  "name": "EPA Method 7000B: numeric design values and acceptance criteria for standard addition",
  "description": "Numeric criteria that EPA SW-846 Method 7000B gives for deciding when to use the method of standard additions and how to design and validate it, compiled into one table from sections 9.8 and 9.10 of the method text. Values are percentages or ratios exactly as stated in the method.",
  "page": "https://casrai.org/guides/standard-addition",
  "publisher": "CASRAI",
  "license": "https://creativecommons.org/licenses/by/4.0/",
  "dateModified": "2026-11-17T05:46:17Z",
  "temporalCoverage": "Revision 2, February 2007",
  "spatialCoverage": "United States (EPA SW-846)",
  "columns": [
    {
      "name": "Parameter",
      "unit": "",
      "description": "What the criterion controls"
    },
    {
      "name": "Value",
      "unit": "",
      "description": "Numeric value stated in the method"
    },
    {
      "name": "Unit",
      "unit": "",
      "description": "Unit of the value"
    },
    {
      "name": "Role in the method",
      "unit": "",
      "description": "Where the value applies"
    },
    {
      "name": "EPA 7000B section",
      "unit": "",
      "description": ""
    }
  ],
  "rows": [
    {
      "Parameter": "Post-digestion spike recovery, lower limit",
      "Value": 80,
      "Unit": "percent",
      "Role in the method": "Recovery below this window fails the matrix-effect test and triggers standard additions for the batch",
      "EPA 7000B section": "9.8"
    },
    {
      "Parameter": "Post-digestion spike recovery, upper limit",
      "Value": 120,
      "Unit": "percent",
      "Role in the method": "Recovery above this window fails the matrix-effect test",
      "EPA 7000B section": "9.8"
    },
    {
      "Parameter": "Dilution test agreement (1:5 dilution)",
      "Value": 10,
      "Unit": "percent (plus or minus)",
      "Role in the method": "Diluted result must agree within this tolerance of the original result",
      "EPA 7000B section": "9.8"
    },
    {
      "Parameter": "Spike level, lower bound",
      "Value": 10,
      "Unit": "times LLOQ",
      "Role in the method": "Spike concentration range for the post-digestion spike",
      "EPA 7000B section": "9.8"
    },
    {
      "Parameter": "Spike level, upper bound",
      "Value": 100,
      "Unit": "times LLOQ",
      "Role in the method": "Spike concentration range for the post-digestion spike",
      "EPA 7000B section": "9.8"
    },
    {
      "Parameter": "Linearity of calibration curve (minimum)",
      "Value": 0.995,
      "Unit": "correlation coefficient",
      "Role in the method": "Validity condition for using standard additions",
      "EPA 7000B section": "9.10.3"
    },
    {
      "Parameter": "Series addition 1",
      "Value": 50,
      "Unit": "percent of expected endogenous absorbance",
      "Role in the method": "First addition in a series of additions",
      "EPA 7000B section": "9.10.2"
    },
    {
      "Parameter": "Series addition 2",
      "Value": 100,
      "Unit": "percent of expected endogenous absorbance",
      "Role in the method": "Second addition in a series of additions",
      "EPA 7000B section": "9.10.2"
    },
    {
      "Parameter": "Series addition 3",
      "Value": 150,
      "Unit": "percent of expected endogenous absorbance",
      "Role in the method": "Third addition in a series of additions",
      "EPA 7000B section": "9.10.2"
    },
    {
      "Parameter": "Single addition signal ratio SA to SB",
      "Value": 2,
      "Unit": "ratio (approximate)",
      "Role in the method": "Spike chosen so the spiked signal is about twice the unspiked signal",
      "EPA 7000B section": "9.10.1"
    }
  ],
  "sources": [
    {
      "title": "U.S. EPA SW-846 Method 7000B, Flame Atomic Absorption Spectrophotometry, Revision 2, February 2007",
      "url": "https://www.epa.gov/sites/default/files/2015-12/documents/7000b.pdf",
      "accessed": "2026-11-17"
    }
  ]
}