  "name": "SQUID magnetometry modes and interpretation checks",
  "description": "Original qualitative comparison compiled from Quantum Design’s MPMS3 product description and separate application notes on AC susceptibility and background subtraction. It distinguishes specimen motion from alternating-field excitation and identifies the mode-specific record needed to interpret the result. This is a method reference, not measured performance data or a ranking.",
  "page": "https://casrai.org/guides/squid-magnetometry",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-21T13:20:00Z",
  "temporalCoverage": null,
  "spatialCoverage": null,
  "columns": [
    {
      "name": "Measurement mode",
      "unit": "",
      "description": ""
    },
    {
      "name": "Signal generation",
      "unit": "",
      "description": ""
    },
    {
      "name": "Response",
      "unit": "",
      "description": ""
    },
    {
      "name": "Interpretation check",
      "unit": "",
      "description": ""
    },
    {
      "name": "Primary source references",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Measurement mode": "DC scan",
      "Signal generation": "Sample moves through pickup gradiometer under a DC field",
      "Response": "Moment obtained by fitting voltage versus sample position",
      "Interpretation check": "For the background correction described in note 1500-023, subtract raw voltage waveforms before refitting the sample waveform.",
      "Primary source references": "https://qdusa.com/products/mpms3.html ; https://qdusa.com/siteDocs/appNotes/1500-023.pdf"
    },
    {
      "Measurement mode": "SQUID-VSM",
      "Signal generation": "Vibrating sample generates a time-dependent detector signal under a DC field",
      "Response": "Magnetic moment",
      "Interpretation check": "Note 1500-023 describes point-by-point moment subtraction; match measurement sequence, holder geometry and rod rotation.",
      "Primary source references": "https://qdusa.com/products/mpms3.html ; https://qdusa.com/siteDocs/appNotes/1500-023.pdf"
    },
    {
      "Measurement mode": "AC susceptibility",
      "Signal generation": "Small alternating magnetic drive is superimposed on a DC field",
      "Response": "In-phase (real) and out-of-phase (imaginary) susceptibility",
      "Interpretation check": "Record frequency, drive amplitude, DC bias field and temperature; the imaginary response is associated with dissipative processes.",
      "Primary source references": "https://qdusa.com/siteDocs/appNotes/1078-201.pdf ; https://qdusa.com/products/mpms3.html"
    }
  ],
  "sources": [
    {
      "title": "Quantum Design MPMS3",
      "url": "https://qdusa.com/products/mpms3.html",
      "accessed": "2026-11-14"
    },
    {
      "title": "Quantum Design 1500-023: Background subtraction using the MPMS3",
      "url": "https://qdusa.com/siteDocs/appNotes/1500-023.pdf",
      "accessed": "2026-11-14"
    },
    {
      "title": "Quantum Design 1078-201: Introduction to AC Susceptibility",
      "url": "https://qdusa.com/siteDocs/appNotes/1078-201.pdf",
      "accessed": "2026-11-14"
    }
  ]
}