0 nm bandwidth": 485.2
    },
    {
      "Holmium oxide band": "Band 13",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 536.42,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 536.56,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 537.21,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.04,
      "USP reference peak listed for 1.0 nm bandwidth": 536.6
    },
    {
      "Holmium oxide band": "Band 14",
      "NIST SRM 2034 certified minimum, 0.1 nm bandwidth": 640.41,
      "NIST SRM 2034 certified minimum, 1 nm bandwidth": 640.5,
      "NIST SRM 2034 certified minimum, 3 nm bandwidth": 641.15,
      "NIST expanded uncertainty U95 at 1 nm bandwidth": 0.04,
      "USP reference peak listed for 1.0 nm bandwidth": 640.5
    }
  ],
  "sources": [
    {
      "title": "NIST SRM 2034 Certificate, Holmium Oxide Solution Wavelength Standard (issued 29 August 2016)",
      "url": "https://tsapps.nist.gov/srmext/certificates/2034.pdf",
      "accessed": "2026-11-22"
    },
    {
      "title": "USP Ultraviolet-Visible Spectroscopy (USP 48-NF 43 text as reproduced)",
      "url": "https://trungtamthuoc.com/usp-en/ultraviolet-visible-spectroscopy",
      "accessed": "2026-11-22"
    }
  ]
}